. 2010 Mar;82(3):412-4.
doi: 10.4269/ajtmh.2010.09-0560.

Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area

Affiliations

Automated detection of malaria-associated pseudoeosinophilia and abnormal WBC scattergram by the Sysmex XE-2100 hematology analyzer: a clinical study with 1,801 patients and real-time quantitative PCR analysis in vivax malaria-endemic area

Jong-Ha Yoo et al. Am J Trop Med Hyg. 2010 Mar.

Abstract

Recently, the XE-2100 hematology analyzer was investigated in a rather small patient group; pseudoeosinophilia or abnormal white blood cell (WBC) scattergrams reported by this instrument were considered as significantly valuable diagnostic parameters in detecting vivax malaria. This study was conducted not only to assess the usefulness of pseudoeosinophilia or abnormal WBC scattergram in vivax malaria-endemic areas with large patient groups (N = 1,801) but also to investigate the correlation of parasitemia and platelet count with pseudoeosinophilia and abnormal WBC scattergrams. Of the 1,801 analyzed patients, 413 (22.9%) were found to have malaria by Wright-Giemsa stained blood smears. Overall, either pseudoeosinophilia or abnormal WBC scattergram was detected in 191 of 413 malaria patients and 4 of 1,388 patients without malaria (sensitivity = 46.2%, specificity = 99.7%). We suggest that clinical hematology laboratories using the XE-2100 analyzer should be aware of such specific parameters, even with the absence of a clinical request.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Correlation graph of parasitemia (y) against pseudoeosinophilia as measured by the Sysmex XE-2100 analyzer (x). Regression line was y = 185.4x + 1368 (r = 0.270; P < 0.001; N = 161).

Similar articles

Cited by

References

    1. Hänscheid T, Pinto BG, Cristino JM, Grobusch MP. Malaria diagnosis with the haematology analyser Cell-Dyn 3500: what does the instrument detect? Clin Lab Haematol. 2000;22:259–261. - PubMed
    1. Scott CS, van Zyl D, Ho E, Meyersfeld D, Ruivo L, Mendelow BV, Coetzer TL. Automated detection of malaria-associated intraleucocytic haemozoin by Cell-Dyn CD4000 depolarization analysis. Clin Lab Haematol. 2003;25:77–86. - PubMed
    1. Wever PC, Henskens YM, Kager PA, Dankert J, van Gool T. Detection of imported malaria with the Cell-Dyn 4000 hematology analyzer. J Clin Microbiol. 2002;40:4729–4731. - PMC - PubMed
    1. Hänscheid T, Melo-Cristino J, Pinto BG. Automated detection of malaria pigment in white blood cells for the diagnosis of malaria in Portugal. Am J Trop Med Hyg. 2001;64:290–292. - PubMed
    1. Hänscheid T, Pinto BG, Pereira I, Cristino JM, Valadas E. Avoiding misdiagnosis of malaria: a novel automated method allows specific diagnosis, even in the absence of clinical suspicion. Emerg Infect Dis. 1999;5:836–838. - PMC - PubMed