Decreased CSF-flow artefacts in T2 imaging of the cervical spine with periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER/BLADE)
- PMID: 20386890
- DOI: 10.1007/s00234-010-0691-8
Decreased CSF-flow artefacts in T2 imaging of the cervical spine with periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER/BLADE)
Abstract
Introduction: The cervical spine is prone to artefacts in T2 MR-imaging due to patient movements and cerebrospinal fluid flow. The periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER/BLADE) acquisition method was developed to reduce motion artefacts. We sought to determine if T2-BLADE is superior to T2-TSE with conventional k-space reading.
Methods: Twenty-five patients were examined using a 1.5 T MR-scanner. T2-weighted imaging of the cervical spine in sagittal and axial orientation using conventional or BLADE k-space reading was performed. Spinal cord, subarachnoid space, vertebrae and discs were evaluated by two independent observers using a scale from 0 (non-diagnostic) to 3 (excellent). Interobserver correlation was assessed as Cohen's kappa. Results of Mann-Whitney U test with p < 0.05 were regarded as significant. Furthermore, the investigators were asked for subjective evaluation in consensus.
Results: Overall interobserver accuracy of κ = 0.91 was obtained. Comparison of sagittal images showed better values for all investigated structures in T2-BLADE: spinal cord (TSE/BLADE: 1.52/2.04; p < 0.001), subarachnoid space (1.36/2.06; p < 0.001) and vertebrae/discs (1.66/2.86; p < 0.001). Comparison of axial images showed better values in T2-BLADE for spinal cord (1.68/1.86; p = 0.149) and vertebrae/discs (1.0/1.96: p < 0.001) while subarachnoid space was better to be evaluated in conventional T2-TSE (1.94/1.12; p < 0.001). In sagittal orientation, motion- and CSF-flow artefacts were reduced in T2-BLADE. In axial orientation, however, CSF-flow artefacts were pronounced in T2-BLADE.
Conclusion: The image quality of the sagittal T2-BLADE sequences was significantly better than the T2-TSE and acquired in less time. In axial orientation, increased CSF-flow artefacts may reduce accuracy of structures in the subarachnoid space.
Similar articles
-
BLADE Sequences in Transverse T2-weighted MR Imaging of the Cervical Spine. Cut-off for Artefacts?Rofo. 2015 May;36(2):102-8. doi: 10.1055/s-0034-1385179. Epub 2014 Sep 22. Rofo. 2015. PMID: 25912327 Review.
-
BLADE sequences in transverse T2-weighted MR imaging of the cervical spine. Cut-off for Artefacts?Rofo. 2015 Feb;187(2):102-8. doi: 10.1055/s-0034-1385179. Epub 2014 Sep 22. Rofo. 2015. PMID: 25243527
-
BLADE in sagittal T2-weighted MR imaging of the cervical spine.AJNR Am J Neuroradiol. 2010 Apr;31(4):674-81. doi: 10.3174/ajnr.A1894. Epub 2009 Nov 26. AJNR Am J Neuroradiol. 2010. PMID: 19942708 Free PMC article.
-
BLADE sequences in sagittal T2-weighted MR imaging of the cervical spine and spinal cord--lesion detection and clinical value.Rofo. 2014 Jan;186(1):47-53. doi: 10.1055/s-0033-1350346. Epub 2013 Sep 2. Rofo. 2014. PMID: 23999783
-
State-of-the-art magnetic resonance imaging sequences for pediatric body imaging.Pediatr Radiol. 2023 Jun;53(7):1285-1299. doi: 10.1007/s00247-022-05528-y. Epub 2022 Oct 18. Pediatr Radiol. 2023. PMID: 36255456 Review.
Cited by
-
Elimination of motion and pulsation artifacts using BLADE sequences in shoulder MR imaging.Skeletal Radiol. 2015 Nov;44(11):1619-26. doi: 10.1007/s00256-015-2232-3. Epub 2015 Aug 26. Skeletal Radiol. 2015. PMID: 26306388
-
Magnetic Resonance Imaging in Cervical Spine Trauma: More Than Soft Tissue Illustration.Cureus. 2022 Jan 22;14(1):e21493. doi: 10.7759/cureus.21493. eCollection 2022 Jan. Cureus. 2022. PMID: 35223270 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical