Comparative Study
. 2010 Aug;24(8):498-504.
doi: 10.1038/jhh.2010.27. Epub 2010 Mar 18.

A new oscillometric method for pulse wave analysis: comparison with a common tonometric method

Affiliations
Comparative Study

A new oscillometric method for pulse wave analysis: comparison with a common tonometric method

S Wassertheurer et al. J Hum Hypertens. 2010 Aug.

Abstract

In the European Society of Cardiology-European Society of Hypertension guidelines of the year 2007, the consequences of arterial stiffness and wave reflection on cardiovascular mortality have a major role. But the investigators claimed the poor availability of devices/methods providing easy and widely suitable measuring of arterial wall stiffness or their surrogates like augmentation index (AIx) or aortic systolic blood pressure (aSBP). The aim of this study was the validation of a novel method determining AIx and aSBP based on an oscillometric method using a common cuff (ARCSolver) against a validated tonometric system (SphygmoCor). aSBP and AIx measured with the SphygmoCor and ARCSolver method were compared for 302 subjects. The mean age was 56 years with an s.d. of 20 years. At least two iterations were performed in each session. This resulted in 749 measurements. For aSBP the mean difference was -0.1 mm Hg with an s.d. of 3.1 mm Hg. The mean difference for AIx was 1.2% with an s.d. of 7.9%. There was no significant difference in reproducibility of AIx for both methods. The variation estimate of inter- and intraobserver measurements was 6.3% for ARCSolver and 7.5% for SphygmoCor. The ARCSolver method is a novel method determining AIx and aSBP based on an oscillometric system with a cuff. The results agree with common accepted tonometric measurements. Its application is easy and for widespread use.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(a) Characteristic modulus amplifications and phase shifts of pressure wave harmonics between aortic root and brachial artery used for the ARCSolver—generalized transfer function and data published by Karamanoglu et al. (b) Principles to derive aSBP and AIx from the brachial waveform.
Figure 2
Figure 2
Bland–Altman analysis of aSBP ARCSolver vs SphygmoCor.
Figure 3
Figure 3
Bland–Altman analysis of Alx ARCSolver vs SphygmoCor.

Comment in

References

    1. MacMahon S, Peto R, Cutler J, Collins R, Sorlie P, Neaton J, et al. Blood pressure, stroke, and coronary heart disease: part 1, prolonged differences in blood pressure: prospective observational studies corrected for the regression dilution bias. Lancet. 1990;335:765–774. - PubMed
    1. Franklin SS, Lopez VA, Wong ND, Mitchell GF, Larson MG, Vasan RS, et al. Single versus combined blood pressure components and risk for cardiovascular disease. The Framingham Heart Study. Circulation. 2008;119:243–250. - PMC - PubMed
    1. Mancia G, Rosei EA, Cifkova R, DeBacker G, Erdine S, Fagard R, et al. European Society of Hypertension—European Society of Cardiology guidelines for the management of arterial hypertension. J Hypertens. 2003;21:1011–1053. - PubMed
    1. Nichols WW, O'Rourke MF.McDonald's Blood Flow in Arteries4th edn.Arnold: London; 1998
    1. Weber T, Auer J, O'Rourke MF, Kvas E, Lassnig E, Berent R, et al. Arterial stiffness, wave reflections, and the risk of coronary artery disease Circulation 2004109184–189.14. - PubMed

MeSH terms