Factors associated with cardiac conduction disorders and permanent pacemaker implantation after percutaneous aortic valve implantation with the CoreValve prosthesis
- PMID: 20211315
- DOI: 10.1016/j.ahj.2009.12.009
Factors associated with cardiac conduction disorders and permanent pacemaker implantation after percutaneous aortic valve implantation with the CoreValve prosthesis
Abstract
Background: Cardiac conduction disorders and requirement for permanent pacemaker implantation (PPI) are not uncommon after surgical aortic valve replacement and have important clinical implications. We aimed to investigate the incidence of cardiac conduction disorders after percutaneous aortic valve implantation (PAVI) and to identify possible clinical factors associated with their development.
Methods: We studied 34 patients (mean age 80 +/- 8 years, 18 male) who underwent PAVI with the CoreValve bioprosthesis (Corevalve Inc, Irvine, CA). Electrocardiographic evaluation was performed pre- and postprocedurally, and at 1-week and 1-month follow-up. Other clinical variables were obtained from the medical history, echocardiography, and angiography.
Results: After PAVI, 7 patients required PPI, all of whom developed total atrioventricular block within 3 days postprocedurally. A smaller left ventricular outflow tract diameter (20.3 +/- 0.5 vs 21.6 +/- 1.8 cm, P = .01), more left-sided heart axis (-20 degrees +/- 29 degrees vs 19 degrees +/- 36 degrees , P = .02), more mitral annular calcification (10 +/- 1 vs 5 +/- 4 mm, P = .008), and a smaller postimplantation indexed effective orifice area (0.86 +/- 0.20 vs 1.10 +/- 0.26 cm(2)/m(2), P = .04) were associated with PPI. The incidence of new left bundle-branch block (LBBB) was 65% and was associated with a deeper implantation of the prosthesis: 10.2 +/- 2.3 mm in the new-LBBB group versus 7.7 +/- 3.1 mm in the non-LBBB group (P = .02).
Conclusions: Percutaneous aortic valve implantation with the CoreValve prosthesis results in a high incidence of total atrioventricular block requiring PPI and new-onset LBBB. Preexisting disturbance of cardiac conduction, a narrow left ventricular outflow tract, and the severity of mitral annular calcification predict the need for permanent pacing, whereas the only factor shown to be predictive for new-onset LBBB is the depth of prosthesis implantation.
Similar articles
-
Conduction disturbances and permanent cardiac pacing after transcatheter implantation of the CoreValve aortic bioprosthesis: initial single centre experience.Kardiol Pol. 2012;70(2):121-8. Kardiol Pol. 2012. PMID: 22427074
-
Predictors for permanent pacemaker requirement after transcatheter aortic valve implantation with the CoreValve bioprosthesis.Am Heart J. 2009 May;157(5):860-6. doi: 10.1016/j.ahj.2009.02.016. Am Heart J. 2009. PMID: 19376312
-
Incidence and predictors of permanent pacemaker requirement after transcatheter aortic valve implantation with a self-expanding bioprosthesis.Pacing Clin Electrophysiol. 2010 Nov;33(11):1364-72. doi: 10.1111/j.1540-8159.2010.02870.x. Epub 2010 Aug 17. Pacing Clin Electrophysiol. 2010. PMID: 20723083
-
[Transcatheter aortic valve implantation and conduction disturbances].Ann Cardiol Angeiol (Paris). 2019 Dec;68(6):443-449. doi: 10.1016/j.ancard.2019.09.024. Epub 2019 Oct 23. Ann Cardiol Angeiol (Paris). 2019. PMID: 31668339 Review. French.
-
Perioperative conduction disturbances after transcatheter aortic valve replacement.J Cardiothorac Vasc Anesth. 2013 Dec;27(6):1414-20. doi: 10.1053/j.jvca.2013.08.001. Epub 2013 Oct 5. J Cardiothorac Vasc Anesth. 2013. PMID: 24103715 Review.
Cited by
-
The next revolution: percutaneous aortic valve replacement.Rambam Maimonides Med J. 2010 Oct 31;1(2):e0016. doi: 10.5041/RMMJ.10016. Print 2010 Oct. Rambam Maimonides Med J. 2010. PMID: 23908788 Free PMC article.
-
Mitral Annulus Calcification and Cardiac Conduction Disturbances: A DANCAVAS Sub-study.J Cardiovasc Imaging. 2022 Jan;30(1):62-75. doi: 10.4250/jcvi.2021.0110. J Cardiovasc Imaging. 2022. PMID: 35086172 Free PMC article.
-
Transcatheter aortic valve implantation: current and future approaches.Nat Rev Cardiol. 2011 Nov 15;9(1):15-29. doi: 10.1038/nrcardio.2011.164. Nat Rev Cardiol. 2011. PMID: 22083020 Review.
-
Monitored anesthesia care with dexmedetomidine in transfemoral percutaneous trans-catheter aortic valve implantation: two cases report.Korean J Anesthesiol. 2014 Apr;66(4):317-21. doi: 10.4097/kjae.2014.66.4.317. Epub 2014 Apr 28. Korean J Anesthesiol. 2014. PMID: 24851170 Free PMC article.
-
Occurrence and Persistency of Conduction Disturbances during Transcatheter Aortic Valve Implantation.Medicina (Kaunas). 2021 Jul 7;57(7):695. doi: 10.3390/medicina57070695. Medicina (Kaunas). 2021. PMID: 34356976 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical