[Measuring Kt by ionic dialysance is a useful tool for assessing dialysis dose in critical patients]
- PMID: 20393622
- DOI: 10.3265/Nefrologia.pre2010.Mar.10335
[Measuring Kt by ionic dialysance is a useful tool for assessing dialysis dose in critical patients]
Abstract
Aim: To evaluate the Kt assessed through ionic dializance (KtOCM) in UCI patients undergoing renal replacement therapy for acute kidney injury, comparing the results with those obtained through the urea removal rate method determined by dialyzate collection (Kturea).
Material and methods: 18 adult UCI staying individuals suffering from renal replacement therapy requiring oliguric acute kidney injury were included in this study. RRT consisted in intermittent or extended hemodialysis performed through a Fresenius 4008E dialysis machine equipped with an on-line clearance monitor (OCM Fresenius). The KtOCM results were provided automatically. The Spearman correlation test was used to assess the relationship between the two exploratory methods and the Student s t test to compare the results obtained by the KtOCM and the Kturea.
Results: 35 treatments were analyzed. There were not statistically significant differences between the results form the KtOCM and the Kturea (34.9 +/- 10.69 vs 32.78 +/- 11.31, p = NS). A remarkable association was find between both methods (r = 0.87; 95CI, 0.76-0.94; p < 0.001).
Conclusions: The assessment of Kt through ionic dialyzance is a simple method to estimate the dose of dialysis in critically ill patients and is and useful tool to monitor and adjust the RRT in real time according to a target dose.
Comment in
-
[Clearance dose in acute kidney injury].Nefrologia. 2010;30(2):147-50. doi: 10.3265/Nefrologia.pre2010.Mar.10327. Nefrologia. 2010. PMID: 20393617 Spanish. No abstract available.
Similar articles
-
[The monitoring of dialysis dose by ionic dialysance-based Kt reveals less dialysis adequacy than the Kt/V(UREA)-based measurement in critically ill patients with acute renal failure].Nefrologia. 2010;30(2):232-5. doi: 10.3265/Nefrologia.pre2010.Mar.10336. Nefrologia. 2010. PMID: 20393623 Spanish.
-
Factors determining a low dose of haemodialysis as measured by ionic dialysance in critical patients with acute kidney injury.Nefrologia. 2012 May 14;32(3):359-66. doi: 10.3265/Nefrologia.pre2012.Jan.11284. Epub 2012 Apr 17. Nefrologia. 2012. PMID: 22535160 English, Spanish.
-
Influence of the ionic dialysance monitor on Kt measurement in hemodialysis.Am J Kidney Dis. 2008 Jul;52(1):85-92. doi: 10.1053/j.ajkd.2008.03.014. Epub 2008 May 2. Am J Kidney Dis. 2008. PMID: 18455852
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
-
Renal replacement therapy in acute renal failure: solute removal mechanisms and dose quantification.Kidney Int Suppl. 1998 May;66:S133-7. Kidney Int Suppl. 1998. PMID: 9573590 Review.