Alarm characterization for a continuous glucose monitor that replaces traditional blood glucose monitoring
Abstract
Background: Continuous glucose monitoring (CGM) devices available in the United States are approved for use as adjuncts to self-monitoring of blood glucose (SMBG); all CGM alarms require SMBG confirmation before treatment. In this report, an analysis method is proposed to determine the CGM threshold alarm accuracy required to eliminate SMBG confirmation.
Method: The proposed method builds on the Clinical and Laboratory Standards Institute (CLSI) guideline for evaluating CGM threshold alarms using data from an in-clinic study of subjects with type 1 diabetes. The CLSI method proposes a maximum time limit of +/-30 minutes for the detection of hypo- and hyperglycemic events but does not include limits for glucose measurement accuracy. The International Standards Organization (ISO) standard for SMBG glucose measurement accuracy (ISO 15197) is +/-15 mg/dl for glucose <75 mg/dl and +/-20% for glucose > or = 75 mg/dl. This standard was combined with the CLSI method to more completely characterize the accuracy of CGM alarms.
Results: Incorporating the ISO 15197 accuracy margins, FreeStyle Navigator CGM system alarms detected 70 mg/dl hypoglycemia within 30 minutes at a rate of 70.3%, with a false alarm rate of 11.4%. The device detected high glucose in the range of 140-300 mg/dl within 30 minutes at an average rate of 99.2%, with a false alarm rate of 2.1%.
Conclusion: Self-monitoring of blood glucose confirmation is necessary for detecting and treating hypoglycemia with the FreeStyle Navigator CGM system, but at high glucose levels, SMBG confirmation adds little incremental value to CGM alarms.
2010 Diabetes Technology Society.
Figures
Similar articles
-
FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.J Diabetes Sci Technol. 2011 Jan 1;5(1):99-106. doi: 10.1177/193229681100500114. J Diabetes Sci Technol. 2011. PMID: 21303631 Free PMC article. Clinical Trial.
-
Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose.J Diabetes Sci Technol. 2010 Jan 1;4(1):41-8. doi: 10.1177/193229681000400106. J Diabetes Sci Technol. 2010. PMID: 20167166 Free PMC article. Clinical Trial.
-
Long-Term Home Study on Nocturnal Hypoglycemic Alarms Using a New Fully Implantable Continuous Glucose Monitoring System in Type 1 Diabetes.Diabetes Technol Ther. 2015 Nov;17(11):780-6. doi: 10.1089/dia.2014.0375. Epub 2015 Jul 15. Diabetes Technol Ther. 2015. PMID: 26177299
-
Hypo- and Hyperglycemic Alarms: Devices and Algorithms.J Diabetes Sci Technol. 2015 Apr 30;9(5):1126-37. doi: 10.1177/1932296815583507. J Diabetes Sci Technol. 2015. PMID: 25931581 Free PMC article. Review.
-
Benefits and limitations of continuous glucose monitoring in type 1 diabetes.Expert Rev Endocrinol Metab. 2020 Jan;15(1):41-49. doi: 10.1080/17446651.2020.1706482. Epub 2020 Jan 11. Expert Rev Endocrinol Metab. 2020. PMID: 31928104 Review.
Cited by
-
FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.J Diabetes Sci Technol. 2011 Jan 1;5(1):99-106. doi: 10.1177/193229681100500114. J Diabetes Sci Technol. 2011. PMID: 21303631 Free PMC article. Clinical Trial.
-
Continuous glucose monitoring and trend accuracy: news about a trend compass.J Diabetes Sci Technol. 2014 Sep;8(5):986-97. doi: 10.1177/1932296814533091. Epub 2014 May 16. J Diabetes Sci Technol. 2014. PMID: 24876437 Free PMC article.
-
Clinically relevant hypoglycemia prediction metrics for event mitigation.Diabetes Technol Ther. 2012 Aug;14(8):719-27. doi: 10.1089/dia.2011.0198. Epub 2012 Jun 12. Diabetes Technol Ther. 2012. PMID: 22690875 Free PMC article.
References
-
- Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL. Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data. Diabetes Care. 2004;27(8):1922–1928. - PubMed
-
- Clinical and Laboratory Standards Institute. Performance metrics for continuous interstitial glucose monitoring; approved guideline. 2008. POCT05-A (28)33.
-
- International Standard Organization. In vitro diagnostic test systems: requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. 2003. ISO 15197:2003(E)
-
- Weinstein RL, Schwartz SL, Brazg RL, Bugler JR, Peyser TA, McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements. Diabetes Care. 2007;30(5):1125–1130. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical