Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy

Affiliations
Comparative Study

Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT: plan quality, delivery efficiency and accuracy

Min Rao et al. Med Phys. 2010 Mar.

Abstract

Purpose: Helical tomotherapy (HT) and volumetric modulated arc therapy (VMAT) are arc-based approaches to IMRT delivery. The objective of this study is to compare VMAT to both HT and fixed field IMRT in terms of plan quality, delivery efficiency, and accuracy.

Methods: Eighteen cases including six prostate, six head-and-neck, and six lung cases were selected for this study. IMRT plans were developed using direct machine parameter optimization in the Pinnacle3 treatment planning system. HT plans were developed using a Hi-Art II planning station. VMAT plans were generated using both the Pinnacle3 SmartArc IMRT module and a home-grown arc sequencing algorithm. VMAT and HT plans were delivered using Elekta's PreciseBeam VMAT linac control system (Elekta AB, Stockholm, Sweden) and a TomoTherapy Hi-Art II system (TomoTherapy Inc., Madison, WI), respectively. Treatment plan quality assurance (QA) for VMAT was performed using the IBA MatriXX system while an ion chamber and films were used for HT plan QA.

Results: The results demonstrate that both VMAT and HT are capable of providing more uniform target doses and improved normal tissue sparing as compared with fixed field IMRT. In terms of delivery efficiency, VMAT plan deliveries on average took 2.2 min for prostate and lung cases and 4.6 min for head-and-neck cases. These values increased to 4.7 and 7.0 min for HT plans.

Conclusions: Both VMAT and HT plans can be delivered accurately based on their own QA standards. Overall, VMAT was able to provide approximately a 40% reduction in treatment time while maintaining comparable plan quality to that of HT.