Valve-related complications with the Hancock I porcine bioprosthesis. A twelve- to fourteen-year follow-up study
- PMID: 2023444
Valve-related complications with the Hancock I porcine bioprosthesis. A twelve- to fourteen-year follow-up study
Abstract
Valve-related morbidity and mortality after heart valve replacement with the Hancock I porcine bioprosthesis has been retrospectively analyzed. From June 1974 through December 1976, 253 Hancock I bioprostheses (150 mitral and 103 aortic) were inserted in 220 selected patients who survived the operation and had follow-up until June 1989 (mean follow-up 13.5 years, with an accumulative follow-up of 2956.4 patient-years). One hundred seventeen patients had mitral valve replacement, 70 had aortic valve replacement, and 33 had combined mitral and aortic valve replacement. There were 27 thromboembolic events. The probability of being free from thromboembolism at 14 years was 81.0% +/- 7.4% for the mitral valve replacement group, 85.4% +/- 6.7% for the aortic group, and 67.1% +/- 18.4% for the mitral-aortic group. Fifteen episodes of prosthetic valve endocarditis occurred. There were 10 instances of nonstructural dysfunction (paravalvular leaks) in seven mitral valves (4.6%) and in three aortic valves (2.9%). One hundred twenty-two bioprostheses in 106 patients resulted in structural deterioration. The probability of freedom from structural deterioration at 14 years was 37.2% +/- 3.9% for the mitral group, 43.9% +/- 7.1% for the aortic group, and 30.1% +/- 8.9% for the mitral-aortic group. The logistic regression analysis between age at the time of operation and bioprosthetic life (structural deterioration-free period) demonstrates a linear regression curve (r = 0.53). There were 56 late deaths (27 patients died at reoperation). The actuarial survival rate (including hospital mortality) at 14 years was 57.2% +/- 5.4% for the entire series, with no statistically significant difference between groups. The probability of remaining free from valve-related morbidity and mortality at 14 years was 16.7% +/- 4.8% for the mitral group, 20.8% +/- 6.2% for the aortic group, and 14.0% +/- 7.0% for the mitral-aortic group. The long-term results of this series show that the clinical performance of the Hancock I porcine valve appears satisfactory during the first 6 years. The behavior of this bioprosthesis at 14 years' follow-up changes drastically, because only a minor group of patients is free from valve-related complications, justifying the restriction of its use for selected patients.
Similar articles
-
An eight-year experience with porcine bioprosthetic cardiac valves.J Thorac Cardiovasc Surg. 1986 Jun;91(6):910-7. J Thorac Cardiovasc Surg. 1986. PMID: 3713240
-
The Carpentier-Edwards standard porcine bioprosthesis. A first-generation tissue valve with excellent long-term clinical performance.J Thorac Cardiovasc Surg. 1990 Mar;99(3):543-61. J Thorac Cardiovasc Surg. 1990. PMID: 2308373
-
Experience with low-dose aspirin as thromboprophylaxis for the Tissuemed porcine aortic bioprosthesis: a survey of five years' experience.J Heart Valve Dis. 1998 Sep;7(5):574-9. J Heart Valve Dis. 1998. PMID: 9793859
-
[Mid-term results of heart valve replacement with 2 types of pericardial heterograft. Apropos of 208 cases].Arch Mal Coeur Vaiss. 1991 Jun;84(6):785-91. Arch Mal Coeur Vaiss. 1991. PMID: 1898212 Review. French.
-
Pregnancy and childbearing in a population with biologic valvular prostheses.J Thorac Cardiovasc Surg. 1991 Aug;102(2):179-86. J Thorac Cardiovasc Surg. 1991. PMID: 1865693 Review.
Cited by
-
Poor clinical performance of the Wessex porcine heart valve bioprosthesis at nine years' follow up.Heart. 1997 Apr;77(4):319-24. doi: 10.1136/hrt.77.4.319. Heart. 1997. PMID: 9155609 Free PMC article.
-
Three-Plug Technique for Mitral Paravalvular Leak Closure in Marfan Syndrome.Cureus. 2025 Jan 24;17(1):e77920. doi: 10.7759/cureus.77920. eCollection 2025 Jan. Cureus. 2025. PMID: 39991344 Free PMC article.