Aims: In young patients with or without a congenital heart disease, transvenous leads for pacemakers or implantable cardioverter defibrillators can cause later vascular obstruction or infection. Removal of non-functional leads is controversial as it bears the risk of vascular disrupture and embolizations. We report the data of a single centre for paediatric cardiology on efficiency and safety of transvenous lead removal.
Methods and results: Between May 2005 and August 2009 in 22 patients with a mean age of 12.9 years (range: 3.6-29.5 years) removal of 28 transvenous leads (mean lead age: 5.1 years) was attempted. The main indications for removal were vascular obstruction, increased threshold, and lead dislocation. Commercially available retraction tools were used, if necessary. Twenty-five leads (89%) were retrieved with clinical success, of which 22 (79%) were removed with complete procedural success. In three leads the lead tips were retained, while three leads could not be retrieved. No major complications occurred. Additional interventions such as recanalization, balloon dilation, or stent implantation were performed as indicated. Procedure and X-ray times could be correlated to the implant age of the leads.
Conclusion: Using non-electrical techniques, transvenous lead removal can be performed with a success rate of 89% in young patients. In the case of vessel obstructions, lead replacement combined with revascularization should be performed early, as the older the lead, the more prolonged and more hazardous the extraction procedure becomes. The use of new leads and precautionary implantation techniques may facilitate later lead removal.
Similar articles
-
Effectiveness of excimer laser-assisted pacing and ICD lead extraction in children and young adults.Pacing Clin Electrophysiol. 2006 May;29(5):461-6. doi: 10.1111/j.1540-8159.2006.00376.x. Pacing Clin Electrophysiol. 2006. PMID: 16689839 Clinical Trial.
-
Single-operator experience with a mechanical approach for removal of pacing and implantable defibrillator leads.Europace. 2009 Nov;11(11):1505-9. doi: 10.1093/europace/eup222. Epub 2009 Aug 7. Europace. 2009. PMID: 19666895
-
Clinical utility of intraprocedural transesophageal echocardiography during transvenous lead extraction.J Am Soc Echocardiogr. 2008 Jul;21(7):861-7. doi: 10.1016/j.echo.2008.01.006. Epub 2008 Mar 4. J Am Soc Echocardiogr. 2008. PMID: 18313266
-
Pacing and implantable cardioverter-defibrillator transvenous lead extraction.Ital Heart J. 2005 Mar;6(3):261-6. Ital Heart J. 2005. PMID: 15875518 Review.
-
Extraction of cardiac rhythm devices: indications, techniques and outcomes for the removal of pacemaker and defibrillator leads.Int J Clin Pract. 2010 Jul;64(8):1140-7. doi: 10.1111/j.1742-1241.2010.02338.x. Int J Clin Pract. 2010. PMID: 20642712 Review.
Cited by
-
Disparities in transvenous lead extraction in young adults.Sci Rep. 2022 Jun 10;12(1):9601. doi: 10.1038/s41598-022-13769-y. Sci Rep. 2022. PMID: 35689031 Free PMC article.
-
Pacing device therapy in infants and children: a review.J Artif Organs. 2013 Mar;16(1):23-33. doi: 10.1007/s10047-012-0668-y. Epub 2012 Oct 27. J Artif Organs. 2013. PMID: 23104398 Review.
-
Performance of steroid eluting bipolar epicardial leads in pediatric and congenital heart disease patients: 15 years of single center experience.J Cardiothorac Surg. 2014 May 12;9:84. doi: 10.1186/1749-8090-9-84. J Cardiothorac Surg. 2014. PMID: 24886320 Free PMC article.
-
[Antibradycardia therapy : Indication and implementation].Herzschrittmacherther Elektrophysiol. 2016 Jun;27(2):88-94. doi: 10.1007/s00399-016-0426-6. Epub 2016 May 24. Herzschrittmacherther Elektrophysiol. 2016. PMID: 27221084 Review. German.
-
Transvenous endocardial pacing with SelectSecure™ 3830 lead in pediatric patients: case series of two infants and a literature review.BMC Cardiovasc Disord. 2024 Mar 5;24(1):145. doi: 10.1186/s12872-024-03820-3. BMC Cardiovasc Disord. 2024. PMID: 38443792 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical