Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2010 Jun;24(3):366-72.
doi: 10.1007/s00540-010-0927-0. Epub 2010 Apr 6.

Distortion of anterior airway anatomy during laryngoscopy with the GlideScope videolaryngoscope

Affiliations
Comparative Study

Distortion of anterior airway anatomy during laryngoscopy with the GlideScope videolaryngoscope

Yoshihiro Hirabayashi et al. J Anesth. 2010 Jun.

Abstract

Purpose: A non-line-of-sight view is expected to cause less movement of the anterior airway anatomy and cervical spine during laryngeal visualization. Reduced distortion of anterior airway anatomy during laryngoscopy with the GlideScope videolaryngoscope (GVL), compared with the Macintosh laryngoscope, could explain the relatively easier nasotracheal intubation with the GVL. The purpose of this radiographic study was to compare the degree of anterior airway distortion and cervical spine movement during laryngoscopy with the GVL and the conventional Macintosh laryngoscope.

Methods: Twenty patients requiring general anesthesia and tracheal intubation were studied. Each patient underwent laryngoscopy using the first-generation GVL and a direct laryngoscope with a Macintosh blade. During each laryngoscopy, a radiograph was taken when the best view of the larynx was obtained. Independent radiologists with subspeciality training in musculoskeletal imaging evaluated anterior airway distortion and cervical spine movement.

Results: The distance between the epiglottis and the posterior pharyngeal wall during the GlideScope procedure was 21% less than that during the Macintosh laryngoscopy (P < 0.05). Anterior deviations of the vertebral bodies from baseline were 27, 32, 36, and 39% less at the atlas, C2, C3, and C4 vertebrae, respectively, during the GlideScope procedure than those measured during Macintosh laryngoscopy (P < 0.01). Cervical extension between the occiput and C4 during the GlideScope procedure was 23% less than that during Macintosh laryngoscopy (P < 0.05).

Conclusion: Both anterior airway distortion and cervical spine movement during laryngeal visualization were less with the GVL than with the Macintosh laryngoscope.

PubMed Disclaimer

References

    1. Anesthesiology. 1997 Dec;87(6):1335-42 - PubMed
    1. Br J Anaesth. 2009 Mar;102(3):424-30 - PubMed
    1. J Clin Anesth. 2006 Dec;18(8):611-5 - PubMed
    1. Anesth Analg. 2008 Mar;106(3):935-41, table of contents - PubMed
    1. Can J Anaesth. 2006 Nov;53(11):1163 - PubMed

Publication types

LinkOut - more resources