Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area
- PMID: 20207865
- PMCID: PMC2829901
- DOI: 10.4269/ajtmh.2010.09-0560
Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area
Abstract
Recently, the XE-2100 hematology analyzer was investigated in a rather small patient group; pseudoeosinophilia or abnormal white blood cell (WBC) scattergrams reported by this instrument were considered as significantly valuable diagnostic parameters in detecting vivax malaria. This study was conducted not only to assess the usefulness of pseudoeosinophilia or abnormal WBC scattergram in vivax malaria-endemic areas with large patient groups (N = 1,801) but also to investigate the correlation of parasitemia and platelet count with pseudoeosinophilia and abnormal WBC scattergrams. Of the 1,801 analyzed patients, 413 (22.9%) were found to have malaria by Wright-Giemsa stained blood smears. Overall, either pseudoeosinophilia or abnormal WBC scattergram was detected in 191 of 413 malaria patients and 4 of 1,388 patients without malaria (sensitivity = 46.2%, specificity = 99.7%). We suggest that clinical hematology laboratories using the XE-2100 analyzer should be aware of such specific parameters, even with the absence of a clinical request.
Figures
Similar articles
-
Malaria detection with the Sysmex XE-2100 hematology analyzer using pseudoeosinophilia and abnormal WBC scattergram.Ann Hematol. 2008 Sep;87(9):755-9. doi: 10.1007/s00277-008-0486-8. Epub 2008 Apr 22. Ann Hematol. 2008. PMID: 18427808
-
Usefulness of automated cell counter in detection of malaria in a cancer set up--our experience.Indian J Pathol Microbiol. 2012 Oct-Dec;55(4):467-73. doi: 10.4103/0377-4929.107782. Indian J Pathol Microbiol. 2012. PMID: 23455781
-
Pseudoeosinophilia associated with malaria infection determined in the Sysmex XE-2100 hematology analyzer.Ann Hematol. 2005 Jun;84(6):400-2. doi: 10.1007/s00277-004-0987-z. Epub 2004 Dec 23. Ann Hematol. 2005. PMID: 15616813
-
Diagnostic role of Sysmex hematology analyzer in the detection of malaria: A systematic review and meta-analysis.PLoS One. 2024 Sep 6;19(9):e0296766. doi: 10.1371/journal.pone.0296766. eCollection 2024. PLoS One. 2024. PMID: 39240990 Free PMC article.
-
[Automated hematology analysers and spurious counts Part 2. Leukocyte count and differential].Ann Biol Clin (Paris). 2012 Mar-Apr;70(2):141-54. doi: 10.1684/abc.2012.0665. Ann Biol Clin (Paris). 2012. PMID: 22484525 Review. French.
Cited by
-
Infected RBC flag/parameter provided by Mindray BC-6800 haematology analyzer aid the diagnosis of malaria.Malar J. 2019 Jul 31;18(1):262. doi: 10.1186/s12936-019-2890-z. Malar J. 2019. PMID: 31366365 Free PMC article.
-
Automated Hematology Analyzers in Diagnosis of Plasmodium vivax Malaria: an Adjunct to Conventional Microscopy.Mediterr J Hematol Infect Dis. 2014 Jun 1;6(1):e2014034. doi: 10.4084/MJHID.2014.034. eCollection 2014. Mediterr J Hematol Infect Dis. 2014. PMID: 24959331 Free PMC article.
-
Application of Sysmex XN-Series Automated Haematology Analyser in the Rapid Detection of Malaria.Indian J Hematol Blood Transfus. 2020 Jul;36(3):512-518. doi: 10.1007/s12288-020-01276-x. Epub 2020 Apr 3. Indian J Hematol Blood Transfus. 2020. PMID: 32647426 Free PMC article.
-
Performance evaluation of machine learning-based infectious screening flags on the HORIBA Medical Yumizen H550 Haematology Analyzer for vivax malaria and dengue fever.Malar J. 2020 Nov 23;19(1):429. doi: 10.1186/s12936-020-03502-3. Malar J. 2020. PMID: 33228680 Free PMC article.
-
Laboratory Diagnosis of Malaria: Comparison of Manual and Automated Diagnostic Tests.Can J Infect Dis Med Microbiol. 2017;2017:9286392. doi: 10.1155/2017/9286392. Epub 2017 Apr 5. Can J Infect Dis Med Microbiol. 2017. PMID: 28479922 Free PMC article.
References
-
- Hänscheid T, Pinto BG, Cristino JM, Grobusch MP. Malaria diagnosis with the haematology analyser Cell-Dyn 3500: what does the instrument detect? Clin Lab Haematol. 2000;22:259–261. - PubMed
-
- Scott CS, van Zyl D, Ho E, Meyersfeld D, Ruivo L, Mendelow BV, Coetzer TL. Automated detection of malaria-associated intraleucocytic haemozoin by Cell-Dyn CD4000 depolarization analysis. Clin Lab Haematol. 2003;25:77–86. - PubMed
-
- Hänscheid T, Melo-Cristino J, Pinto BG. Automated detection of malaria pigment in white blood cells for the diagnosis of malaria in Portugal. Am J Trop Med Hyg. 2001;64:290–292. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical