Trapeziometacarpal joint implants can be evaluated by roentgen stereophotogrammetric analysis
- PMID: 20181771
- DOI: 10.1177/1753193409355627
Trapeziometacarpal joint implants can be evaluated by roentgen stereophotogrammetric analysis
Abstract
Both marker-based roentgen stereophotogrammetric analysis (RSA) and model-based RSA have been helpful evaluation tools in hip and knee arthroplasty. The purpose of this study was to test both model-based and marker-based RSA in the evaluation of total joint prostheses of the trapeziometacarpal joint. In a phantom study, the precision of marker-based RSA was tested with a cemented polyethylene cup and compared with the precision of model-based RSA in an uncemented Elektra screw cup. The precision of model-based RSA of the metacarpal stem was tested using an uncemented Elektra metacarpal stem. In a clinical study 11 patients had double stereo radiographs followed by RSA analysis. The precision of translation in both marker-based and model-based RSA was sufficient for clinical use, but rotation cannot be estimated with sufficient precision.
Similar articles
-
Primary cup fixation with different designs of trapeziometacarpal total joint trapezium components: a radiostereometric analysis in a pig bone model.J Hand Surg Eur Vol. 2011 May;36(4):285-90. doi: 10.1177/1753193410394526. Epub 2011 Jan 31. J Hand Surg Eur Vol. 2011. PMID: 21282215
-
Precision assessment of model-based RSA for a total knee prosthesis in a biplanar set-up.Knee. 2008 Oct;15(5):396-402. doi: 10.1016/j.knee.2008.05.001. Epub 2008 Jul 16. Knee. 2008. PMID: 18635360
-
Clinical validation of model-based RSA for a total knee prosthesis.Clin Orthop Relat Res. 2007 Nov;464:205-9. Clin Orthop Relat Res. 2007. PMID: 18062052
-
Implant arthroplasty of the carpometacarpal joint of the thumb.Hand Clin. 2010 Aug;26(3):327-37, v. doi: 10.1016/j.hcl.2010.05.007. Hand Clin. 2010. PMID: 20670798 Review.
-
Distraction arthroplasty of the trapeziometacarpal joint.J Hand Surg Am. 2007 Nov;32(9):1443-6. doi: 10.1016/j.jhsa.2007.08.017. J Hand Surg Am. 2007. PMID: 17996782 Review.
Cited by
-
Superior bone fixation of conical compared with hemispherical trapezial cup design: an experimental radiostereometry study.J Exp Orthop. 2023 Nov 30;10(1):127. doi: 10.1186/s40634-023-00692-y. J Exp Orthop. 2023. PMID: 38032446 Free PMC article.
-
Mechanical Stability of the Prodisc-C Vivo Cervical Disc Arthroplasty: A Preliminary, Observational Study Using Radiostereometric Analysis.Global Spine J. 2020 May;10(3):294-302. doi: 10.1177/2192568219850763. Epub 2019 May 23. Global Spine J. 2020. PMID: 32313795 Free PMC article.
-
Comparison of Marker-Based RSA and CT-RSA for Analyzing Micromotions After Distal Radius Osteotomy: A 1-Year Retrospective Study of 24 Patients.J Orthop Res. 2025 Mar;43(3):660-670. doi: 10.1002/jor.26031. Epub 2024 Dec 27. J Orthop Res. 2025. PMID: 39731268 Free PMC article.
-
The accuracy and precision of radiostereometric analysis in upper limb arthroplasty.Acta Orthop. 2017 Jun;88(3):320-325. doi: 10.1080/17453674.2017.1291872. Epub 2017 Mar 2. Acta Orthop. 2017. PMID: 28464752 Free PMC article.
-
Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up.Acta Orthop. 2013 Feb;84(1):98-105. doi: 10.3109/17453674.2013.765625. Epub 2013 Jan 23. Acta Orthop. 2013. PMID: 23343372 Free PMC article. Clinical Trial.