Determination of reference geometry for polyethylene tibial insert wear analysis
- PMID: 20347254
- DOI: 10.1016/j.arth.2010.01.096
Determination of reference geometry for polyethylene tibial insert wear analysis
Abstract
Geometric wear analysis techniques require unworn geometries to serve as a reference in wear measurement. A method to create a reference geometrical model is described for retrieval studies when the actual unworn geometry is unavailable. Never-implanted tibial inserts were scanned with micro-computed tomography. Two, 3, or 6 insert surfaces were coaligned and averaged to create reference geometries. Individual inserts were compared with each other (manufacturing variability) and with the reference geometries (reference variability). The 3-dimensional deviations between the surfaces were recorded. The reference variability was reduced to 8.3 ± 39 μm, vs manufacturing variability of 15 ± 59 μm. Deviations were smallest on the articular surfaces where most wear occurs and were significantly less than the reported insert wear rate of 20 μm/y.
Copyright © 2011 Elsevier Inc. All rights reserved.