Use of a flexible robotic transgastric natural orifice translumenal endoscopic surgery (NOTES) platform in a cadaver to test access, navigation, maneuverability, and stability
- PMID: 20174943
- DOI: 10.1007/s00464-010-0905-5
Use of a flexible robotic transgastric natural orifice translumenal endoscopic surgery (NOTES) platform in a cadaver to test access, navigation, maneuverability, and stability
Abstract
Background: The advancement of natural orifice translumenal endoscopic surgery (NOTES) depends on the availability of a suitable platform. A 2008 Natural Orifice Surgery Consortium for Assessment and Research (NOSCAR) Working Group identified access, navigation, maneuverability, and stability to withstand instrument forces as the essential requirements for a successful NOTES platform [1]. No single NOTES platform can adequately achieve all four of these key capabilities. In this study, the authors tested a novel flexible endoscopic robotic platform in a cadaver to determine how it performs with respect to these specific requirements.
Methods: A highly maneuverable multichannel flexible robotic endoscopic platform developed for NOTES was used transgastrically in an adult cadaver to test the platform's ability to reach multiple intraabdominal targets. The device was under the control of the surgeon throughout the test. The surgeon was blinded to a control laparoscopic view. At each target organ, the platform was maneuvered via a joystick to provide different camera views and tool access orientation appropriate for the surgical tasks of each procedure. Standard endoscopic tools were used in the platform's two working channels to demonstrate the platform's ability to withstand the forces generated during tissue manipulation and clipping.
Results: The platform reached each target organ site without difficulty, relying on the endoscopic view only. The platform's unique ability to maintain stability in two- and three-dimensional space resisted forces exerted by tools in the tool channels used for tissue manipulation and clipping. Endoscopic visualization allowed the surgeon to reposition the platform easily to gain a different camera view or tool access orientation about an organ (see video in Supplementary material).
Conclusion: The authors demonstrated that the four capabilities identified by NOSCAR can be provided by the flexible robotic endoscopic platform. These essential capabilities for the clinical implementation of NOTES were addressed specifically by this platform. The provision of these capabilities in a single device may further the advancement and adoption of NOTES.
Similar articles
-
A technical review of flexible endoscopic multitasking platforms.Int J Surg. 2012;10(7):345-54. doi: 10.1016/j.ijsu.2012.05.009. Epub 2012 May 26. Int J Surg. 2012. PMID: 22641123 Review.
-
Multifunction robotic platform for natural orifice surgery.Stud Health Technol Inform. 2011;163:740-2. Stud Health Technol Inform. 2011. PMID: 21335891
-
Robotics in endoscopy.Curr Opin Gastroenterol. 2012 Sep;28(5):477-82. doi: 10.1097/MOG.0b013e328356ac5e. Curr Opin Gastroenterol. 2012. PMID: 22885946 Review.
-
Robotic natural orifice translumenal endoscopic surgery and laparoendoscopic single-site surgery: current status.Curr Opin Urol. 2011 Jan;21(1):71-7. doi: 10.1097/MOU.0b013e32833fd602. Curr Opin Urol. 2011. PMID: 20962649 Review.
-
Robotic retroperitoneal transvaginal natural orifice translumenal endoscopic surgery (NOTES) nephrectomy: feasibility study in a cadaver model.Urology. 2013 Jun;81(6):1232-7. doi: 10.1016/j.urology.2012.11.083. Epub 2013 Mar 27. Urology. 2013. PMID: 23541230
Cited by
-
Natural orifice translumenal endoscopic surgery (NOTES(®)): a technical review.Surg Endosc. 2011 Oct;25(10):3135-48. doi: 10.1007/s00464-011-1718-x. Epub 2011 May 7. Surg Endosc. 2011. PMID: 21553172 Review.
-
Triangulation: the holy grail of endoscopic surgery?Surg Endosc. 2011 May;25(5):1355-7. doi: 10.1007/s00464-011-1650-0. Surg Endosc. 2011. PMID: 21424191 No abstract available.
-
A Survey of Transoral Robotic Mechanisms: Distal Dexterity, Variable Stiffness, and Triangulation.Cyborg Bionic Syst. 2023;4:0007. doi: 10.34133/cbsystems.0007. Epub 2023 Mar 13. Cyborg Bionic Syst. 2023. PMID: 37058618 Free PMC article. Review.
-
On the suitability of Thiel cadavers for natural orifice transluminal endoscopic surgery (NOTES): surgical training, feasibility studies, and anatomical education.Surg Endosc. 2015 Mar;29(3):737-46. doi: 10.1007/s00464-014-3734-0. Epub 2014 Jul 25. Surg Endosc. 2015. PMID: 25060684
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources