Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2010 Sep;99(9):1314-8.
doi: 10.1111/j.1651-2227.2010.01816.x.

Neopuff T-piece mask resuscitator: is mask leak related to watching the pressure dial?

Affiliations
Randomized Controlled Trial

Neopuff T-piece mask resuscitator: is mask leak related to watching the pressure dial?

Mark B Tracy et al. Acta Paediatr. 2010 Sep.

Abstract

Aim: The aim of the study is to compare mask leak and delivered ventilation during Neopuff (NP) mask ventilation in two modes: (i) with NP pressure dial hidden and resuscitator watching chest wall (CW) rise with, (ii) CW movement hidden and resuscitator watching NP pressure dial.

Methods: Thirty-six participants gave mask ventilation to a modified manikin designed to measure mask leak and delivered ventilation for two minutes in each mode randomly assigned. Paired t-tests were used to analyse differences in mean values. Linear regression was used to determine the association of mask leak with delivered ventilation.

Results: Of 7277 inflations analysed, 3621 were observing chest wall mode (CWM) and 3656 observing NP mode (NPM). Mask leak was similar between the groups; 31.6% for CWM and 31.5% (p = 0.56) for NPM. There were no significant differences in airways pressures and expired tidal volumes (TVe) between modes. Mask leak was strongly associated with TVe (R = -0.86 p < 0.0001) and with peak inspiratory pressure (PIP) (R = -0.51 p < 0.0001). TVe was associated with PIP (R = 0.51 p < 0.0001).

Conclusion: This study provides reassurance that NP mask leak is not greater when resuscitators watch the NP pressure dial. Mask leak is related to TVe. Mask ventilation training with manikins should include tidal volume measurements.

PubMed Disclaimer

Publication types

LinkOut - more resources