[The monitoring of dialysis dose by ionic dialysance-based Kt reveals less dialysis adequacy than the Kt/V(UREA)-based measurement in critically ill patients with acute renal failure]
- PMID: 20393623
- DOI: 10.3265/Nefrologia.pre2010.Mar.10336
[The monitoring of dialysis dose by ionic dialysance-based Kt reveals less dialysis adequacy than the Kt/V(UREA)-based measurement in critically ill patients with acute renal failure]
Abstract
Introduction: Measurement of dialysis dose by methods based on urea kinetics (Kt/VUREA) are hardly applicable to critical ill patients with acute renal failure (ARF). However, it is the base of the ADQI consensus recommendation for the target minimum dose.
Objective: To evaluate the usefulness of the real-time measurement of delivered dialysis dose (Kt) by means of the ionic dialysance (KtID) in the critically ill patient and to compare adequacy of dialysis dose between KtID and traditional Kt/V(UREA).
Material and methods: Prospective observational study in 17 critically ill patients with ARF requiring acute hemodialysis with a predefined prescription for the study (51 measures).
Results: The mean delivered Kt/V(UREA) was 1.19 +/- 0.14, with 59% of the sessions with values equal or above the ADQI recommendation. On the contrary, the mean KtID values obtained was 37.6 +/- 1 l, with only 29.4% of the sessions being equal or greater than the recommended values.
Conclusions: Dialysis dose monitoring by means of KtID reveals a lower degree of adequacy as compared to the traditional Kt/V(UREA) method. The dynamic character of KtID monitoring can allow the adaptation of each dialysis session ("K" and/or "t") in order to achieve the recommended dose.
Comment in
-
[Clearance dose in acute kidney injury].Nefrologia. 2010;30(2):147-50. doi: 10.3265/Nefrologia.pre2010.Mar.10327. Nefrologia. 2010. PMID: 20393617 Spanish. No abstract available.
Similar articles
-
[Measuring Kt by ionic dialysance is a useful tool for assessing dialysis dose in critical patients].Nefrologia. 2010;30(2):227-31. doi: 10.3265/Nefrologia.pre2010.Mar.10335. Nefrologia. 2010. PMID: 20393622 Spanish.
-
Influence of the ionic dialysance monitor on Kt measurement in hemodialysis.Am J Kidney Dis. 2008 Jul;52(1):85-92. doi: 10.1053/j.ajkd.2008.03.014. Epub 2008 May 2. Am J Kidney Dis. 2008. PMID: 18455852
-
Disease severity adversely affects delivery of dialysis in acute renal failure.Nephron Clin Pract. 2007;107(4):c163-9. doi: 10.1159/000110592. Epub 2007 Nov 1. Nephron Clin Pract. 2007. PMID: 17975323
-
[Ionic dialysance and quality control in hemodialysis].Nephrologie. 2001;22(5):191-7. Nephrologie. 2001. PMID: 11572165 Review. French.
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
Cited by
-
Impact of the intensity of intermittent renal replacement therapy in critically ill patients.J Nephrol. 2021 Feb;34(1):105-112. doi: 10.1007/s40620-020-00760-x. Epub 2020 Jun 3. J Nephrol. 2021. PMID: 32495232
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical