Comparative assessment of energy-mapping approaches in CT-based attenuation correction for PET
- PMID: 20387123
- DOI: 10.1007/s11307-010-0303-3
Comparative assessment of energy-mapping approaches in CT-based attenuation correction for PET
Abstract
Introduction: Reliable quantification in positron emission tomography (PET) requires accurate attenuation correction of emission data, which in turn entails accurate determination of the attenuation map (µ-map) of the object under study. One of the main steps involved in CT-based attenuation correction (CTAC) is energy-mapping, or the conversion of linear attenuation coefficients (µ) calculated at the effective CT energy to those corresponding to 511 keV.
Materials and methods: The aim of this study is to compare different energy-mapping techniques including scaling, segmentation, the hybrid method, the bilinear calibration curve technique and the dual-energy approach to generate the µ-maps required for attenuation correction. In addition, our newly proposed method involving a quadratic polynomial calibration curve was also assessed. The µ-maps generated for both phantom and clinical studies were assessed qualitatively and quantitatively. A cylindrical polyethylene phantom containing different concentrations of K(2)HPO(4) in water was scanned and the µ-maps calculated from the corresponding CT images using the above-referenced energy-mapping methods. The CT images of five whole-body data sets acquired on a GE Discovery LS PET/CT scanner were employed to generate µ-maps using different energy-mapping approaches that were compared with the µ-maps generated at 511 keV using (68)Ge/(68)Ga rod sources. In another experiment, the evaluation was performed on PET images of a clinical study corrected for attenuation using µ-maps generated using the above described methods. The evaluation was performed for three different tissue types, namely, soft tissue, lung, and bone.
Results and discussion: All energy-mapping methods yielded almost similar results for soft tissues. The mean relative differences between scaling, segmentation, hybrid, bilinear, and quadratic polynomial calibration curve methods and the transmission scan serving as reference were 6.60%, 6.56%, 6.60%, 5.96%, and 7.36%, respectively. However, the scaling method produced the largest difference (16%) for bone tissues. For lung tissues, the segmentation method produced the largest difference (14.9%). The results for reconstructed PET images followed a similar trend. For soft tissues, all energy-mapping methods yield results in nearly the same range. However, in bone tissues, the scaling method resulted in considerable bias in the µ-maps and the reconstructed PET images. The segmentation method also produced noticeable bias especially in regions with variable densities such as the lung, since a single µ is assigned to the lungs. Apart from the aforementioned case, despite small differences in the generated µ-maps, the use of different energy-mapping methods does not affect, to a visible or measurable extent, the reconstructed PET images.
Similar articles
-
A novel energy mapping approach for CT-based attenuation correction in PET.Med Phys. 2012 Apr;39(4):2078-89. doi: 10.1118/1.3694108. Med Phys. 2012. PMID: 22482629
-
CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PET.Eur J Nucl Med Mol Imaging. 2003 Mar;30(3):344-53. doi: 10.1007/s00259-002-1070-4. Epub 2002 Dec 20. Eur J Nucl Med Mol Imaging. 2003. PMID: 12634961
-
Correction for oral contrast artifacts in CT attenuation-corrected PET images obtained by combined PET/CT.J Nucl Med. 2003 Dec;44(12):1940-4. J Nucl Med. 2003. PMID: 14660720
-
Computed tomography-based attenuation correction in neurological positron emission tomography: evaluation of the effect of the X-ray tube voltage on quantitative analysis.Nucl Med Commun. 2006 Apr;27(4):339-46. doi: 10.1097/01.mnm.0000203631.23407.23. Nucl Med Commun. 2006. PMID: 16531919
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
Cited by
-
Investigating the feasibility of generating dual-energy CT from one 120-kVp CT scan: a phantom study.J Appl Clin Med Phys. 2021 Feb;22(2):126-137. doi: 10.1002/acm2.13148. Epub 2020 Oct 14. J Appl Clin Med Phys. 2021. PMID: 33426800 Free PMC article.
-
Anomaly Detection and Artifact Recovery in PET Attenuation-Correction Images Using the Likelihood Function.IEEE J Sel Top Signal Process. 2013 Feb;7(1):10.1109/JSTSP.2012.2237380. doi: 10.1109/JSTSP.2012.2237380. IEEE J Sel Top Signal Process. 2013. PMID: 24198866 Free PMC article.
-
Respiratory-induced errors in tumor quantification and delineation in CT attenuation-corrected PET images: effects of tumor size, tumor location, and respiratory trace: a simulation study using the 4D XCAT phantom.Mol Imaging Biol. 2013 Dec;15(6):655-65. doi: 10.1007/s11307-013-0656-5. Mol Imaging Biol. 2013. PMID: 23780352
-
PET/MRI attenuation estimation in the lung: A review of past, present, and potential techniques.Med Phys. 2020 Feb;47(2):790-811. doi: 10.1002/mp.13943. Epub 2020 Jan 1. Med Phys. 2020. PMID: 31794071 Free PMC article. Review.
-
Effects of attenuation map accuracy on attenuation-corrected micro-SPECT images.EJNMMI Res. 2013 Jan 31;3(1):7. doi: 10.1186/2191-219X-3-7. EJNMMI Res. 2013. PMID: 23369630 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical