Comparison of the clinical information provided by the FreeStyle Navigator continuous interstitial glucose monitor versus traditional blood glucose readings
- PMID: 20388046
- DOI: 10.1089/dia.2009.0136
Comparison of the clinical information provided by the FreeStyle Navigator continuous interstitial glucose monitor versus traditional blood glucose readings
Abstract
Background: The purpose of the analysis was to compare the clinical utility of data from traditional self-monitoring of blood glucose (SMBG) to that of continuous glucose monitoring (CGM).
Methods: A clinical study of the clinical accuracy of the FreeStyle Navigator CGM System (Abbott Diabetes Care, Alameda, CA), which includes SMBG capabilities, was conducted by comparison to the YSI blood glucose analyzer (YSI Inc., Yellow Springs, OH) using 58 subjects with type 1 diabetes. The Continuous Glucose-Error Grid Analysis (CG-EGA) was used as the analytical tool.
Results: Using CG-EGA, the "clinically accurate," "benign errors," and "clinical errors" were 86.8%, 8.7%, and 4.5% for SMBG and 92.7%, 3.7%, and 3.6% for CGM, respectively. If blood glucose is viewed as a process in time, SMBG would provide accurate information about this process 86.8% of the time, whereas CGM would provide accurate information about this process 92.7% of the time (P < 0.0001). In the hypoglycemic range, however, SMBG is more accurate as the "clinically accurate," "benign errors," and "clinical errors" were 83.5%, 6.4%, and 10.1% for SMBG and 57.1%, 8.4%, and 34.5% (P < 0.0001) for CGM, respectively.
Conclusions: While SMBG produces more accurate instantaneous glucose values than CGM, control of blood glucose involves a system in flux, and CGM provides more detailed insight into the dynamics of that system. In the normal and elevated glucose ranges, the additional information about the direction and rate of glucose change provided by the FreeStyle Navigator CGM System increases the ability to make correct clinical decisions when compared to episodic SMBG tests.
Similar articles
-
Comparison of accuracy and safety of the SEVEN and the Navigator continuous glucose monitoring systems.Diabetes Technol Ther. 2009 Feb;11(2):65-72. doi: 10.1089/dia.2008.0109. Diabetes Technol Ther. 2009. PMID: 19243265
-
Detection of hypoglycemia with continuous interstitial and traditional blood glucose monitoring using the FreeStyle Navigator Continuous Glucose Monitoring System.Diabetes Technol Ther. 2009 Mar;11(3):145-50. doi: 10.1089/dia.2008.0047. Diabetes Technol Ther. 2009. PMID: 19216686
-
FreeStyle Navigator Continuous Glucose Monitoring System with TRUstart algorithm, a 1-hour warm-up time.J Diabetes Sci Technol. 2011 Jan 1;5(1):99-106. doi: 10.1177/193229681100500114. J Diabetes Sci Technol. 2011. PMID: 21303631 Free PMC article. Clinical Trial.
-
Diabetes technology and treatments in the paediatric age group.Int J Clin Pract Suppl. 2011 Feb;(170):76-82. doi: 10.1111/j.1742-1241.2010.02582.x. Int J Clin Pract Suppl. 2011. PMID: 21323816 Review.
-
Nonadjunctive Use of Continuous Glucose Monitoring for Diabetes Treatment Decisions.J Diabetes Sci Technol. 2016 Aug 22;10(5):1169-73. doi: 10.1177/1932296816631569. Print 2016 Sep. J Diabetes Sci Technol. 2016. PMID: 26880390 Free PMC article. Review.
Cited by
-
Accuracy and reliability of continuous glucose monitoring systems: a head-to-head comparison.Diabetes Technol Ther. 2013 Aug;15(8):722-7. doi: 10.1089/dia.2013.0049. Epub 2013 May 7. Diabetes Technol Ther. 2013. PMID: 23650900 Free PMC article. Clinical Trial.
-
Assessment of patient-led or physician-driven continuous glucose monitoring in patients with poorly controlled type 1 diabetes using basal-bolus insulin regimens: a 1-year multicenter study.Diabetes Care. 2012 May;35(5):965-71. doi: 10.2337/dc11-2021. Epub 2012 Mar 28. Diabetes Care. 2012. PMID: 22456864 Free PMC article. Clinical Trial.
-
Assessing sensor accuracy for non-adjunct use of continuous glucose monitoring.Diabetes Technol Ther. 2015 Mar;17(3):177-86. doi: 10.1089/dia.2014.0272. Epub 2014 Dec 1. Diabetes Technol Ther. 2015. PMID: 25436913 Free PMC article.
-
Clinical Performance Evaluation of Continuous Glucose Monitoring Systems: A Scoping Review and Recommendations for Reporting.J Diabetes Sci Technol. 2023 Nov;17(6):1506-1526. doi: 10.1177/19322968231190941. Epub 2023 Aug 20. J Diabetes Sci Technol. 2023. PMID: 37599389 Free PMC article.
-
The challenges of measuring glycemic variability.J Diabetes Sci Technol. 2012 May 1;6(3):712-5. doi: 10.1177/193229681200600328. J Diabetes Sci Technol. 2012. PMID: 22768904 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical