Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel
- PMID: 20357149
- DOI: 10.1213/ANE.0b013e3181d2aad7
Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel
Abstract
Background: The Airway Scope (AWS) (Pentax-AWS, Hoya Corp., Tokyo, Japan) and the Airtraq (ATQ) (Prodol, Vizcaya, Spain) have similarities in the novel structures of their blades. In this study, we evaluated the ease of use of the AWS and ATQ compared with the Macintosh laryngoscope (ML) by inexperienced personnel in a simulated manikin difficult airway.
Methods: Twenty-four fifth-year medical students with no previous experience in tracheal intubation participated in this study. We used an advanced patient simulator (SimMan(R), Laerdal Medical, Stavanger, Norway) to simulate difficult airway scenarios including cervical spine rigidity, limited mouth opening, and pharyngeal obstruction. The sequences in selecting devices and scenarios were randomized. Success rates for tracheal intubation, and the time required for visualization of the glottis, tracheal intubation, and inflation of the lungs, and the number of optimization maneuvers and dental click sounds were analyzed. The 3 different intubation devices were tested in 4 different scenarios by 24 students.
Results: Both the AWS and ATQ had very high success rates of tracheal intubation compared with the ML (AWS 100%*; ATQ 98%*; and ML 89%; *P < 0.05 AWS, ATQ versus ML). The time to intubation with the AWS was significantly shorter than with the ATQ and ML (AWS 11 +/- 6 seconds; ATQ 16 +/- 12 seconds; and ML 16 +/- 11 seconds; *P < 0.05 AWS versus ATQ, ML). The number of optimization maneuvers with the AWS was significantly lower than with the ATQ and ML. There were significantly more audible dental click sounds with the ML than with the AWS and ATQ.
Conclusion: Both the AWS and ATQ may be suitable devices for difficult intubation by inexperienced personnel in this manikin simulated scenario. Further studies in a clinical setting are necessary to confirm these findings.
Similar articles
-
Tracheal intubation by inexperienced medical residents using the Airtraq and Macintosh laryngoscopes--a manikin study.Am J Emerg Med. 2006 Nov;24(7):769-74. doi: 10.1016/j.ajem.2006.03.014. Am J Emerg Med. 2006. PMID: 17098095
-
A comparison of the Glidescope, Pentax AWS, and Macintosh laryngoscopes when used by novice personnel: a manikin study.Can J Anaesth. 2009 Nov;56(11):802-11. doi: 10.1007/s12630-009-9165-z. Epub 2009 Aug 5. Can J Anaesth. 2009. PMID: 19655213 Clinical Trial.
-
Utility of the Pentax-AWS without interruption of chest compression: comparison of the Macintosh laryngoscope with the Pentax-AWS in manikin model.Resuscitation. 2010 Jan;81(1):69-73. doi: 10.1016/j.resuscitation.2009.09.031. Epub 2009 Nov 17. Resuscitation. 2010. PMID: 19919888
-
[Airtraq optical laryngoscopy].An Sist Sanit Navar. 2009 Jan-Apr;32(1):75-83. An Sist Sanit Navar. 2009. PMID: 19430513 Review. Spanish.
-
[Glottiscopes and videolaryngoscopes: a rational choice?].Ann Fr Anesth Reanim. 2013 Dec;32(12):e211-5. doi: 10.1016/j.annfar.2013.10.021. Epub 2013 Nov 16. Ann Fr Anesth Reanim. 2013. PMID: 24246658 Review. French.
Cited by
-
Recent advance in patient monitoring.Korean J Anesthesiol. 2010 Sep;59(3):144-59. doi: 10.4097/kjae.2010.59.3.144. Epub 2010 Sep 20. Korean J Anesthesiol. 2010. PMID: 20877698 Free PMC article.
-
Comparison of three video laryngoscopes and direct laryngoscopy for emergency endotracheal intubation: a retrospective cohort study.BMJ Open. 2019 Mar 30;9(3):e024927. doi: 10.1136/bmjopen-2018-024927. BMJ Open. 2019. PMID: 30928937 Free PMC article.
-
A comparison of the force applied on oral structures during intubation attempts between the Pentax-AWS airwayscope and the Macintosh laryngoscope: a high-fidelity simulator-based study.BMJ Open. 2014 Oct 8;4(10):e006416. doi: 10.1136/bmjopen-2014-006416. BMJ Open. 2014. PMID: 25296656 Free PMC article.
-
Influence of clinical experience of the Macintosh laryngoscope on performance with the Pentax-AWS Airway Scope(®), a rigid video-laryngoscope, by paramedics in Japan.J Anesth. 2015 Oct;29(5):672-7. doi: 10.1007/s00540-014-1960-1. Epub 2014 Dec 20. J Anesth. 2015. PMID: 25527323
-
A comparison of the Airtraq®, McGrath®, and Macintosh laryngoscopes for difficult paediatric intubation: A manikin study.PLoS One. 2017 Feb 10;12(2):e0171889. doi: 10.1371/journal.pone.0171889. eCollection 2017. PLoS One. 2017. PMID: 28187213 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical