Setup accuracy of spine radiosurgery using cone beam computed tomography image guidance in patients with spinal implants
- PMID: 20367378
- DOI: 10.3171/2009.10.SPINE09249
Setup accuracy of spine radiosurgery using cone beam computed tomography image guidance in patients with spinal implants
Abstract
Object: Cone beam computed tomography (CBCT) image guidance technology has been adopted for use in spine radiosurgery. There is concern regarding the ability to safely and accurately perform spine radiosurgery without the use of implanted fiducials for image guidance in postsurgical cases in which titanium instrumentation and/or methylmethacrylate (MMA) has been implanted. In this study the authors prospectively evaluated the accuracy of the patient setup for spine radiosurgery by using CBCT image guidance in the context of orthopedic hardware at the site of disease.
Methods: The positioning deviations of 31 single-fraction spine radiosurgery treatments in patients with spinal implants were prospectively evaluated using the Elekta Synergy S 6-MV linear accelerator with a beam modulator and CBCT image guidance combined with a robotic couch that allows positioning correction in 3 translational and 3 rotational directions. To measure patient movement, 3 quality-assurance CBCT studies were performed and recorded: before, halfway through, and after radiosurgical treatment. The positioning data and fused images of planning CTs and CBCTs from the treatments were analyzed to determine intrafractional patient movements. From each of 3 CBCTs, 3 translational and 3 rotational coordinates were obtained.
Results: The prescribed dose to the gross tumor volume for the cohort was 12-18 Gy (mean 14 Gy) utilizing 9-14 coplanar intensity-modulated radiation therapy (IMRT) beams (mean 10 beams). At the halfway point of the radiosurgery, the translational variations and standard deviations were 0.6 +/- 0.6, 0.4 +/- 0.4, and 0.5 +/- 0.5 mm in the lateral (X), longitudinal (Y), and anteroposterior (Z) directions, respectively. The magnitude of the 3D vector (X,Y,Z) was 1.1 +/- 0.7 mm. Similarly, the variations immediately after treatment were 0.5 +/- 0.3, 0.4 +/- 0.4, and 0.5 +/- 0.6 mm along the X, Y, and Z directions, respectively. The 3D vector was 1.0 +/- 0.6 mm. The mean rotational angles were 0.3 +/- 0.4, 0.5 +/- 0.6, and 0.3 +/- 0.4 degrees along yaw, roll, and pitch, respectively, at the halfway point and 0.3 +/- 0.4, 0.6 +/- 0.6, and 0.4 +/- 0.5 degrees immediately after treatment.
Conclusions: Cone beam CT image guidance used for patient setup for spine radiosurgery was highly accurate despite the presence of spinal instrumentation and/or MMA at the level of the target volume. The presence of such spinal implants does not preclude safe treatment via spine radiosurgery in these patients.
Similar articles
-
Stereotactic radiosurgery for intradural spine tumors using cone-beam CT image guidance.Neurosurg Focus. 2017 Jan;42(1):E11. doi: 10.3171/2016.9.FOCUS16356. Neurosurg Focus. 2017. PMID: 28041317
-
Prospective evaluation of a dedicated spine radiosurgery program using the Elekta Synergy S system.J Neurosurg. 2010 Dec;113 Suppl:236-41. doi: 10.3171/2010.8.GKS10949. J Neurosurg. 2010. PMID: 21121807
-
Radiosurgery for benign tumors of the spine using the Synergy S with cone-beam computed tomography image guidance.J Neurosurg. 2012 Dec;117 Suppl:197-202. doi: 10.3171/2012.8.GKS12981. J Neurosurg. 2012. PMID: 23205810
-
Spinal radiosurgery: technology and clinical outcomes.Neurosurg Rev. 2009 Jan;32(1):1-12; discussion 12-3. doi: 10.1007/s10143-008-0167-z. Epub 2008 Sep 24. Neurosurg Rev. 2009. PMID: 18813965 Review.
-
Radiosurgery for spinal malignant tumors.Dtsch Arztebl Int. 2009 Feb;106(7):106-12. doi: 10.3238/arztebl.2009.0106. Epub 2009 Feb 13. Dtsch Arztebl Int. 2009. PMID: 19562022 Free PMC article. Review.
Cited by
-
Comparison of setup accuracy and efficiency between the Klarity system and BodyFIX system for spine stereotactic body radiation therapy.J Appl Clin Med Phys. 2022 Nov;23(11):e13804. doi: 10.1002/acm2.13804. Epub 2022 Oct 9. J Appl Clin Med Phys. 2022. PMID: 36210179 Free PMC article.
-
Efficiency gains for spinal radiosurgery using multicriteria optimization intensity modulated radiation therapy guided volumetric modulated arc therapy planning.Pract Radiat Oncol. 2015 Jan-Feb;5(1):49-55. doi: 10.1016/j.prro.2014.04.003. Epub 2014 May 27. Pract Radiat Oncol. 2015. PMID: 25413420 Free PMC article.
-
Prospective evaluation of target and spinal cord motion and dosimetric changes with respiration in spinal stereotactic body radiation therapy utilizing 4-D CT.J Radiosurg SBRT. 2016;4(3):191-201. J Radiosurg SBRT. 2016. PMID: 29296444 Free PMC article.
-
Clinical Evaluation of an Auto-Segmentation Tool for Spine SBRT Treatment.Front Oncol. 2022 Mar 14;12:842579. doi: 10.3389/fonc.2022.842579. eCollection 2022. Front Oncol. 2022. PMID: 35359361 Free PMC article.
-
Intrafractional motion in stereotactic body radiotherapy of spinal metastases utilizing cone beam computed tomography image guidance.Phys Imaging Radiat Oncol. 2019 Nov 2;12:1-6. doi: 10.1016/j.phro.2019.10.001. eCollection 2019 Oct. Phys Imaging Radiat Oncol. 2019. PMID: 33458287 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials