Prospective controlled animal study on biopsy sampling with new flexible cryoprobes versus forceps: evaluation of biopsy size, histological quality and bleeding risk

Affiliations
Comparative Study

Prospective controlled animal study on biopsy sampling with new flexible cryoprobes versus forceps: evaluation of biopsy size, histological quality and bleeding risk

Karl-Josef Franke et al. Respiration. 2010.

Abstract

Background: Cryoextraction is a procedure used for the recanalization of obstructed airways caused by visible exophytic endobronchial tumor. Biopsy samples obtained by this technique have been shown to be useful for histological assessment.

Objectives: The aim of the present animal study was to systematically evaluate biopsy size, histological quality and bleeding risk after cryobiopsy with new, flexible cryoprobes in comparison with forceps biopsy, serving as the gold standard.

Methods: Biopsies were obtained from anesthetized pigs with the flexible bronchoscopy technique, and evaluated histologically with respect to their size and quality. Bleeding frequency, bleeding duration and histological changes in the biopsy bed were also recorded.

Results: Cryobiopsies were significantly larger than forceps biopsies. The size of cryobiopsies was dependent on the freezing time. The histological quality of the cryobiopsy specimenswas not impaired by the freezing process, whereas forceps biopsies showed typical crush artifacts. Despite the larger defects left in the tracheobronchial system after cryobiopsy, bleeding frequency and duration were not higher compared to forceps biopsy.

Conclusions: Since cryobiopsy sampling is not associated with a higher bleeding risk compared with forceps biopsy, this new biopsy technique offers--in addition to a good specimen quality--a safe and valuable tool with the potential of improving the outcome of diagnostic endoscopy.