Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography
- PMID: 20352443
- DOI: 10.1007/s00417-010-1360-7
Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography
Abstract
Background: To assess the intra-examiner repeatability and agreement of corneal pachymetry maps measured by time-domain (Visante OCT) and Fourier-domain optical coherence tomography (RTVue OCT).
Methods: This observational cross-sectional study enrolled 72 eyes from 72 volunteers. Only one eye of each patient was chosen randomly to receive repeated scanning with both devices by the same examiner, in order to test the intra-observer repeatability. The first scan by each device from all enrolled eyes was used to analyze the difference and agreement between the two devices. The agreement between the two devices was analyzed by the Bland-Altman method. Intra-observer repeatability of each OCT device was analyzed by intra-class correlation (ICC).
Results: The mean corneal thickness of the central 2 mm zone was 524.3 ± 35.7 µm and 525.4 ± 35.3 µm by the Visante OCT and the RTVue OCT respectively (p = 0.089). The intra-observer repeatability of the RTVue OCT (ICC = 0.994) was superior to that of the Visante OCT (ICC = 0.989) in the central 2 mm zone. However, the intra-observer repeatability of the RTVue OCT in the pericentral 2 to 5 mm zone was not superior to that of the Visante OCT. Both OCT devices had similar repeatability in the pericentral 2 to 5 mm zone [ICC of the Visante OCT = 0.991; 95% confidence interval (CI) 0.986-0.995), ICC of the RTVue OCT = 0.991; 95% CI 0.985-0.994)).
Conclusions: The difference in CCT measurement by the Visante OCT and the RTVue OCT is probably too small to influence clinical decision making for refractive surgery and glaucoma management. The RTVue OCT demonstrated better intra-observer repeatability in the central 2 mm zone, which probably was related to its rapid image acquisition capability.
Similar articles
-
Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography.Ophthalmology. 2008 May;115(5):796-801.e2. doi: 10.1016/j.ophtha.2007.07.006. Epub 2007 Oct 4. Ophthalmology. 2008. PMID: 17916376
-
Intrasession repeatability of pachymetry measurements with RTVue XR 100 optical coherence tomography in normal cornea.Saudi J Ophthalmol. 2017 Apr-Jun;31(2):65-68. doi: 10.1016/j.sjopt.2017.04.003. Epub 2017 Apr 9. Saudi J Ophthalmol. 2017. PMID: 28559715 Free PMC article.
-
The role of axial resolution of optical coherence tomography on the measurement of corneal and epithelial thicknesses.Invest Ophthalmol Vis Sci. 2013 Jan 28;54(1):746-55. doi: 10.1167/iovs.11-9308. Invest Ophthalmol Vis Sci. 2013. PMID: 23139281
-
Optical Coherence Tomography Features of Ocular Cysticercosis: A Review of Literature With Observer Variation.Ophthalmic Surg Lasers Imaging Retina. 2022 Aug;53(8):446-454. doi: 10.3928/23258160-20220629-02. Epub 2022 Aug 1. Ophthalmic Surg Lasers Imaging Retina. 2022. PMID: 35951713 Review.
-
Fourier-domain optical coherence tomography: recent advances toward clinical utility.Curr Opin Biotechnol. 2009 Feb;20(1):111-8. doi: 10.1016/j.copbio.2009.02.007. Epub 2009 Mar 4. Curr Opin Biotechnol. 2009. PMID: 19264475 Free PMC article. Review.
Cited by
-
Repeatability of corneal epithelial thickness measurements using Fourier-domain optical coherence tomography in normal and post-LASIK eyes.Cornea. 2013 Dec;32(12):1544-8. doi: 10.1097/ICO.0b013e3182a7f39d. Cornea. 2013. PMID: 24145634 Free PMC article.
-
Comparison of central corneal thickness measurements with three new optical devices and a standard ultrasonic pachymeter.Int J Ophthalmol. 2014 Apr 18;7(2):302-8. doi: 10.3980/j.issn.2222-3959.2014.02.19. eCollection 2014. Int J Ophthalmol. 2014. PMID: 24790874 Free PMC article.
-
Evaluation of rigid gas permeable lens fitting in keratoconic patients with optical coherence tomography.Graefes Arch Clin Exp Ophthalmol. 2013 Jun;251(6):1565-70. doi: 10.1007/s00417-013-2271-1. Epub 2013 Feb 8. Graefes Arch Clin Exp Ophthalmol. 2013. PMID: 23392819
-
Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.Graefes Arch Clin Exp Ophthalmol. 2018 Jan;256(1):113-123. doi: 10.1007/s00417-017-3832-5. Epub 2017 Nov 2. Graefes Arch Clin Exp Ophthalmol. 2018. PMID: 29098383
-
Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Measurement of Thinnest Corneal Thickness in Keratoconus.J Ophthalmol. 2015;2015:402925. doi: 10.1155/2015/402925. Epub 2015 Jun 25. J Ophthalmol. 2015. PMID: 26199737 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical