A comparison of response time to desaturation between tracheal oximetry and peripheral oximetry
- PMID: 20224879
- DOI: 10.1007/s10877-010-9227-3
A comparison of response time to desaturation between tracheal oximetry and peripheral oximetry
Abstract
Objective: Trachea is an alternative site for pulse oxygen saturation monitoring. The response time of the oximetry probe has been reported more rapid when placed in central than in peripheral. The purpose of this study is to compare the performance of the oximetry probes placed in trachea and peripheral site during rapid desaturation.
Methods: Endotracheal tubes with an oximetry sensor were intubated in ten anesthesia dogs. Both the oxygen saturation signals from trachea (S(t)O(2)) and tail (S(p)O(2)) were shown on the same monitoring screen. The mechanical ventilation was disconnected to produce a rapid desaturation when both S(t)O(2) and S(p)O(2) were 100%, and the mechanical ventilation was reconnected when S(p)O(2) decreased to 80%. The time of S(t)O(2) and S(p)O(2) dropped to 95, 90, 85, and 80% was recorded, respectively during the mechanical ventilation disconnection, and the arterial blood was sampled for arterial oxygen saturation (S(a)O(2)) measurement simultaneously. The levels of measurement agreement between two oximetry readings (S(p)O(2), S(t)O(2)) and S(a)O(2) were analyzed, respectively with the Bland and Altman method.
Results: The mean response time of S(t)O(2) was significantly shorter than S(p)O(2) when both of them decreased from 100 to 80% (172.6+/-68.9 vs 220.7+/-72.3 s) during rapid desaturation. The 95% confidence interval for absolute difference between S(p)O(2) and S(a)O(2) was 4.12+/-6.47%, and between S(t)O(2) and S(a)O(2) was 3.33-3.46%.
Conclusions: Oxymetry placed in trachea provides a better monitoring for detecting rapid desaturation than in peripheral.
Similar articles
-
A pilot study of continuous transtracheal mixed venous oxygen saturation monitoring.Anesth Analg. 2005 Aug;101(2):440-443. doi: 10.1213/01.ANE.0000156949.91614.E9. Anesth Analg. 2005. PMID: 16037159 Clinical Trial.
-
[The pulmonary arterial oxygen saturation monitoring via trachea in hypoxia dogs].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2009 Aug;26(4):752-6. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2009. PMID: 19813603 Chinese.
-
Significant Delay in the Detection of Desaturation between Finger Transmittance and Earlobe Reflectance Oximetry Probes during Fiberoptic Bronchoscopy: Analysis of 104 Cases.Lung. 2019 Feb;197(1):67-72. doi: 10.1007/s00408-018-0180-0. Epub 2018 Nov 24. Lung. 2019. PMID: 30474708
-
Retinal oximetry and systemic arterial oxygen levels.Acta Ophthalmol. 2018 Nov;96 Suppl A113:1-44. doi: 10.1111/aos.13932. Acta Ophthalmol. 2018. PMID: 30460761 Review.
-
Pulse oximetry in the oesophagus.Physiol Meas. 2006 Jan;27(1):R1-35. doi: 10.1088/0967-3334/27/1/R01. Epub 2005 Nov 29. Physiol Meas. 2006. PMID: 16365505 Review.
Cited by
-
Can the descending aortic stroke volume be estimated by transesophageal descending aortic photoplethysmography?J Anesth. 2017 Jun;31(3):337-344. doi: 10.1007/s00540-017-2338-y. Epub 2017 Mar 27. J Anesth. 2017. PMID: 28349203
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources