Determination of reference geometry for polyethylene tibial insert wear analysis
- PMID: 20347254
- DOI: 10.1016/j.arth.2010.01.096
Determination of reference geometry for polyethylene tibial insert wear analysis
Abstract
Geometric wear analysis techniques require unworn geometries to serve as a reference in wear measurement. A method to create a reference geometrical model is described for retrieval studies when the actual unworn geometry is unavailable. Never-implanted tibial inserts were scanned with micro-computed tomography. Two, 3, or 6 insert surfaces were coaligned and averaged to create reference geometries. Individual inserts were compared with each other (manufacturing variability) and with the reference geometries (reference variability). The 3-dimensional deviations between the surfaces were recorded. The reference variability was reduced to 8.3 ± 39 μm, vs manufacturing variability of 15 ± 59 μm. Deviations were smallest on the articular surfaces where most wear occurs and were significantly less than the reported insert wear rate of 20 μm/y.
Copyright © 2011 Elsevier Inc. All rights reserved.
Similar articles
-
How do CAD models compare with reverse engineered manufactured components for use in wear analysis?Clin Orthop Relat Res. 2012 Jul;470(7):1847-54. doi: 10.1007/s11999-011-2143-0. Clin Orthop Relat Res. 2012. PMID: 22016002 Free PMC article.
-
Manufacturing lot affects polyethylene tibial insert volume, thickness, and surface geometry.Proc Inst Mech Eng H. 2013 Aug;227(8):884-9. doi: 10.1177/0954411913486755. Epub 2013 Apr 26. Proc Inst Mech Eng H. 2013. PMID: 23636769
-
In vitro quantification of wear in tibial inserts using microcomputed tomography.Clin Orthop Relat Res. 2011 Jan;469(1):107-12. doi: 10.1007/s11999-010-1490-6. Clin Orthop Relat Res. 2011. PMID: 20676810 Free PMC article.
-
The role of the cemented all-polyethylene tibial component in total knee replacement: a 30-year patient follow-up and review of the literature.Knee. 2010 Dec;17(6):412-6. doi: 10.1016/j.knee.2009.11.008. Epub 2010 Jan 8. Knee. 2010. PMID: 20060725 Review.
-
Pop-top tibial component: elimination of back-side wear with potential modularity.Surg Technol Int. 2004;13:253-60. Surg Technol Int. 2004. PMID: 15744698 Review.
Cited by
-
An autonomous mathematical reconstruction to effectively measure volume loss on retrieved polyethylene tibial inserts.J Biomed Mater Res B Appl Biomater. 2013 Apr;101(3):449-57. doi: 10.1002/jbm.b.32782. Epub 2012 Aug 22. J Biomed Mater Res B Appl Biomater. 2013. PMID: 22915471 Free PMC article.
-
Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.J Biomed Mater Res B Appl Biomater. 2017 Oct;105(7):2053-2059. doi: 10.1002/jbm.b.33684. Epub 2016 Jul 12. J Biomed Mater Res B Appl Biomater. 2017. PMID: 27401236 Free PMC article.
-
The 2012 Mark Coventry award: a retrieval analysis of high flexion versus posterior-stabilized tibial inserts.Clin Orthop Relat Res. 2013 Jan;471(1):56-63. doi: 10.1007/s11999-012-2387-3. Clin Orthop Relat Res. 2013. PMID: 22661024 Free PMC article.
-
How do CAD models compare with reverse engineered manufactured components for use in wear analysis?Clin Orthop Relat Res. 2012 Jul;470(7):1847-54. doi: 10.1007/s11999-011-2143-0. Clin Orthop Relat Res. 2012. PMID: 22016002 Free PMC article.
-
Inter and intra-system size variability of reverse shoulder arthroplasty polyethylene inserts.Int J Shoulder Surg. 2016 Jan-Mar;10(1):10-4. doi: 10.4103/0973-6042.174512. Int J Shoulder Surg. 2016. PMID: 26980984 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical