Photoplethysmography. Part 2. Influence of light source wavelength
- PMID: 2016920
- DOI: 10.1007/BF02446295
Photoplethysmography. Part 2. Influence of light source wavelength
Abstract
When the microvascular blood perfusion in human skin is measured by photoplethysmography (PPG), infra-red light (800-960 nm) is normally used as the light source. The PPG signal, which consists of a pulsatile (AC) and a slowly fluctuating (DC) component, was studied at different optical wavelengths utilising optical fibres for guiding the light to and from the skin surface. Finger and forearm skin was examined and high and low skin blood perfusion was brought about by local water-induced temperature provocation. The analysis of the measurement results provided evidence that the use of shorter wavelengths in PPG (AC) for monitoring skin perfusion changes could be applicable. The use of different optical wavelengths also raises the possibility of recording perfusing changes at different depths in the superficial tissue. The sweat water content in stratum corneum of human skin will probably determine the total amount of reflected and backscattered radiation reaching the photodetector. This is important when the skin perfusion is changed by alterations in the environmental temperature conditions activating the sweat glands in tissue. Temperature-dependent optical characteristics of blood-free skin tissue may explain the limited ability of the DC component of PPG to monitor skin perfusion changes.
Similar articles
-
Photoplethysmography. Part 1. Comparison with laser Doppler flowmetry.Med Biol Eng Comput. 1991 Jan;29(1):40-7. doi: 10.1007/BF02446294. Med Biol Eng Comput. 1991. PMID: 2016919
-
Investigating optical path and differential pathlength factor in reflectance photoplethysmography for the assessment of perfusion.J Biomed Opt. 2018 Jul;23(7):1-11. doi: 10.1117/1.JBO.23.7.075005. J Biomed Opt. 2018. PMID: 29998648
-
Influence of skin type and wavelength on light wave reflectance.J Clin Monit Comput. 2013 Jun;27(3):313-7. doi: 10.1007/s10877-013-9436-7. Epub 2013 Feb 9. J Clin Monit Comput. 2013. PMID: 23397431
-
Photoplethysmography and its application in clinical physiological measurement.Physiol Meas. 2007 Mar;28(3):R1-39. doi: 10.1088/0967-3334/28/3/R01. Epub 2007 Feb 20. Physiol Meas. 2007. PMID: 17322588 Review.
-
[Methods of microcirculatory monitoring (laser Doppler flowmetry, photoplethysmography and computer-assisted venous occlusion plethysmography)].Anasthesiol Intensivmed Notfallmed Schmerzther. 1996 May;31(1 Suppl):S37-43. Anasthesiol Intensivmed Notfallmed Schmerzther. 1996. PMID: 8737539 Review. German. No abstract available.
Cited by
-
Visual and psychological stress during computer work in healthy, young females-physiological responses.Int Arch Occup Environ Health. 2018 Oct;91(7):811-830. doi: 10.1007/s00420-018-1324-5. Epub 2018 May 30. Int Arch Occup Environ Health. 2018. PMID: 29850947 Free PMC article.
-
iPhone 4s photoplethysmography: which light color yields the most accurate heart rate and normalized pulse volume using the iPhysioMeter Application in the presence of motion artifact?PLoS One. 2014 Mar 11;9(3):e91205. doi: 10.1371/journal.pone.0091205. eCollection 2014. PLoS One. 2014. PMID: 24618594 Free PMC article.
-
Monitoring of cardiovascular reactivity to cold stress using digital volume pulse characteristics in health and diabetes.J Clin Monit Comput. 2009 Apr;23(2):123-30. doi: 10.1007/s10877-009-9174-z. Epub 2009 Mar 24. J Clin Monit Comput. 2009. PMID: 19308667
-
Evaluation of biases in remote photoplethysmography methods.NPJ Digit Med. 2021 Jun 3;4(1):91. doi: 10.1038/s41746-021-00462-z. NPJ Digit Med. 2021. PMID: 34083724 Free PMC article.
-
Monitoring of respiratory and heart rates using a fibre-optic sensor.Med Biol Eng Comput. 1992 Sep;30(5):533-7. doi: 10.1007/BF02457833. Med Biol Eng Comput. 1992. PMID: 1293445
References
Publication types
MeSH terms
LinkOut - more resources
Other Literature Sources