Independent evaluation of a commercial deconvolution reporting software for gas chromatography mass spectrometry analysis of pesticide residues in fruits and vegetables
- PMID: 20172528
- DOI: 10.1016/j.chroma.2010.01.022
Independent evaluation of a commercial deconvolution reporting software for gas chromatography mass spectrometry analysis of pesticide residues in fruits and vegetables
Abstract
The gas chromatography mass spectrometry (GC-MS) deconvolution reporting software (DRS) from Agilent Technologies has been evaluated for its ability as a screening tool to detect a large number of pesticides in incurred and fortified samples extracted with acetone/dichloromethane/light petroleum (Mini-Luke method). The detection of pesticides is based on fixed retention times using retention time locking (RTL) and full scan mass spectral comparison with a partly customer built automated mass spectral deconvolution and identification system (AMDIS) database. The GC-MS was equipped with a programmable temperature vaporising (PTV) injector system which enables more sample to be injected. In a blind study of 52 real samples a total number of 158 incurred pesticides were found. In addition to the 85 pesticides found by manual interpretation of GC-NPD/ECD chromatograms, the DRS revealed 73 more pesticides (+46%). The DRS system also shows its potential to discover pesticides which are normally not searched for (EPN in long beans from Thailand). A spiking experiment was performed to blank matrices of apple, orange and lettuce with 177 different pesticides at concentration levels 0.02 and 0.1 mg/kg. The samples were analysed on GC-MS full scan and the AMDIS match factor was used as a mass spectral quality criterion. The threshold level of the AMDIS match factor was set at 20 to eliminate most of the false positives. AMDIS match factors from 20 up to 69 are regarded only as indication of a positive hit and must be followed by manual interpretation. Pesticides giving AMDIS match factors at > or = 70 are regarded as identified. To simplify and decrease the large amount of data generated at each concentration level, the AMDIS match factors > or = 20 was averaged (mean AMF) for each pesticide including the commodities and their replicates. Among 177 different pesticides spiked at 0.02 and 0.1 mg/kg level, the percentage of mean AMF values > or = 70 were 23% and 80%, respectively. For 531 individual detections of pesticides (177 pesticides x 3 replicates) giving AMDIS match factor 20 in apple, orange and lettuce, the detection rates at 0.02 mg/kg were 71%, 63% and 72%, respectively. For the 0.1 mg/kg level the detection rates were 89%, 85% and 89%, respectively. In real samples some manual interpretation must be performed in addition. However, screening by GC-MS/DRS is about 5-10 times faster compared to screening with GC-NPD/ECD because the time used for manual interpretation is much shorter and there is no need for re-injection on GC-MS for the identification of suspect peaks found on GC-NPD/ECD.
2010 Elsevier B.V. All rights reserved.
Similar articles
-
Qualitative aspects in the analysis of pesticide residues in fruits and vegetables using fast, low-pressure gas chromatography-time-of-flight mass spectrometry.J Agric Food Chem. 2011 Jul 27;59(14):7544-56. doi: 10.1021/jf104606j. Epub 2011 Apr 13. J Agric Food Chem. 2011. PMID: 21452898
-
Fast and sensitive determination of pesticide residues in vegetables using low-pressure gas chromatography with a triple quadrupole mass spectrometer.J Chromatogr A. 2006 Sep 22;1128(1-2):236-43. doi: 10.1016/j.chroma.2006.06.044. Epub 2006 Jul 3. J Chromatogr A. 2006. PMID: 16820160
-
Application of automated mass spectrometry deconvolution and identification software for pesticide analysis in surface waters.J Environ Sci Health B. 2006;41(8):1259-71. doi: 10.1080/03601230600962211. J Environ Sci Health B. 2006. PMID: 17090491
-
Analytical methods for estimation of organophosphorus pesticide residues in fruits and vegetables: a review.Talanta. 2010 Sep 15;82(4):1077-89. doi: 10.1016/j.talanta.2010.06.043. Epub 2010 Jul 23. Talanta. 2010. PMID: 20801302 Review.
-
Liquid chromatography-high-resolution mass spectrometry for pesticide residue analysis in fruit and vegetables: screening and quantitative studies.J Chromatogr A. 2013 Apr 26;1287:24-37. doi: 10.1016/j.chroma.2013.02.065. Epub 2013 Feb 27. J Chromatogr A. 2013. PMID: 23535633 Review.
Cited by
-
Leaching potential of phenylurea herbicides in a calcareous soil: comparison of column elution and batch studies.Environ Sci Pollut Res Int. 2014 Apr;21(7):4906-13. doi: 10.1007/s11356-012-1244-y. Epub 2012 Oct 23. Environ Sci Pollut Res Int. 2014. PMID: 23097070
-
Screening for More than 1,000 Pesticides and Environmental Contaminants in Cannabis by GC/Q-TOF.Med Cannabis Cannabinoids. 2020 Jan 9;3(1):14-24. doi: 10.1159/000504391. eCollection 2020 Aug. Med Cannabis Cannabinoids. 2020. PMID: 34676338 Free PMC article.
-
Rapid Simultaneous Determination of 43 Pesticide Residues in Schizonepeta tenuifolia by Gas Chromatography Mass Spectrometry.Int J Anal Chem. 2021 Dec 13;2021:8934998. doi: 10.1155/2021/8934998. eCollection 2021. Int J Anal Chem. 2021. PMID: 34938338 Free PMC article.
-
Evaluation of a Simplified Method for GC/MS Qualitative Analysis of Polycyclic Aromatic Hydrocarbons, Polychlorinated Biphenyls, and Organic Pesticides Using PARADISe Computer Program.Molecules. 2020 Aug 15;25(16):3727. doi: 10.3390/molecules25163727. Molecules. 2020. PMID: 32824143 Free PMC article.
-
An untargeted metabolomic workflow to improve structural characterization of metabolites.Anal Chem. 2013 Aug 20;85(16):7713-9. doi: 10.1021/ac400751j. Epub 2013 Aug 2. Anal Chem. 2013. PMID: 23829391 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous