Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy with various seed models
- PMID: 20175499
- DOI: 10.1118/1.3271104
Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy with various seed models
Abstract
Purpose: Dose distributions are calculated for various models of 125I and 103Pd seeds in the standardized plaques of the Collaborative Ocular Melanoma Study (COMS). The sensitivity to seed model of dose distributions and dose distributions relative to TG-43 are investigated.
Methods: Monte Carlo simulations are carried out with the EGSnrc user-code BrachyDose. Brachytherapy seeds and eye plaques are fully modeled. Simulations of one seed in the central slot of a 20 mm Modulay (gold alloy) plaque backing with and without the Silastic (silicone polymer) insert and of a 16 mm fully loaded Modulay/Silastic plaque are performed. Dose distributions are compared to those calculated under TG-43 assumptions, i.e., ignoring the effects of the plaque backing and insert and interseed attenuation. Three-dimensional dose distributions for different 125I and 103Pd seed models are compared via depth-dose curves, isodose contours, and tabulation of doses at points of interest in the eye. Results are compared to those of our recent BrachyDose study for COMS plaques containing model 6711 (125I) or 200 (103Pd) seeds [R. M. Thomson et al., Med. Phys. 35, 5530-5543 (2008)].
Results: Along the central axis of a plaque containing one seed, variations of less than 1% are seen in the effect of the Modulay backing alone for different seed models; for the Modulay/Silastic combination, variations are 2%. For a 16 mm plaque fully loaded with 125I (103Pd) seeds, dose decreases relative to TG-43 doses are 11%-12% (19%-20%) and 14%-15% (20%) at distances of 0.5 and 1 cm from the inner sclera along the plaque's central axis, respectively. For the same prescription dose, doses at points of interest vary by up to 8% with seed model. Doses to critical normal structures are lower for all 103Pd seed models than for 125I with the possible exception of the sclera adjacent to the plaque; scleral doses vary with seed model and are not always higher for 103Pd than for 125I.
Conclusions: Dose decreases relative to doses calculated under TG-43 assumptions vary slightly with seed model (for each radionuclide). Dose distributions are sensitive to seed model; however, variations are generally no larger than the magnitudes of other systematic uncertainties in eye plaque therapy.
Similar articles
-
Monte Carlo dosimetry for 125I and 103Pd eye plaque brachytherapy.Med Phys. 2008 Dec;35(12):5530-43. doi: 10.1118/1.3002412. Med Phys. 2008. PMID: 19175111
-
Monte Carlo dosimetry for 103Pd, 125I, and 131Cs ocular brachytherapy with various plaque models using an eye phantom.Med Phys. 2014 Mar;41(3):031706. doi: 10.1118/1.4864474. Med Phys. 2014. PMID: 24593710
-
Modified COMS plaques for 125I and 103Pd iris melanoma brachytherapy.Int J Radiat Oncol Biol Phys. 2010 Nov 15;78(4):1261-9. doi: 10.1016/j.ijrobp.2009.12.002. Epub 2010 May 14. Int J Radiat Oncol Biol Phys. 2010. PMID: 20472360
-
Recommendations of the American Association of Physicists in Medicine regarding the impact of implementing the 2004 task group 43 report on dose specification for 103Pd and 125I interstitial brachytherapy.Med Phys. 2005 May;32(5):1424-39. doi: 10.1118/1.1884925. Med Phys. 2005. PMID: 15984693 Review.
-
Intratumoral injection therapies for locally advanced pancreatic cancer: systematic review.BJS Open. 2023 May 5;7(3):zrad052. doi: 10.1093/bjsopen/zrad052. BJS Open. 2023. PMID: 37254902 Free PMC article.
Cited by
-
Comparison of dose calculation methods for brachytherapy of intraocular tumors.Med Phys. 2011 Jan;38(1):306-16. doi: 10.1118/1.3523614. Med Phys. 2011. PMID: 21361199 Free PMC article.
-
Development of GATE Monte Carlo Code for Simulation and Dosimetry of New I-125 Seeds in Eye Plaque Brachytherapy.Nucl Med Mol Imaging. 2021 Apr;55(2):86-95. doi: 10.1007/s13139-020-00680-5. Epub 2021 Jan 11. Nucl Med Mol Imaging. 2021. PMID: 33968275 Free PMC article.
-
Monte Carlo investigation of dose distribution of uniformly and non-uniformly loaded standard and notched eye plaques.J Appl Clin Med Phys. 2023 Dec;24(12):e14149. doi: 10.1002/acm2.14149. Epub 2023 Sep 22. J Appl Clin Med Phys. 2023. PMID: 37738654 Free PMC article.
-
Keeping an eye on the ring: COMS plaque loading optimization for improved dose conformity and homogeneity.J Contemp Brachytherapy. 2012 Sep;4(3):165-75. doi: 10.5114/jcb.2012.30683. Epub 2012 Sep 29. J Contemp Brachytherapy. 2012. PMID: 23346146 Free PMC article.
-
Sensitivity of contributing factors to heterogeneity corrections in dosimetry of ocular brachytherapy using iodine-125 COMS plaques.J Contemp Brachytherapy. 2024 Oct;16(5):371-382. doi: 10.5114/jcb.2024.144232. Epub 2024 Oct 18. J Contemp Brachytherapy. 2024. PMID: 39719954 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous