Photon beam quality variations of a flattening filter free linear accelerator
- PMID: 20175465
- DOI: 10.1118/1.3264617
Photon beam quality variations of a flattening filter free linear accelerator
Abstract
Purpose: Recently, there has been an increasing interest in operating conventional linear accelerators without a flattening filter. The aim of this study was to determine beam quality variations as a function of off-axis ray angle for unflattened beams. In addition, a comparison was made with the off-axis energy variation in flattened beams.
Methods: Two Elekta Precise linear accelerators were modified in order to enable radiation delivery with and without the flattening filter in the beam line. At the Medical University Vienna (Vienna, Austria), half value layer (HVL) measurements were performed for 6 and 10 MV with an in-house developed device that can be easily mounted on the gantry. At St. Luke's Hospital (Dublin, Ireland), measurements were performed at 6 MV in narrow beam geometry with the gantry tilted around 270 degrees with pinhole collimators, an attenuator, and the chamber positioned on the table. All attenuation measurements were performed with ionization chambers and a buildup cap (2 mm brass) or a PMMA mini phantom (diameter 3 cm, measurement depth 2.5 cm).
Results: For flattened 6 and 10 MV photon beams from the Elekta linac the relative HVL(theta) varies by about 11% for an off-axis ray angle theta = 10 degrees. These results agree within +/- 2% with a previously proposed generic off-axis energy correction. For unflattened beams, the variation was less than 5% in the whole range of off-axis ray angles up to 10 degrees. The difference in relative HVL data was less than 1% for unflattened beams at 6 and 10 MV.
Conclusions: Off-axis energy variation is rather small in unflattened beams and less than half the one for flattened beams. Thus, ignoring the effect of off-axis energy variation for dose calculations in unflattened beams can be clinically justified.
Similar articles
-
Experimental investigation of the response of an a-Si EPID to an unflattened photon beam from an Elekta Precise linear accelerator.Med Phys. 2009 Apr;36(4):1318-29. doi: 10.1118/1.3089424. Med Phys. 2009. PMID: 19472639
-
Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses.Z Med Phys. 2011 May;21(2):91-101. doi: 10.1016/j.zemedi.2010.07.003. Epub 2010 Oct 2. Z Med Phys. 2011. PMID: 20888199
-
Beam characteristics of energy-matched flattening filter free beams.Med Phys. 2014 May;41(5):052103. doi: 10.1118/1.4871615. Med Phys. 2014. PMID: 24784392
-
[A Proposal for the Absorbed Dose to Water Dosimetry for Flattening Filter-free Beams].Igaku Butsuri. 2016;36(2):79-84. doi: 10.11323/jjmp.36.2_79. Igaku Butsuri. 2016. PMID: 28428458 Review. Japanese.
-
The determination of phantom and collimator scatter components of the output of megavoltage photon beams: measurement of the collimator scatter part with a beam-coaxial narrow cylindrical phantom.Radiother Oncol. 1991 Apr;20(4):250-7. doi: 10.1016/0167-8140(91)90124-y. Radiother Oncol. 1991. PMID: 1906190 Review.
Cited by
-
Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.J Appl Clin Med Phys. 2015 May 8;16(3):5219. doi: 10.1120/jacmp.v16i3.5219. J Appl Clin Med Phys. 2015. PMID: 26103482 Free PMC article.
-
Acceptance criteria for flattening filter-free photon beam from standard medical electron linear accelerator: AERB task group recommendations.J Med Phys. 2014 Oct;39(4):206-11. doi: 10.4103/0971-6203.144482. J Med Phys. 2014. PMID: 25525307 Free PMC article.
-
Comparison of detector performance in small 6 MV and 6 MV FFF beams using a Versa HD accelerator.PLoS One. 2019 Mar 11;14(3):e0213253. doi: 10.1371/journal.pone.0213253. eCollection 2019. PLoS One. 2019. PMID: 30856183 Free PMC article.
-
Unflattened photon beams from the standard flattening filter free accelerators for radiotherapy: Advantages, limitations and challenges.J Med Phys. 2011 Jul;36(3):123-5. doi: 10.4103/0971-6203.83464. J Med Phys. 2011. PMID: 21897556 Free PMC article. No abstract available.
-
Commissioning and quality assurance of HalcyonTM 2.0 linear accelerator.Rep Pract Oncol Radiother. 2021 Jun 9;26(3):433-444. doi: 10.5603/RPOR.a2021.0065. eCollection 2021. Rep Pract Oncol Radiother. 2021. PMID: 34277097 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous