Accuracy and reproducibility of left ventricular ejection fraction measurements using an ambulatory radionuclide left ventricular function monitor
- PMID: 2022984
Accuracy and reproducibility of left ventricular ejection fraction measurements using an ambulatory radionuclide left ventricular function monitor
Abstract
The accuracy and reproducibility of a new ambulatory radionuclide detector system (the VEST) for ejection fraction measurement has not been fully validated. Thirty-six subjects, (19 volunteers and 17 patients) underwent repetitive bicycle exercise using sequences of both VEST monitoring and gamma camera imaging. A high intraclass correlation was noted for both absolute ejection fraction [0.84 (0.56, 0.95)] and delta ejection fraction [0.87 (0.63, 0.96)] during repeat VEST monitoring. The intraclass correlation for ejection fraction was comparable for data averaged over 30 sec versus 2 min. These correlations compared favorably to those obtained for assessment of absolute and delta ejection fraction as derived by gamma camera determination by the same computer operator (intraobserver variability), two different computer operators (interobserver variability), and during repeat exercise using gamma camera imaging. In concordance, correlations between VEST and gamma camera measurements were relatively high for both absolute ejection fraction (0.78 [0.61, 0.88]) and delta ejection fraction (0.63 [0.39, 0.79]). Thus, the VEST represents a reproducible means of measuring ejection fraction change during dynamic physical activity. Its accuracy in ejection fraction measurements is similar to gamma camera imaging during exercise testing.
Similar articles
-
Regional left ventricular ejection fraction with single crystal gammacamera and gated equilibrium technique: repeatability and inter-observer variability.J Nucl Med Allied Sci. 1990 Jan-Mar;34(1):10-8. J Nucl Med Allied Sci. 1990. PMID: 2384819 Clinical Trial.
-
Ambulatory ventricular function monitoring for serial assessments of cardiac function during exercise.J Cardiol. 1987 Dec;17(4):875-85. J Cardiol. 1987. PMID: 3506612
-
[Determination of ejection fraction and left ventricular fraction using isotopic ventriculography and bidimensional echocardiography. Comparison with contrast ventriculography].Rev Esp Cardiol. 1998;51 Suppl 1:10-8. Rev Esp Cardiol. 1998. PMID: 9549395 Spanish.
-
Myocardial dysfunction in silent myocardial ischemia as demonstrated by ambulatory radionuclide left ventricular function studies.Cardiol Clin. 1992 Aug;10(3):473-8. Cardiol Clin. 1992. PMID: 1504978 Review.
-
[Continuous ambulatory measurement of left ventricular function by an isotopic method].Arch Mal Coeur Vaiss. 1995 Oct;88(10):1453-63. Arch Mal Coeur Vaiss. 1995. PMID: 8745618 Review. French.
Cited by
-
Ambulatory monitoring of left ventricular function in patients with Parkinson's disease and postural hypotension.Eur J Nucl Med. 1994 Dec;21(12):1312-7. doi: 10.1007/BF02426695. Eur J Nucl Med. 1994. PMID: 7875169
-
Nonperfusion applications in nuclear cardiology: report of a task force of the American Society of Nuclear Cardiology.J Nucl Cardiol. 1998 Mar-Apr;5(2):218-31. doi: 10.1016/s1071-3581(98)90206-9. J Nucl Cardiol. 1998. PMID: 9588675 No abstract available.
-
Assessment of the optimal atrioventricular delay in dual chamber-paced patients by a portable scintillation probe (VEST).J Nucl Cardiol. 1996 Jul-Aug;3(4):321-6. doi: 10.1016/s1071-3581(96)90092-6. J Nucl Cardiol. 1996. PMID: 8799251
-
Pathophysiology and time course of silent myocardial ischaemia during mental stress: clinical, anatomical, and physiological correlates.Br Heart J. 1995 Mar;73(3):242-9. doi: 10.1136/hrt.73.3.242. Br Heart J. 1995. PMID: 7727184 Free PMC article.
-
Left ventricular end-diastolic volume is decreased at maximal exercise in athletes with marked repolarisation abnormalities: a continuous radionuclide monitoring study.Eur J Nucl Med Mol Imaging. 2005 Feb;32(2):203-10. doi: 10.1007/s00259-004-1652-4. Epub 2004 Sep 16. Eur J Nucl Med Mol Imaging. 2005. PMID: 15378284 Clinical Trial.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources