Abstract
A lack of initial stability of the fixation is associated with aseptic loosening of the tibial components of cementless knee prostheses. With sufficient stability after surgery, minimal relative motion between the prosthesis and bone interfaces allows osseointegation to occur thereby providing a strong prosthesis-to-bone biological attachment. Finite element modelling was used to investigate the bone-prosthesis interface micromotion and the relative risk of aseptic loosening. It was anticipated that by prescribing different joint loads representing gait and other activities, and the consideration of varying tibial-femoral contact points during knee flexion, it would influence the computational prediction of the interface micromotion. In this study, three-dimensional finite element models were set up with applied loads representing walking and stair climbing, and the relative micromotions were predicted. These results were correlated to in-vitro measurements and to the results of prior retrieval studies. Two load conditions, (i) a generic vertical joint load of 3 x body weight with 70%/30%M/L load share and antero-posterior/medial-lateral shear forces, acted at the centres of the medial and lateral compartments of the tibial tray, and (ii) a peak vertical joint load at 25% of the stair climbing cycle with corresponding antero-posterior shear force applied at the tibial-femoral contact points of the specific knee flexion angle, were found to generate interface micromotion responses which corresponded to in-vivo observations. The study also found that different loads altered the interface micromotion predicted, so caution is needed when comparing the fixation performance of various reported cementless tibial prosthetic designs if each design was evaluated with a different loading condition.
Copyright 2010 Elsevier Ltd. All rights reserved.
Similar articles
-
[Noncemented total hip arthroplasty: influence of extramedullary parameters on initial implant stability and on bone-implant interface stresses].Rev Chir Orthop Reparatrice Appar Mot. 2000 Oct;86(6):590-7. Rev Chir Orthop Reparatrice Appar Mot. 2000. PMID: 11060433 French.
-
ESB Clinical Biomechanics Award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6-10 months.Clin Biomech (Bristol). 2009 May;24(4):315-26. doi: 10.1016/j.clinbiomech.2009.01.011. Epub 2009 Mar 13. Clin Biomech (Bristol). 2009. PMID: 19285767
-
The influence of tibial component fixation techniques on resorption of supporting bone stock after total knee replacement.J Biomech. 2011 Mar 15;44(5):948-54. doi: 10.1016/j.jbiomech.2010.11.026. Epub 2011 Jan 14. J Biomech. 2011. PMID: 21236431
-
The role of the design of tibial components and stems in knee replacement.J Bone Joint Surg Br. 2012 Aug;94(8):1009-15. doi: 10.1302/0301-620X.94B8.28289. J Bone Joint Surg Br. 2012. PMID: 22844039 Review.
-
Strength requirements for internal and external prostheses.J Biomech. 1999 Apr;32(4):381-93. doi: 10.1016/s0021-9290(98)00190-0. J Biomech. 1999. PMID: 10213028 Review.
Cited by
-
Emerging ideas: Instability-induced periprosthetic osteolysis is not dependent on the fibrous tissue interface.Clin Orthop Relat Res. 2013 Jun;471(6):1758-62. doi: 10.1007/s11999-013-2896-8. Epub 2013 Mar 6. Clin Orthop Relat Res. 2013. PMID: 23463289 Free PMC article.
-
Vascular endothelial growth factor pathway promotes osseointegration and CD31hiEMCNhi endothelium expansion in a mouse tibial implant model: an animal study.Bone Joint J. 2019 Jul;101-B(7_Supple_C):108-114. doi: 10.1302/0301-620X.101B7.BJJ-2018-1473.R1. Bone Joint J. 2019. PMID: 31256654 Free PMC article.
-
Comparison between uncemented and cemented fixation for the tibial component in distal femoral replacement: a clinical and radiological study.Int Orthop. 2018 Sep;42(9):2249-2261. doi: 10.1007/s00264-018-3847-5. Epub 2018 Feb 24. Int Orthop. 2018. PMID: 29478210
-
Gut microbial-derived short-chain fatty acids and bone: a potential role in fracture healing.Eur Cell Mater. 2021 Apr 21;41:454-470. doi: 10.22203/eCM.v041a29. Eur Cell Mater. 2021. PMID: 33881768 Free PMC article. Review.
-
Additive manufactured push-fit implant fixation with screw-strength pull out.J Orthop Res. 2018 May;36(5):1508-1518. doi: 10.1002/jor.23771. Epub 2017 Nov 22. J Orthop Res. 2018. PMID: 29023901 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical