Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials
- PMID: 20162104
Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials
Abstract
Purpose: Resonance frequency analysis (RFA) offers the opportunity to monitor the osseointegration of an implant in a simple, noninvasive way. A better comprehension of the relationship between RFA and parameters related to bone quality would therefore help clinicians improve diagnoses. In this study, a bone analog made from polyurethane foam was used to isolate the influences of bone density and cortical thickness in RFA.
Materials and methods: Straumann standard implants were inserted in polyurethane foam blocks, and primary implant stability was measured with RFA. The blocks were composed of two superimposed layers with different densities. The top layer was dense to mimic cortical bone, whereas the bottom layer had a lower density to represent trabecular bone. Different densities for both layers and different thicknesses for the simulated cortical layer were tested, resulting in eight different block combinations. RFA was compared with two other mechanical evaluations of primary stability: removal torque and axial loading response.
Results: The primary stability measured with RFA did not correlate with the two other methods, but there was a significant correlation between removal torque and the axial loading response (P < .005). Statistical analysis revealed that each method was sensitive to different aspects of bone quality. RFA was the only method able to detect changes in both bone density and cortical thickness. However, changes in trabecular bone density were easier to distinguish with removal torque and axial loading than with RFA.
Conclusions: This study shows that RFA, removal torque, and axial loading are sensitive to different aspects of the bone-implant interface. This explains the absence of correlation among the methods and proves that no standard procedure exists for the evaluation of primary stability.
Similar articles
-
Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study.Clin Oral Implants Res. 2010 Feb;21(2):213-20. doi: 10.1111/j.1600-0501.2009.01823.x. Clin Oral Implants Res. 2010. PMID: 20070754
-
Meta-analysis of methods used to assess implant stability.Int J Oral Maxillofac Implants. 2009 Nov-Dec;24(6):1015-32. Int J Oral Maxillofac Implants. 2009. PMID: 20162105
-
Differences in implant stability associated with various methods of preparation of the implant bed: an in vitro study.J Prosthet Dent. 2012 Jun;107(6):366-72. doi: 10.1016/S0022-3913(12)60092-4. J Prosthet Dent. 2012. PMID: 22633592
-
Validity and clinical significance of biomechanical testing of implant/bone interface.Clin Oral Implants Res. 2006 Oct;17 Suppl 2:2-7. doi: 10.1111/j.1600-0501.2006.01365.x. Clin Oral Implants Res. 2006. PMID: 16968377 Review.
-
Assessment of implant stability as a prognostic determinant.Int J Prosthodont. 1998 Sep-Oct;11(5):491-501. Int J Prosthodont. 1998. PMID: 9922740 Review.
Cited by
-
Relationship between Implant Geometry and Primary Stability in Different Bony Defects and Variant Bone Densities: An In Vitro Study.Materials (Basel). 2020 Sep 30;13(19):4349. doi: 10.3390/ma13194349. Materials (Basel). 2020. PMID: 33007841 Free PMC article.
-
Effects of implant tilting and the loading direction on the displacement and micromotion of immediately loaded implants: an in vitro experiment and finite element analysis.J Periodontal Implant Sci. 2017 Aug;47(4):251-262. doi: 10.5051/jpis.2017.47.4.251. Epub 2017 Aug 28. J Periodontal Implant Sci. 2017. PMID: 28861289 Free PMC article.
-
Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis.J Clin Exp Dent. 2016 Jul 1;8(3):e307-11. doi: 10.4317/jced.52873. eCollection 2016 Jul. J Clin Exp Dent. 2016. PMID: 27398182 Free PMC article.
-
Influence of cortical bone anchorage on the primary stability of dental implants.Oral Maxillofac Surg. 2018 Sep;22(3):297-301. doi: 10.1007/s10006-018-0705-y. Epub 2018 Jun 6. Oral Maxillofac Surg. 2018. PMID: 29876688
-
Retrospective clinical study of ultrawide implants more than 6 mm in diameter.Maxillofac Plast Reconstr Surg. 2016 Aug 5;38(1):30. doi: 10.1186/s40902-016-0075-z. eCollection 2016 Dec. Maxillofac Plast Reconstr Surg. 2016. PMID: 27547748 Free PMC article.