Orbit-related variation in spatial resolution as a source of artifactual defects in thallium-201 SPECT
- PMID: 2022998
Orbit-related variation in spatial resolution as a source of artifactual defects in thallium-201 SPECT
Abstract
The cause of 180-degree diametrical artifactual defects in clinical thallium-201 SPECT imaging was investigated using phantom simulation. This artifact was observed on SPECT images acquired with a "body contour" or "peanut" orbit. It was hypothesized that this artifact was caused by differences in spatial resolution that occur when the heart-to-detector distance changes employing noncircular orbits. To test this hypothesis, a series of planar static images of a normal cylindrical phantom was obtained at varying distances from the camera detector head. From these images, tomographic acquisition files were created that simulated tomographic data acquired with circular orbits and elliptical orbits. The reconstructed phantom short-axis slices showed no artifacts for circular orbits. However, for various elliptical orbits, significant regional nonuniformity, similar to the artifacts noted in patients, was observed. The degree of nonuniformity correlated with the long-short axis ratio of elliptical orbits (r = 0.98). In addition, circular orbits with the phantom in an eccentric position resulted in similar nonuniformities. It is concluded that a noncircular tomographic orbit can create characteristic artifacts on thallium-201 SPECT images. For rotational thallium 201 SPECT, a circular orbit with the heart in the center of rotation should be employed.
Comment in
-
SPECT and artifacts--in search of the imaginary lesion.J Nucl Med. 1991 May;32(5):875-7. J Nucl Med. 1991. PMID: 2022999 No abstract available.
Similar articles
-
Differential effect of 180 degrees and 360 degrees acquisition orbits on the accuracy of SPECT imaging: quantitative evaluation in phantoms.J Nucl Med. 2002 Aug;43(8):1115-24. J Nucl Med. 2002. PMID: 12163639
-
Body-contour versus circular orbit acquisition in cardiac SPECT: assessment of defect detectability with channelized Hotelling observer.Nucl Med Commun. 2007 Dec;28(12):937-42. doi: 10.1097/MNM.0b013e3282f1b9a3. Nucl Med Commun. 2007. PMID: 18090221
-
Effect of tomographic orbit and type of rotation on apparent myocardial activity.Nucl Med Commun. 2005 Jan;26(1):25-30. doi: 10.1097/00006231-200501000-00005. Nucl Med Commun. 2005. PMID: 15604944
-
Artifacts in planar and SPECT myocardial perfusion imaging.Am J Card Imaging. 1992 Mar;6(1):42-57; discussion 58. Am J Card Imaging. 1992. PMID: 10147614 Review. No abstract available.
-
Anatomy of the Orbits: Skeletal Features and Some Notes on the Periorbital Lining.Atlas Oral Maxillofac Surg Clin North Am. 2021 Mar;29(1):1-18. doi: 10.1016/j.cxom.2020.10.001. Atlas Oral Maxillofac Surg Clin North Am. 2021. PMID: 33516532 Review. No abstract available.
Cited by
-
Investigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study.J Biomed Phys Eng. 2015 Mar 4;5(1):39-48. eCollection 2015 Mar. J Biomed Phys Eng. 2015. PMID: 25973410 Free PMC article.
-
Quantitative myocardial perfusion SPECT.J Nucl Cardiol. 1998 Sep-Oct;5(5):507-22. doi: 10.1016/s1071-3581(98)90182-9. J Nucl Cardiol. 1998. PMID: 9796898 Review.
-
Use of the selective linogram in cardiac tomography quality control.J Nucl Cardiol. 1995 Jul-Aug;2(4):303-8. doi: 10.1016/s1071-3581(05)80074-1. J Nucl Cardiol. 1995. PMID: 9420804
-
Estimation of attenuation maps from scatter and photopeak window single photon-emission computed tomographic images of technetium 99m-labeled sestamibi.J Nucl Cardiol. 1997 Jan-Feb;4(1 Pt 1):42-51. doi: 10.1016/s1071-3581(97)90048-9. J Nucl Cardiol. 1997. PMID: 9138839
-
Myocardial perfusion and function single photon emission computed tomography.J Nucl Cardiol. 2006 Nov;13(6):e97-120. doi: 10.1016/j.nuclcard.2006.08.008. J Nucl Cardiol. 2006. PMID: 17174800 No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources