Abstract
A prospective study was performed to compare the clinical and radiological results of mobile- and fixed-bearing total knee arthroplasty with specific attention to rotational alignment and range of motion. Sixty-one knees were assigned to total knee arthroplasty with either the NexGen LPS Flex fixed-bearing or with the NexGen LPS Flex mobile-bearing prosthesis. Postoperatively, knees were compared with regard to range of motion, clinical score, and radiographic findings. Rotational alignment of the femoral and tibial components was evaluated by computed tomography. The median follow-up period was 5.9 years (range 2.1-8.8 years). Median postoperative Knee Society scores were 99 points (68-100) for the fixed-bearing group and 100 points (66-100) for the mobile-bearing group (n.s.). The median postoperative flexion angles of 120° (90°-150°) for the fixed-bearing group and 125° (90°-145°) for the mobile-bearing group were not significantly different from each other (n.s.). No knee required revision surgery due to wear of polyethylene or loosening of the component in either group. Computed tomography showed that 11 knees had rotational mismatches of more than 10° between the femoral and tibial components, but no significant difference was found in the postoperative extension and flexion angles or in the clinical score between the two treatment groups. Using the identical design for both fixed- and mobile-bearing prostheses, this prospective, randomized study did not show any clinical advantages of the mobile-bearing knee. Analysis of rotational alignment by CT scan did not reveal a particular advantage of the self-aligning mechanism of mobile-bearing implants.
Similar articles
-
Fixation of high-flexion total knee prostheses: five-year follow-up results of a four-arm randomized controlled clinical and roentgen stereophotogrammetric analysis study.J Bone Joint Surg Am. 2013 Oct 2;95(19):e1411-11. doi: 10.2106/JBJS.L.01523. J Bone Joint Surg Am. 2013. PMID: 24088976 Clinical Trial.
-
LCS mobile-bearing total knee replacement. A 10-year's follow-up study.Orthop Traumatol Surg Res. 2009 May;95(3):177-82. doi: 10.1016/j.otsr.2009.02.002. Epub 2009 Mar 27. Orthop Traumatol Surg Res. 2009. PMID: 19328763
-
Mobile and fixed-bearing (all-polyethylene tibial component) total knee arthroplasty designs: surgical technique.J Bone Joint Surg Am. 2010 Sep;92 Suppl 1 Pt 2:240-9. doi: 10.2106/JBJS.J.00157. J Bone Joint Surg Am. 2010. PMID: 20844179 Clinical Trial.
-
Superior Mid- to Long-Term Clinical Outcomes of Mobile-Bearing Total Knee Arthroplasty Compared to Fixed-Bearing: A Meta-Analysis Based on a Minimum of 5 Years of Study.J Knee Surg. 2021 Oct;34(12):1368-1378. doi: 10.1055/s-0040-1709490. Epub 2020 Jun 5. J Knee Surg. 2021. PMID: 32503063
-
Fixed-bearing vs mobile-bearing prostheses for total knee arthroplasty after approximately 10 years of follow-up: a meta-analysis.J Orthop Surg Res. 2021 Jul 6;16(1):437. doi: 10.1186/s13018-021-02560-w. J Orthop Surg Res. 2021. PMID: 34229702 Free PMC article. Review.
Cited by
-
Comparison of fixed- and mobile-bearing total knee arthroplasty with a mean five-year follow-up: A meta-analysis.Exp Ther Med. 2013 Jul;6(1):45-51. doi: 10.3892/etm.2013.1122. Epub 2013 May 17. Exp Ther Med. 2013. PMID: 23935716 Free PMC article.
-
A clinical study of the rotational alignment of the femoral component in total knee arthroplasty.J Phys Ther Sci. 2015 Jul;27(7):2077-81. doi: 10.1589/jpts.27.2077. Epub 2015 Jul 22. J Phys Ther Sci. 2015. PMID: 26311929 Free PMC article.
-
Mobile-bearing versus fixed-bearing total knee arthroplasty: a meta-analysis of randomized controlled trials.Eur J Orthop Surg Traumatol. 2022 Apr;32(3):481-495. doi: 10.1007/s00590-021-02999-x. Epub 2021 May 22. Eur J Orthop Surg Traumatol. 2022. PMID: 34021791 Free PMC article.
-
Combined Rotational Alignment Change after Total Knee Arthroplasty in Different Tibial Component Designs: Implications for Optimal Tibial Component Rotational Alignment.Knee Surg Relat Res. 2018 Mar 1;30(1):74-83. doi: 10.5792/ksrr.17.048. Knee Surg Relat Res. 2018. PMID: 29482307 Free PMC article.
-
Mid-term clinical and radiological results do not differ between fixed- and mobile-bearing total knee arthroplasty using titanium-nitride-coated posterior-stabilized prostheses: a prospective randomized controlled trial.Knee Surg Sports Traumatol Arthrosc. 2019 Apr;27(4):1165-1173. doi: 10.1007/s00167-018-5095-z. Epub 2018 Aug 7. Knee Surg Sports Traumatol Arthrosc. 2019. PMID: 30088030 Clinical Trial.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous