Carboxyhaemoglobinaemia and pulse oximetry
- PMID: 2031826
- DOI: 10.1093/bja/66.5.625
Carboxyhaemoglobinaemia and pulse oximetry
Abstract
We compared measurements obtained with a pulse oximeter (SpO2) against values obtained from a CO-oximeter in a patient with carbon monoxide poisoning. SpO2 was equal to the sum of the oxyhaemoglobin (HbO) and carboxyhaemoglobin (HbCO) values over the range of HbCO from 30 to 1%. This confirms the experimental findings that pulse oximeters measure HbCO as HbO. The patient was treated with oxygen (FlO2 = 50%) and recovered without any sequelae. Under these circumstances, the half-life of HbCO was approximately 2 h.
Comment in
-
Management of carboxyhaemoglobinaemia.Br J Anaesth. 1992 Jan;68(1):113. doi: 10.1093/bja/68.1.113. Br J Anaesth. 1992. PMID: 1739549 No abstract available.
-
Management of carboxyhaemoglobinaemia.Br J Anaesth. 1992 Jan;68(1):113. doi: 10.1093/bja/68.1.113-a. Br J Anaesth. 1992. PMID: 1739550 No abstract available.
Similar articles
-
Management of carboxyhaemoglobinaemia.Br J Anaesth. 1992 Jan;68(1):113. doi: 10.1093/bja/68.1.113-a. Br J Anaesth. 1992. PMID: 1739550 No abstract available.
-
Diagnostic Performance of Carbon Monoxide Testing by Pulse Oximetry in the Emergency Department.Respir Care. 2019 Nov;64(11):1351-1357. doi: 10.4187/respcare.06365. Epub 2019 Apr 30. Respir Care. 2019. PMID: 31040204
-
Non-invasive monitoring of carboxyhemoglobin during hyperbaric oxygen therapy.Undersea Hyperb Med. 2021 First Quarter;48(1):33-42. doi: 10.22462/01.03.2021.4. Undersea Hyperb Med. 2021. PMID: 33648031
-
Carbon monoxide poisoning: forgotten not gone!Br J Hosp Med. 1988 Apr;39(4):298-300, 302, 304-5. Br J Hosp Med. 1988. PMID: 3291995 Review.
-
Carboxyhemoglobin: a primer for clinicians.Undersea Hyperb Med. 2018 Mar-Apr;45(2):165-171. Undersea Hyperb Med. 2018. PMID: 29734568 Review.
Cited by
-
Pitfalls in diagnosis and management of carbon monoxide poisoning.J Accid Emerg Med. 1996 Jan;13(1):62-3. doi: 10.1136/emj.13.1.62. J Accid Emerg Med. 1996. PMID: 8821232 Free PMC article.
-
Pulse Oximetry: The Working Principle, Signal Formation, and Applications.Adv Exp Med Biol. 2022;1384:205-218. doi: 10.1007/978-3-031-06413-5_12. Adv Exp Med Biol. 2022. PMID: 36217086
-
The effect of dyshemoglobins on pulse oximetry: Part I, Theoretical approach and Part II, Experimental results using an in vitro test system.J Clin Monit. 1993 Apr;9(2):81-90. doi: 10.1007/BF01616919. J Clin Monit. 1993. PMID: 8478653
-
Accuracy of pulse oximetry at various haematocrits and during haemolysis in an in vitro model.Med Biol Eng Comput. 1993 Mar;31(2):135-41. doi: 10.1007/BF02446671. Med Biol Eng Comput. 1993. PMID: 8331993
-
Evaluating blood oxygen saturation measurements by popular fitness trackers in postoperative patients: A prospective clinical trial.iScience. 2023 Oct 6;26(11):108155. doi: 10.1016/j.isci.2023.108155. eCollection 2023 Nov 17. iScience. 2023. PMID: 37876822 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical