[Accuracy of 2 electronic canal length measuring instruments]
- PMID: 2020777
[Accuracy of 2 electronic canal length measuring instruments]
Abstract
Two electronic root canal length measuring instruments were used to measure the working length to "the apex" in 88 teeth scheduled for extraction. After extraction, the canal length was measured using X-Ray and microscopic observation after section of the teeth. The accuracy of the RCM apex locator in determining canal measurement within +/- 0.5 mm from the apical foramen varied around 62%. The "NO'X" allows a success rate of 41% in the same conditions. The conditions of this experiment shows that electronic root-canal length measuring instruments do not work in case of hemorrhage or swelling.
Similar articles
-
Evaluation of accuracy of an electric device (Neosono D-SE) for the measurement of tooth length.Indian J Dent Res. 1989 Apr-Sep;1(2-3):58-65. Indian J Dent Res. 1989. PMID: 2490122
-
Clinical evaluation of the accuracy of the Evident RCM Mark II Apex Locator.J Endod. 1991 Nov;17(11):567-9. doi: 10.1016/S0099-2399(06)81723-1. J Endod. 1991. PMID: 1812206 Clinical Trial.
-
An in vivo evaluation of Endex electronic apex locator.J Endod. 1993 Apr;19(4):177-9. doi: 10.1016/S0099-2399(06)80683-7. J Endod. 1993. PMID: 8326263
-
The fundamental operating principles of electronic root canal length measurement devices.Int Endod J. 2006 Aug;39(8):595-609. doi: 10.1111/j.1365-2591.2006.01131.x. Int Endod J. 2006. PMID: 16872454 Review.
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
Publication types
MeSH terms
LinkOut - more resources
Miscellaneous