MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences
Abstract
One of the challenges in PET/MRI is the derivation of an attenuation map to correct the PET image for attenuation. Different methods have been suggested for deriving the attenuation map from an MR image. Because the low signal intensity of cortical bone on images acquired with conventional MRI sequences makes it difficult to detect this tissue type, these methods rely on some sort of anatomic precondition to predict the attenuation map, raising the question of whether these methods will be usable in the clinic when patients may exhibit anatomic abnormalities.
Methods: We propose the use of the transverse relaxation rate, derived from images acquired with an ultrashort echo time sequence to classify the voxels into 1 of 3 tissue classes (bone, soft tissue, or air), without making any assumptions on patient anatomy. Each voxel is assigned a linear attenuation coefficient corresponding to its tissue class. A reference CT scan is used to determine the voxel-by-voxel accuracy of the proposed method. The overall accuracy of the MRI-based attenuation correction is evaluated using a method that takes into account the nonlocal effects of attenuation correction.
Results: As a proof of concept, the head of a pig was used as a phantom for imaging. The new method yielded a correct tissue classification in 90% of the voxels. Five human brain PET/CT and MRI datasets were also processed, yielding slightly worse voxel-by-voxel performance, compared to a CT-derived attenuation map. The PET datasets were reconstructed using the segmented MRI attenuation map derived with the new method, and the resulting images were compared with segmented CT-based attenuation correction. An average error of around 5% was found in the brain.
Conclusion: The feasibility of using the transverse relaxation rate map derived from ultrashort echo time MR images for the estimation of the attenuation map was shown on phantom and clinical brain data. The results indicate that the new method, compared with CT-based attenuation correction, yields clinically acceptable errors. The proposed method does not make any assumptions about patient anatomy and could therefore also be used in cases in which anatomic abnormalities are present.
Similar articles
-
MR-based attenuation correction for PET/MRI neurological studies with continuous-valued attenuation coefficients for bone through a conversion from R2* to CT-Hounsfield units.Neuroimage. 2015 May 15;112:160-168. doi: 10.1016/j.neuroimage.2015.03.009. Epub 2015 Mar 14. Neuroimage. 2015. PMID: 25776213 Free PMC article.
-
MRI-based attenuation correction for hybrid PET/MRI systems: a 4-class tissue segmentation technique using a combined ultrashort-echo-time/Dixon MRI sequence.J Nucl Med. 2012 May;53(5):796-804. doi: 10.2967/jnumed.111.092577. Epub 2012 Apr 13. J Nucl Med. 2012. PMID: 22505568
-
Magnetic resonance-based attenuation correction for PET/MR hybrid imaging using continuous valued attenuation maps.Invest Radiol. 2013 May;48(5):323-32. doi: 10.1097/RLI.0b013e318283292f. Invest Radiol. 2013. PMID: 23442772
-
Vision 20/20: Magnetic resonance imaging-guided attenuation correction in PET/MRI: Challenges, solutions, and opportunities.Med Phys. 2016 Mar;43(3):1130-55. doi: 10.1118/1.4941014. Med Phys. 2016. PMID: 26936700 Review.
-
Challenges and current methods for attenuation correction in PET/MR.MAGMA. 2013 Feb;26(1):81-98. doi: 10.1007/s10334-012-0334-7. Epub 2012 Aug 9. MAGMA. 2013. PMID: 22875599 Review.
Cited by
-
Comparison of standardized uptake values in normal structures between PET/CT and PET/MRI in an oncology patient population.Mol Imaging Biol. 2013 Dec;15(6):776-85. doi: 10.1007/s11307-013-0629-8. Mol Imaging Biol. 2013. PMID: 23632951 Free PMC article.
-
Quantitative characterizations of ultrashort echo (UTE) images for supporting air-bone separation in the head.Phys Med Biol. 2015 Apr 7;60(7):2869-80. doi: 10.1088/0031-9155/60/7/2869. Epub 2015 Mar 17. Phys Med Biol. 2015. PMID: 25776205 Free PMC article.
-
A realistic phantom of the human head for PET-MRI.EJNMMI Phys. 2020 Aug 5;7(1):52. doi: 10.1186/s40658-020-00320-z. EJNMMI Phys. 2020. PMID: 32757099 Free PMC article.
-
Accurate hybrid template-based and MR-based attenuation correction using UTE images for simultaneous PET/MR brain imaging applications.BMC Med Imaging. 2018 Nov 6;18(1):41. doi: 10.1186/s12880-018-0283-3. BMC Med Imaging. 2018. PMID: 30400875 Free PMC article.
-
A comparison of CT- and MR-based attenuation correction in neurological PET.Eur J Nucl Med Mol Imaging. 2014 Jun;41(6):1176-89. doi: 10.1007/s00259-013-2652-z. Epub 2014 Jan 15. Eur J Nucl Med Mol Imaging. 2014. PMID: 24425423
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials