Slow gantry rotation acquisition technique for on-board four-dimensional digital tomosynthesis
- PMID: 20229901
- DOI: 10.1118/1.3285291
Slow gantry rotation acquisition technique for on-board four-dimensional digital tomosynthesis
Abstract
Purpose: Four-dimensional cone-beam computed tomography (4D CBCT) has been investigated for motion imaging in the radiotherapy treatment room. The drawbacks of 4D CBCT are long scan times and high imaging doses. The aims of this study were to develop and investigate a slow gantry rotation acquisition protocol for four-dimensional digital tomosynthesis (4D DTS) as a faster, lower dose alternative to 4D CBCT.
Methods: This technique was implemented using an On-Board Imager kV imaging system (Varian Medical Systems, Palo Alto, CA) mounted on the gantry of a linear accelerator. The general procedure for 4D DTS imaging using slow gantry rotation acquisition consists of the following steps: (1) acquire projections over a limited gantry rotation angle in a single motion with constant frame rate and gantry rotation speed; (2) generate a respiratory signal and temporally match projection images with appropriate points from the respiratory signal; (3) use the respiratory signal to assign phases to each of the projection images; (4) sort projection images into phase bins; and (5) reconstruct phase images. Phantom studies were conducted to validate theoretically derived relationships between acquisition and respiratory parameters. Optimization of acquisition parameters was then conducted by simulating lung scans using patient data. Lung tumors with approximate volumes ranging from 0.12 to 1.53 cm3 were studied.
Results: A protocol for slow gantry rotation 4D DTS was presented. Equations were derived to express relationships between acquisition parameters (frame rate, phase window, and angular intervals between projections), respiratory cycle durations, and resulting acquisition times and numbers of projections. The phantom studies validated the relationships, and the patient studies resulted in determinations of appropriate acquisition parameters. The phase window must be set according to clinical goals. For 10% phase windows, we found that appropriate frame rates ranging from 2 to 5 frames/s, gantry rotation speeds ranging from 0.44 to 1.03 degrees/s, and aiming for an approximate maximum angular interval of 3.4 degrees between projections in phase bins were appropriate for dose, scan time, and tumor visibility optimization. Adequate tumor visibility was achieved for coronal 4D DTS images of all three lung tumors with acquisition times ranging from 0.45 to 2.12 min. vs. 1.84 to 4.24 min for 4D CBCT. 4D DTS imaging doses ranged from 0.12 to 0.72 times the dose of a standard CBCT scan vs 0.48 to 1.44 times the dose of a standard CBCT scan for 4D CBCT.
Conclusions: A slow gantry rotation acquisition technique for 4D DTS was developed and investigated. Study results indicated that 4D DTS is a feasible technique for imaging lung tumor motion in the treatment room and requires shorter acquisition times and less imaging dose than 4D CBCT for larger tumors that do not require large scan angles for sagittal views and for situations where only coronal views are needed to meet clinical needs.
Similar articles
-
Image acquisition optimization of a limited-angle intrafraction verification (LIVE) system for lung radiotherapy.Med Phys. 2018 Jan;45(1):340-351. doi: 10.1002/mp.12647. Epub 2017 Nov 30. Med Phys. 2018. PMID: 29091287 Free PMC article.
-
On-board four-dimensional digital tomosynthesis: first experimental results.Med Phys. 2008 Aug;35(8):3574-83. doi: 10.1118/1.2953561. Med Phys. 2008. PMID: 18777918
-
Four dimensional digital tomosynthesis using on-board imager for the verification of respiratory motion.PLoS One. 2014 Dec 26;9(12):e115795. doi: 10.1371/journal.pone.0115795. eCollection 2014. PLoS One. 2014. PMID: 25541710 Free PMC article.
-
Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography.J Cancer Res Ther. 2019 Jul-Sep;15(3):470-474. doi: 10.4103/jcrt.JCRT_949_17. J Cancer Res Ther. 2019. PMID: 31169206 Review.
-
[Computed tomography of the lungs. A step into the fourth dimension].Radiologe. 2009 Aug;49(8):698-704. doi: 10.1007/s00117-009-1879-y. Radiologe. 2009. PMID: 19693618 Review. German.
Cited by
-
Effects of irregular respiratory motion on the positioning accuracy of moving target with free breathing cone-beam computerized tomography.Int J Med Phys Clin Eng Radiat Oncol. 2018 May;7(2):173-183. doi: 10.4236/ijmpcero.2018.72015. Epub 2018 May 8. Int J Med Phys Clin Eng Radiat Oncol. 2018. PMID: 29951344 Free PMC article.
-
Estimating 4D-CBCT from prior information and extremely limited angle projections using structural PCA and weighted free-form deformation for lung radiotherapy.Med Phys. 2017 Mar;44(3):1089-1104. doi: 10.1002/mp.12102. Med Phys. 2017. PMID: 28079267 Free PMC article.
-
Technical Note: Imaging dose resulting from optimized procedures with limited-angle intrafractional verification system during stereotactic body radiation therapy lung treatment.Med Phys. 2019 Jun;46(6):2709-2715. doi: 10.1002/mp.13511. Epub 2019 Apr 29. Med Phys. 2019. PMID: 30937910 Free PMC article.
-
Image acquisition optimization of a limited-angle intrafraction verification (LIVE) system for lung radiotherapy.Med Phys. 2018 Jan;45(1):340-351. doi: 10.1002/mp.12647. Epub 2017 Nov 30. Med Phys. 2018. PMID: 29091287 Free PMC article.
-
Preliminary clinical evaluation of a 4D-CBCT estimation technique using prior information and limited-angle projections.Radiother Oncol. 2015 Apr;115(1):22-9. doi: 10.1016/j.radonc.2015.02.022. Epub 2015 Mar 26. Radiother Oncol. 2015. PMID: 25818396 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources