Multi-slice computed tomography: A tool for non-invasive temperature measurement?
- PMID: 20377364
- DOI: 10.3109/02656731003605654
Multi-slice computed tomography: A tool for non-invasive temperature measurement?
Abstract
Purpose: To investigate the potential of multi-slice computed tomography (MSCT) as a tool for non-invasive temperature measurement.
Materials and methods: Samples of water, 0.9% saline, sunflower oil and dilutions of (1:32, 1:64, 1:128) contrast agent (Iopromid 370, BayerSchering Pharma, Berlin) were heated in a plexiglass phantom. In a first set-up, samples of 0.9% saline solution were scanned at defined temperatures (25 degrees -75 degrees C; 5 degrees C intervals) using a clinical CT scanner. Scan parameters (tube current-time product, tube voltage, collimation, slice thickness) were systematically varied. In a second set-up samples of the different fluids (water, sunflower oil, contrast agent dilutions) were scanned using the following scan protocol: 250 mAs, 140 kV, 1.2 mm collimation, 9.6 mm slice thickness. CT numbers were measured in reconstructed axial images at the different temperatures. A regression analysis was performed to investigate the relationship between temperature and CT number.
Results: Standard deviation of measured CT numbers decreased with increasing tube current-time product, increasing tube voltage, thicker collimation and higher slice thickness. Regression analysis showed an inverse relationship between temperature and CT number for all fluids with regression coefficients of -0.471 (0.9% saline), 0.447 (water), -0.679 (sunflower oil), -0.420 (contrast agent 1:32), -0.414 (contrast agent 1:64) and -0.441 (contrast agent 1:128), respectively.
Conclusion: Multi-slice computed tomography can depict thermal density expansion of different fluids. Based on these results the implementation of a temperature discrimination of several degrees C at a high spatial resolution is achievable.
Similar articles
-
Relationships of clinical protocols and reconstruction kernels with image quality and radiation dose in a 128-slice CT scanner: study with an anthropomorphic and water phantom.Eur J Radiol. 2012 May;81(5):e699-703. doi: 10.1016/j.ejrad.2011.01.078. Epub 2011 Feb 12. Eur J Radiol. 2012. PMID: 21316888
-
The influence of focal spot blooming on high-contrast spatial resolution in CT imaging.Med Phys. 2015 Oct;42(10):6011-20. doi: 10.1118/1.4931053. Med Phys. 2015. PMID: 26429276
-
Optimization of low-contrast detectability in thin-collimated modern multidetector CT using an interactive sliding-thin-slab averaging algorithm.Invest Radiol. 2008 Apr;43(4):229-35. doi: 10.1097/RLI.0b013e3181614f2d. Invest Radiol. 2008. PMID: 18340246
-
Effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral pediatric and adult CT angiography: a phantom study.Invest Radiol. 2013 Apr;48(4):192-9. doi: 10.1097/RLI.0b013e31827efc17. Invest Radiol. 2013. PMID: 23344518
-
The importance of spectral separation: an assessment of dual-energy spectral separation for quantitative ability and dose efficiency.Invest Radiol. 2015 Feb;50(2):114-8. doi: 10.1097/RLI.0000000000000109. Invest Radiol. 2015. PMID: 25373305
Cited by
-
How large is the periablational zone after radiofrequency and microwave ablation? Computer-based comparative study of two currently used clinical devices.Int J Hyperthermia. 2020;37(1):1131-1138. doi: 10.1080/02656736.2020.1823022. Int J Hyperthermia. 2020. PMID: 32996794 Free PMC article.
-
Assessment of thermal sensitivity of CT during heating of liver: an ex vivo study.Br J Radiol. 2012 Sep;85(1017):e661-5. doi: 10.1259/bjr/23942179. Br J Radiol. 2012. PMID: 22919016 Free PMC article.
-
CT-based thermometry with virtual monoenergetic images by dual-energy of fat, muscle and bone using FBP, iterative and deep learning-based reconstruction.Eur Radiol. 2022 Jan;32(1):424-431. doi: 10.1007/s00330-021-08206-z. Epub 2021 Jul 29. Eur Radiol. 2022. PMID: 34327575 Free PMC article.
-
Perspectives of breast cancer thermotherapies.J Cancer. 2014 May 29;5(6):472-9. doi: 10.7150/jca.8693. eCollection 2014. J Cancer. 2014. PMID: 24959300 Free PMC article. Review.
-
Evaluation of Different Registration Algorithms to Reduce Motion Artifacts in CT-Thermography (CTT).Diagnostics (Basel). 2023 Jun 15;13(12):2076. doi: 10.3390/diagnostics13122076. Diagnostics (Basel). 2023. PMID: 37370971 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical