A computerized assessment to compare the impact of standard, stereoscopic, and high-definition laparoscopic monitor displays on surgical technique
- PMID: 20361211
- DOI: 10.1007/s00464-010-1038-6
A computerized assessment to compare the impact of standard, stereoscopic, and high-definition laparoscopic monitor displays on surgical technique
Abstract
Background: Surgeons performing laparoscopic surgery have strong biases regarding the quality and nature of the laparoscopic video monitor display. In a comparative study, we used a unique computerized sensing and analysis system to evaluate the various types of monitors employed in laparoscopic surgery.
Methods: We compared the impact of different types of monitor displays on an individual's performance of a laparoscopic training task which required the subject to move the instrument to a set of targets. Participants (varying from no laparoscopic experience to board-certified surgeons) were asked to perform the assigned task while using all three display systems, which were randomly assigned: a conventional laparoscopic monitor system (2D), a high-definition monitor system (HD), and a stereoscopic display (3D). The effects of monitor system on various performance parameters (total time consumed to finish the task, average speed, and movement economy) were analyzed by computer. Each of the subjects filled out a subjective questionnaire at the end of their training session.
Results: A total of 27 participants completed our study. Performance with the HD monitor was significantly slower than with either the 3D or 2D monitor (p < 0.0001). Movement economy with the HD monitor was significantly reduced compared with the 3D (p < 0.0004) or 2D (p < 0.0001) monitor. In terms of average time required to complete the task, performance with the 3D monitor was significantly faster than with the HD (p < 0.0001) or 2D (p < 0.0086) monitor. However, the HD system was the overwhelming favorite according to subjective evaluation.
Conclusion: Computerized sensing and analysis is capable of quantitatively assessing the seemingly minor effect of monitor display on surgical training performance. The study demonstrates that, while users expressed a decided preference for HD systems, actual quantitative analysis indicates that HD monitors offer no statistically significant advantage and may even worsen performance compared with standard 2D or 3D laparoscopic monitors.
Similar articles
-
Comparative evaluation of HD 2D/3D laparoscopic monitors and benchmarking to a theoretically ideal 3D pseudodisplay: even well-experienced laparoscopists perform better with 3D.Surg Endosc. 2014 Aug;28(8):2387-97. doi: 10.1007/s00464-014-3487-9. Epub 2014 Mar 21. Surg Endosc. 2014. PMID: 24651895 Clinical Trial.
-
The effect on surgical skills of expert surgeons using 3D/HD and 2D/4K resolution monitors in laparoscopic phantom tasks.Surg Endosc. 2018 Oct;32(10):4228-4234. doi: 10.1007/s00464-018-6169-1. Epub 2018 Mar 30. Surg Endosc. 2018. PMID: 29603005 Clinical Trial.
-
Three-dimensional imaging improved the laparoscopic performance of inexperienced operators: a prospective trial.Surg Endosc. 2020 Nov;34(11):5083-5091. doi: 10.1007/s00464-019-07308-6. Epub 2019 Dec 6. Surg Endosc. 2020. PMID: 31811456
-
3D HD versus 2D HD: surgical task efficiency in standardised phantom tasks.Surg Endosc. 2012 May;26(5):1454-60. doi: 10.1007/s00464-011-2055-9. Epub 2011 Dec 17. Surg Endosc. 2012. PMID: 22179446 Clinical Trial.
-
The Conflicting Evidence of Three-dimensional Displays in Laparoscopy: A Review of Systems Old and New.Ann Surg. 2016 Feb;263(2):234-9. doi: 10.1097/SLA.0000000000001504. Ann Surg. 2016. PMID: 26501704 Review.
Cited by
-
Identifying specific surgical tools and methods for laparoscopic colorectal operations in obese patients.J Gastrointest Surg. 2012 Dec;16(12):2304-11. doi: 10.1007/s11605-012-1937-z. Epub 2012 Jul 14. J Gastrointest Surg. 2012. PMID: 22798184
-
What Are the Advantages of 3D Cameras in Gynaecological Laparoscopy?Geburtshilfe Frauenheilkd. 2017 Jan;77(1):45-51. doi: 10.1055/s-0042-120845. Geburtshilfe Frauenheilkd. 2017. PMID: 28190888 Free PMC article.
-
Comparison of 3D endoscopy and conventional 2D endoscopy in gastric endoscopic submucosal dissection: an ex vivo animal study.Surg Endosc. 2019 Dec;33(12):4164-4170. doi: 10.1007/s00464-019-06726-w. Epub 2019 Mar 6. Surg Endosc. 2019. PMID: 30843096
-
The usefulness of 3-dimensional endoscope systems in endoscopic surgery.Surg Endosc. 2016 Oct;30(10):4562-8. doi: 10.1007/s00464-016-4793-1. Epub 2016 Feb 19. Surg Endosc. 2016. PMID: 26895893
-
3-Dimensional (3D) laparoscopy improves operating time in small spaces without impact on hemodynamics and psychomental stress parameters of the surgeon.Surg Endosc. 2015 May;29(5):1231-9. doi: 10.1007/s00464-015-4083-3. Epub 2015 Feb 12. Surg Endosc. 2015. PMID: 25673344
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources