AMS 700CX/CXM inflatable penile prosthesis has high mechanical reliability at long-term follow-up
- PMID: 20384938
- DOI: 10.1111/j.1743-6109.2010.01801.x
AMS 700CX/CXM inflatable penile prosthesis has high mechanical reliability at long-term follow-up
Abstract
Introduction: AMS 700CX/CXM inflatable penile prosthesis is increasingly applied for the treatment of erectile dysfunction (ED). However, there are a few long-term survival data of the inflatable penile prosthesis (IPP) over 10 years.
Aim: To determine the long-term mechanical reliability of AMS 700CX/CXM inflatable penile prosthesis in patients with ED.
Methods: A total of 438 consecutive patients with ED received implantation of an AMS 700CX/CXM penile prosthesis at our institution from January 1991 to April 2009. In 397 patients (90.7%), the medical records were available and current status of penile prosthesis could be obtained by a direct telephone interview. The overall and mechanical survival rates of penile prosthesis were evaluated using Kaplan-Meier method.
Main outcome measures: Assessing the mechanical and overall survival rates of the AMS 700CX/CXM penile prosthesis using Kaplan-Meier analysis, and looking for clinical factors related to survival of the CX/CXM using log-rank test.
Results: Mean age of 397 patients was 63.1 years (range, 24-93) and follow-up duration was 113 months (range 1-219). Eighty-two patients (20.6%) experienced mechanical failure at a median follow-up of 82 months. Mechanical survival rate of the penile prosthesis was 97.6%, 93.2% and 78.2% at 3, 5, and 10 years after implantation, respectively. 12 patients (3.0%) experienced nonmechanical failure including infections, tissue erosion resulting in cylinder protrusion at the meatus and chronic discomfort. Overall survival rate of the penile prosthesis was 95.0%, 91.0% and 75.5% at 3, 5, and 10 years after implantation, respectively. Patients with neurogenic cause for ED showed lower median overall survival of penile prosthesis compared with patients with non-neurogenic cause. Patient age, obesity, and diabetes mellitus had no association with overall survival of penile prosthesis after implantation.
Conclusions: The AMS 700CX/CXM could be accepted and applied in more patients as a reliable treatment alternative of ED.
Similar articles
-
Long-term mechanical reliability of AMS 700CX/CXM inflatable penile prosthesis.J Urol. 2006 Dec;176(6 Pt 1):2599-601; discussion 2601. doi: 10.1016/j.juro.2006.08.012. J Urol. 2006. PMID: 17085168
-
Long term patient satisfaction and quality of life with AMS700CX inflatable penile prosthesis.Arch Ital Urol Androl. 2013 Sep 26;85(3):133-7. doi: 10.4081/aiua.2013.3.133. Arch Ital Urol Androl. 2013. PMID: 24085235
-
Long-term mechanical reliability of AMS 700 series inflatable penile prostheses: comparison of CX/CXM and Ultrex cylinders.J Urol. 1997 Oct;158(4):1400-2. J Urol. 1997. PMID: 9302130
-
Mid-term outcomes of minimally invasive infrapubic approach for inflatable penile prosthesis implantation: A single-center study and literature review.Andrology. 2024 Mar;12(3):624-632. doi: 10.1111/andr.13497. Epub 2023 Jul 20. Andrology. 2024. PMID: 37452742 Review.
-
Ambicor Two-Piece Inflatable Penile Prosthesis: Background and Contemporary Outcomes.Sex Med Rev. 2018 Apr;6(2):319-327. doi: 10.1016/j.sxmr.2017.07.010. Epub 2017 Sep 2. Sex Med Rev. 2018. PMID: 28874326 Review.
Cited by
-
A population-based analysis of contemporary rates of reoperation for penile prosthesis procedures.Urology. 2014 Jul;84(1):112-6. doi: 10.1016/j.urology.2014.02.028. Epub 2014 Apr 29. Urology. 2014. PMID: 24785986 Free PMC article.
-
Narrative review of penile prosthetic implant technology and surgical results, including transgender patients.Transl Androl Urol. 2021 Jun;10(6):2629-2647. doi: 10.21037/tau-20-1279. Transl Androl Urol. 2021. PMID: 34295749 Free PMC article. Review.
-
Current management of penile implant infections, device reliability, and optimizing cosmetic outcome.Curr Urol Rep. 2014 Jun;15(6):413. doi: 10.1007/s11934-014-0413-6. Curr Urol Rep. 2014. PMID: 24740273 Review.
-
Techniques and considerations of prosthetic surgery after phalloplasty in the transgender male.Transl Androl Urol. 2019 Jun;8(3):273-282. doi: 10.21037/tau.2019.06.02. Transl Androl Urol. 2019. PMID: 31380234 Free PMC article. Review.
-
Role of Penile Prosthesis in Priapism: A Review.World J Mens Health. 2018 Jan;36(1):4-14. doi: 10.5534/wjmh.17040. World J Mens Health. 2018. PMID: 29299902 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous