Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions
- PMID: 20353963
- DOI: 10.1093/europace/euq092
Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions
Abstract
Aims: The purpose of the current study is to evaluate the safety profile of patients with pacemakers or implantable cardioverter-defibrillators (ICDs) undergoing a medically necessary magnetic resonance imaging (MRI) scan without limitation on peak specific absorption rate (SAR). Recent series suggest that MRI scanning can be performed safely in select patients with pacemakers or ICDs. These studies, though, limited peak SAR.
Methods and results: One-hundred and three patients with a total of 240 leads underwent a total of 127 scans of any body landmark using usual protocols with standard peak SAR settings for the scan. No patient was pacemaker dependent. Thresholds were obtained immediately before and after the scan. For all scans, the median (25th and 75th percentiles) peak SAR was 2.5 (1.3, 3.2) W/kg whereas the median scan time was 1650 (1236, 2099) s. Pre- and post-scan pacing thresholds were unchanged [0.7 (0.5, 0.8) vs. 0.6 (0.5, 0.8) V at 0.5 ms, P=NS], though the sensed amplitudes [6.7 (2.9, 11.5) vs. 6.1 (2.9, 11.2) mV, P<0.0001] and pacing impedances [500 (440, 609) vs. 491 (437, 593) Omega, P<0.0001] both decreased significantly.
Conclusion: The current series suggests that MRI scans may be performed safely in appropriately selected patients up to a peak SAR of 3.2 W/kg. Furthermore, peak SAR level poorly predicts the safety profile of patients with pacemakers or ICDS who are exposed to an MRI environment.
Comment in
-
The safety of MRI scanning of pacemakers and ICDs: what are the critical elements of safe scanning? Ask me again at 10,000.Europace. 2010 Jul;12(7):915-7. doi: 10.1093/europace/euq174. Epub 2010 Jun 12. Europace. 2010. PMID: 20543196 No abstract available.
-
Clarification to the article 'Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions'.Europace. 2010 Dec;12(12):1798. doi: 10.1093/europace/euq308. Epub 2010 Aug 14. Europace. 2010. PMID: 20710022 No abstract available.
Similar articles
-
The safety of MRI scanning of pacemakers and ICDs: what are the critical elements of safe scanning? Ask me again at 10,000.Europace. 2010 Jul;12(7):915-7. doi: 10.1093/europace/euq174. Epub 2010 Jun 12. Europace. 2010. PMID: 20543196 No abstract available.
-
Cardiac biomarkers in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan.Pacing Clin Electrophysiol. 2008 Oct;31(10):1241-5. doi: 10.1111/j.1540-8159.2008.01172.x. Pacing Clin Electrophysiol. 2008. PMID: 18811802 Clinical Trial.
-
Ectopy in patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing an MRI scan.Pacing Clin Electrophysiol. 2009 Jun;32(6):772-8. doi: 10.1111/j.1540-8159.2009.02364.x. Pacing Clin Electrophysiol. 2009. PMID: 19545340 Clinical Trial.
-
Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices.Europace. 2008 Mar;10(3):336-46. doi: 10.1093/europace/eun021. Europace. 2008. PMID: 18308754 Review.
-
Implications of pacemakers and implantable cardioverter defibrillators in urological practice.J Urol. 2011 Oct;186(4):1198-205. doi: 10.1016/j.juro.2011.02.2697. Epub 2011 Aug 16. J Urol. 2011. PMID: 21849189 Review.
Cited by
-
Safety of implanted cardiac devices in an MRI environment.Curr Cardiol Rep. 2015 Jul;17(7):605. doi: 10.1007/s11886-015-0605-9. Curr Cardiol Rep. 2015. PMID: 26026996 Review.
-
A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.Ann Intern Med. 2011 Oct 4;155(7):415-24. doi: 10.7326/0003-4819-155-7-201110040-00004. Ann Intern Med. 2011. PMID: 21969340 Free PMC article. Clinical Trial.
-
Cardiac Implantable Electronic Device Safety during Magnetic Resonance Imaging.Korean Circ J. 2016 Nov;46(6):804-810. doi: 10.4070/kcj.2016.46.6.804. Epub 2016 Nov 1. Korean Circ J. 2016. PMID: 27826339 Free PMC article.
-
Magnetic resonance imaging in patients with cardiac implantable electronic devices: a single-center prospective study.J Interv Card Electrophysiol. 2017 Oct;50(1):95-104. doi: 10.1007/s10840-017-0262-6. Epub 2017 Jul 22. J Interv Card Electrophysiol. 2017. PMID: 28733766
-
Magnetic resonance imaging and devices: a mesmerising combination.Neth Heart J. 2014 Jun;22(6):267-8. doi: 10.1007/s12471-014-0562-8. Neth Heart J. 2014. PMID: 24807837 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous