Finite element modeling of subcutaneous implantable defibrillator electrodes in an adult torso
- PMID: 20230927
- PMCID: PMC3103844
- DOI: 10.1016/j.hrthm.2010.01.030
Finite element modeling of subcutaneous implantable defibrillator electrodes in an adult torso
Abstract
Background: Total subcutaneous implantable subcutaneous defibrillators are in development, but optimal electrode configurations are not known.
Objective: We used image-based finite element models (FEM) to predict the myocardial electric field generated during defibrillation shocks (pseudo-DFT) in a wide variety of reported and innovative subcutaneous electrode positions to determine factors affecting optimal lead positions for subcutaneous implantable cardioverter-defibrillators (S-ICD).
Methods: An image-based FEM of an adult man was used to predict pseudo-DFTs across a wide range of technically feasible S-ICD electrode placements. Generator location, lead location, length, geometry and orientation, and spatial relation of electrodes to ventricular mass were systematically varied. Best electrode configurations were determined, and spatial factors contributing to low pseudo-DFTs were identified using regression and general linear models.
Results: A total of 122 single-electrode/array configurations and 28 dual-electrode configurations were simulated. Pseudo-DFTs for single-electrode orientations ranged from 0.60 to 16.0 (mean 2.65 +/- 2.48) times that predicted for the base case, an anterior-posterior configuration recently tested clinically. A total of 32 of 150 tested configurations (21%) had pseudo-DFT ratios </=1, indicating the possibility of multiple novel, efficient, and clinically relevant orientations. Favorable alignment of lead-generator vector with ventricular myocardium and increased lead length were the most important factors correlated with pseudo-DFT, accounting for 70% of the predicted variation (R(2) = 0.70, each factor P < .05) in a combined general linear model in which parameter estimates were calculated for each factor.
Conclusion: Further exploration of novel and efficient electrode configurations may be of value in the development of the S-ICD technologies and implant procedure. FEM modeling suggests that the choice of configurations that maximize shock vector alignment with the center of myocardial mass and use of longer leads is more likely to result in lower DFT.
Copyright 2010 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
Conflict of interest statement
Matthew Jolley: No conflicts to report
Jeroen Stinstra: No conflicts to report
Jess Tate No conflicts to report
Steve Pieper: No conflicts to report
Rob MacLeod: No conflicts to report
Larry Chu: No conflicts to report
Paul Wang: No conflicts to report
John K. Triedman: Honorarium, Boston Scientific CRM
Figures






Comment in
-
The infinite value in subcutaneous defibrillation.Heart Rhythm. 2010 May;7(5):699-700. doi: 10.1016/j.hrthm.2010.02.022. Epub 2010 Feb 20. Heart Rhythm. 2010. PMID: 20176140 No abstract available.
References
-
- Bokhari F, Newman D, Greene M, Korley V, Mangat I, Dorian P. Long-term comparison of the implantable cardioverter defibrillator versus amiodarone: eleven-year follow-up of a subset of patients in the Canadian Implantable Defibrillator Study (CIDS) Circulation. 2004;110:112–116. - PubMed
-
- Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. New Engl J Med. 1999;341:1882–1890. - PubMed
-
- Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. New Engl J Med. 1996;335:1933–1940. - PubMed
-
- Burke MC, Coman JA, Cates AW, et al. Defibrillation energy requirements using a left anterior chest cutaneous to subcutaneous shocking vector: implications for a total subcutaneous implantable defibrillator. Heart Rhythm. 2005;2:1332–1338. - PubMed
-
- Lieberman R, Havel WJ, Rashba E, DeGroot PJ, Stromberg K, Shorofsky SR. Acute defibrillation performance of a novel, non-transvenous shock pathway in adult ICD indicated patients. Heart Rhythm. 2008;5:28–34. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical