Validation of Embletta portable diagnostic system for identifying patients with suspected obstructive sleep apnoea syndrome (OSAS)
- PMID: 20199644
- DOI: 10.1111/j.1440-1843.2009.01697.x
Validation of Embletta portable diagnostic system for identifying patients with suspected obstructive sleep apnoea syndrome (OSAS)
Abstract
The Embletta portable diagnostic system is highly sensitive and specific in quantifying the AHI and differentiating obstructive and supine events when compared against hospital-based PSG in patients with suspected OSAS. The simple device may be useful for screening and diagnostic purpose when access to PSG is limited.
Background and objectives: This study aimed to evaluate the diagnostic accuracy of Embletta portable diagnostic system (PDS, Medcare, Reykjavik, Iceland) for the screening of sleep apnoea in clinical practice.
Methods: The Embletta PDS is a digital three-channel recording device that measures airflow through a nasal cannula connected to a pressure transducer, oxygen saturation plus both respiratory and abdominal movements via built-in effort and body position sensors. An AHI is determined based on recording time. Nocturnal polysomnography (Alice 4, Healthdyne, Atlanta, USA), with airflow measured by a nasal pressure transducer (PTAF2, Pro-Tech, Woodinville, WA, USA)) and Embletta PDS recordings, was performed simultaneously in consecutive patients with suspected OSA syndrome. The PSG recordings were analysed manually by a blinded investigator.
Results: Ninety subjects were recruited and 10 failed Embletta PDS studies due to measurement failure. Among the remaining 80 subjects, 63 were males. The mean age (SD) was 51.4 (11.9) years old, BMI 27.1(4.2) kg/m2, neck circumference 38.6 (3.6) cm and Epworth Sleepiness Score 9.7 (5.3). The AHI obtained by the Embletta PDS correlated closely with that obtained by PSG (Pearson correlation, r=0.979, P<0.001). Comparison of AHI based on the Embletta PDS against the PSG demonstrated high sensitivity at AHI>or=5/h (sensitivity 0.924 and specificity 0.857) and high specificity at AHI>or=20/h (sensitivity 0.853 and specificity 0.957).
Conclusions: The Embletta PDS is a highly sensitive and specific screening device in quantifying AHI when compared against PSG in patients with suspected OSA syndrome.
Similar articles
-
Validation of a portable recording device (ApneaLink) for identifying patients with suspected obstructive sleep apnoea syndrome.Intern Med J. 2009 Nov;39(11):757-62. doi: 10.1111/j.1445-5994.2008.01827.x. Epub 2008 Nov 3. Intern Med J. 2009. PMID: 19220528
-
Obstructive sleep apnoea syndrome: comparison between polysomnography and portable sleep monitoring based on jaw recordings.Neurophysiol Clin. 2011 Oct;41(4):191-8. doi: 10.1016/j.neucli.2011.09.001. Epub 2011 Oct 5. Neurophysiol Clin. 2011. PMID: 22078731
-
Validation study of a portable monitoring device for identifying OSA in a symptomatic patient population.Respirology. 2009 Mar;14(2):270-5. doi: 10.1111/j.1440-1843.2008.01439.x. Epub 2008 Dec 11. Respirology. 2009. PMID: 19210658
-
Improving diagnostic ability of blood oxygen saturation from overnight pulse oximetry in obstructive sleep apnea detection by means of central tendency measure.Artif Intell Med. 2007 Sep;41(1):13-24. doi: 10.1016/j.artmed.2007.06.002. Epub 2007 Jul 23. Artif Intell Med. 2007. PMID: 17643971 Clinical Trial.
-
Utility of portable monitoring in the diagnosis of obstructive sleep apnea.J Postgrad Med. 2015 Oct-Dec;61(4):223-9. doi: 10.4103/0022-3859.166509. J Postgrad Med. 2015. PMID: 26440391 Free PMC article. Review.
Cited by
-
Positional sensitivity as a confounder in diagnosis of severity of obstructive sleep apnea.Sleep Breath. 2013 Mar;17(1):173-9. doi: 10.1007/s11325-012-0666-6. Epub 2012 Mar 1. Sleep Breath. 2013. PMID: 22382651
-
Obstructive sleep apnea linked to wake-up strokes.J Neurol. 2012 Jul;259(7):1433-9. doi: 10.1007/s00415-011-6370-9. Epub 2012 Jan 4. J Neurol. 2012. PMID: 22215237
-
WELCOME: improving WEight controL and CO-Morbidities in children with obesity via Executive function training: study protocol for a randomized controlled trial.BMC Public Health. 2018 Aug 29;18(1):1075. doi: 10.1186/s12889-018-5950-3. BMC Public Health. 2018. PMID: 30157826 Free PMC article. Clinical Trial.
-
Comparing a combination of validated questionnaires and level III portable monitor with polysomnography to diagnose and exclude sleep apnea.J Clin Sleep Med. 2013 Dec 15;9(12):1259-66. doi: 10.5664/jcsm.3264. J Clin Sleep Med. 2013. PMID: 24340287 Free PMC article.
-
Validation of a new unattended sleep apnea monitor using two methods for the identification of hypopneas.J Clin Sleep Med. 2020 May 15;16(5):695-703. doi: 10.5664/jcsm.8324. Epub 2020 Feb 6. J Clin Sleep Med. 2020. PMID: 32024586 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous