Numerical and experimental investigations of pulsatile blood flow pattern through a dysfunctional mechanical heart valve
- PMID: 20188372
- DOI: 10.1016/j.jbiomech.2010.01.029
Numerical and experimental investigations of pulsatile blood flow pattern through a dysfunctional mechanical heart valve
Abstract
Around 250,000 heart valve replacements are performed every year around the world. Due their higher durability, approximately 2/3 of these replacements use mechanical prosthetic heart valves (mainly bileaflet valves). Although very efficient, these valves can be subject to valve leaflet malfunctions. These malfunctions are usually the consequence of pannus ingrowth and/or thrombus formation and represent serious and potentially fatal complications. Hence, it is important to investigate the flow field downstream of a dysfunctional mechanical heart valve to better understand its impact on blood components (red blood cells, platelets and coagulation factors) and to improve the current diagnosis techniques. Therefore, the objective of this study will be to numerically and experimentally investigate the pulsatile turbulent flow downstream of a dysfunctional bileaflet mechanical heart valve in terms of velocity field, vortex formation and potential negative effect on blood components. The results show that the flow downstream of a dysfunctional valve was characterized by abnormally elevated velocities and shear stresses as well as large scale vortices. These characteristics can predispose to blood components damage. Furthermore, valve malfunction led to an underestimation of maximal transvalvular pressure gradient, using Doppler echocardiography, when compared to numerical results. This could be explained by the shifting of the maximal velocity towards the normally functioning leaflet. As a consequence, clinicians should try, when possible, to check the maximal velocity position not only at the central orifice but also through the lateral orifices. Finding the maximal velocity in the lateral orifice could be an indication of valve dysfunction.
Copyright 2010 Elsevier Ltd. All rights reserved.
Similar articles
-
Numerical simulation of opening process in a bileaflet mechanical heart valve under pulsatile flow condition.J Heart Valve Dis. 2003 Mar;12(2):245-55. J Heart Valve Dis. 2003. PMID: 12701798
-
Principal stress analysis in LDA measurement of the flow field downstream of 19-mm Sorin Bicarbon heart valve.Technol Health Care. 1998 Nov;6(4):259-70. Technol Health Care. 1998. PMID: 9924953
-
Flow through a defective mechanical heart valve: a steady flow analysis.Med Eng Phys. 2009 Apr;31(3):295-305. doi: 10.1016/j.medengphy.2008.07.003. Epub 2008 Aug 15. Med Eng Phys. 2009. PMID: 18707915
-
Review of numerical methods for simulation of mechanical heart valves and the potential for blood clotting.Med Biol Eng Comput. 2017 Sep;55(9):1519-1548. doi: 10.1007/s11517-017-1688-9. Epub 2017 Jul 26. Med Biol Eng Comput. 2017. PMID: 28744828 Review.
-
Relationship between some design characteristics and wear in the Bicarbon heart valve prosthesis.Int J Artif Organs. 1994 May;17(5):280-93. Int J Artif Organs. 1994. PMID: 7960197 Review.
Cited by
-
Lagrangian postprocessing of computational hemodynamics.Ann Biomed Eng. 2015 Jan;43(1):41-58. doi: 10.1007/s10439-014-1070-0. Epub 2014 Jul 25. Ann Biomed Eng. 2015. PMID: 25059889 Free PMC article. Review.
-
Purely phase-encoded MRI of turbulent flow through a dysfunctional bileaflet mechanical heart valve.MAGMA. 2014 Jun;27(3):227-35. doi: 10.1007/s10334-013-0408-1. Epub 2013 Sep 24. MAGMA. 2014. PMID: 24061612
-
Numerical Modeling of Intraventricular Flow during Diastole after Implantation of BMHV.PLoS One. 2015 May 11;10(5):e0126315. doi: 10.1371/journal.pone.0126315. eCollection 2015. PLoS One. 2015. PMID: 25961285 Free PMC article.
-
Hemodynamic Performance of Dysfunctional Prosthetic Heart Valve with the Concomitant Presence of Subaortic Stenosis: In Silico Study.Bioengineering (Basel). 2020 Aug 7;7(3):90. doi: 10.3390/bioengineering7030090. Bioengineering (Basel). 2020. PMID: 32784661 Free PMC article.
-
Smoothed particle hydrodynamics based FSI simulation of the native and mechanical heart valves in a patient-specific aortic model.Sci Rep. 2024 Mar 21;14(1):6762. doi: 10.1038/s41598-024-57177-w. Sci Rep. 2024. PMID: 38514703 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials