Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
- PMID: 20397019
- DOI: 10.1007/s00330-010-1792-9
Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
Abstract
Objective: To evaluate coronary stents in vitro using 128-slice-dual-source computed tomography (CT).
Methods: Twelve different coronary stents placed in a non-moving cardiac/chest phantom were examined by 128-slice dual-source CT using three CT protocols [high-pitch spiral (HPS), sequential (SEQ) and conventional spiral (SPIR)]. Artificial in-stent lumen narrowing (ALN), visible inner stent area (VIA), artificial in-stent lumen attenuation (ALA) in percent, image noise inside/outside the stent and CTDIvol were measured.
Results: Mean ALN was 46% for HPS, 44% for SEQ and 47% for SPIR without significant difference. Mean VIA was similar with 31% for HPS, 30% for SEQ and 33% for SPIR. Mean ALA was, at 5% for HPS, significantly lower compared with -11% for SPIR (p = 0.024), but not different from SEQ with -1%. Mean image noise was significantly higher for HPS compared with SEQ and SPIR inside and outside the stent (p < 0.001). CTDIvol was lower for HPS (5.17 mGy), compared with SEQ (9.02 mGy) and SPIR (55.97 mGy), respectively.
Conclusion: The HPS mode of 128-slice dual-source CT yields fewer artefacts inside the stent lumen compared with SPIR and SEQ, but image noise is higher. ALN is still too high for routine stent evaluation in clinical practice. Radiation dose of the HPS mode is markedly (less than about tenfold) reduced.
Similar articles
-
Coronary stent imaging with dual-source CT: assessment of lumen visibility using different convolution kernels and postprocessing filters.Acta Radiol. 2015 Jan;56(1):42-50. doi: 10.1177/0284185113517229. Epub 2014 Jan 7. Acta Radiol. 2015. PMID: 24399513
-
High-pitch 128-slice dual-source CT for the assessment of coronary stents in a phantom model.Acad Radiol. 2010 Nov;17(11):1366-74. doi: 10.1016/j.acra.2010.06.017. Acad Radiol. 2010. PMID: 20801697
-
Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents: first in vivo experience.Eur J Radiol. 2013 Apr;82(4):617-22. doi: 10.1016/j.ejrad.2012.11.033. Epub 2012 Dec 23. Eur J Radiol. 2013. PMID: 23265926
-
In vitro imaging of coronary artery stents: are there differences between 16- and 64-slice CT scanners?Eur J Radiol. 2008 Dec;68(3):465-70. doi: 10.1016/j.ejrad.2007.08.032. Epub 2007 Oct 29. Eur J Radiol. 2008. PMID: 17913428
-
Coronary artery stent evaluation by combining iterative reconstruction and high-resolution kernel at coronary CT angiography.Acad Radiol. 2012 Nov;19(11):1324-31. doi: 10.1016/j.acra.2012.06.013. Epub 2012 Aug 28. Acad Radiol. 2012. PMID: 22951109
Cited by
-
A systematic approach for analysis, interpretation, and reporting of coronary CTA studies.Insights Imaging. 2012 Jun;3(3):215-28. doi: 10.1007/s13244-012-0167-y. Epub 2012 May 1. Insights Imaging. 2012. PMID: 22696084 Free PMC article.
-
CT Imaging of Coronary Stents: Past, Present, and Future.ISRN Cardiol. 2012;2012:139823. doi: 10.5402/2012/139823. Epub 2012 Sep 11. ISRN Cardiol. 2012. PMID: 22997590 Free PMC article.
-
Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease.Eur Radiol. 2012 Oct;22(10):2057-66. doi: 10.1007/s00330-012-2479-1. Epub 2012 May 18. Eur Radiol. 2012. PMID: 22592808
-
Assessment of coronary artery by prospective ECG-triggered 256 multi-slice CT on children with congenital heart disease.Int J Cardiovasc Imaging. 2017 Dec;33(12):2021-2028. doi: 10.1007/s10554-017-1150-y. Epub 2017 May 18. Int J Cardiovasc Imaging. 2017. PMID: 28523472
-
Prospectively electrocardiogram-gated high-pitch spiral acquisition mode dual-source CT coronary angiography in patients with high heart rates: comparison with retrospective electrocardiogram-gated spiral acquisition mode.Korean J Radiol. 2012 Nov-Dec;13(6):684-93. doi: 10.3348/kjr.2012.13.6.684. Epub 2012 Oct 12. Korean J Radiol. 2012. PMID: 23118566 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical