Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves
- PMID: 20439787
- DOI: 10.1161/CIRCULATIONAHA.109.901272
Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves
Abstract
Background: Prosthesis-patient mismatch (P-PtM) after aortic valve replacement results in disturbed valve performance associated with increased pressure gradients. However, it is unknown whether this can be related to future structural valve deterioration (SVD) of the bioprosthesis.
Methods and results: In 564 patients (mean age, 74+/-5 years) receiving an aortic valve bioprosthesis, clinical follow-up (median, 6.1 years; maximum, 16.4 years) was analyzed including echocardiography. SVD was diagnosed in 40 patients (7%) as substantially increased stenosis (n=24) or regurgitation (n=16) of the operated valve over time. When patients with P-PtM (effective orifice area index <0.85 cm(2)/m(2); n=285) developed SVD, it was preferentially of the stenosis type, whereas when patients without P-PtM (n=279) developed SVD, the majority was of the incompetence type (P<0.05). Multivariable analysis including patient- and valve-related variables revealed that P-PtM and label size </=21 were independent predictors of SVD (P=0.04 and P=0.02, respectively). A nonparametric Turnbull estimate analysis showed that SVD is virtually nonexistent for up to 9 years in patients without P-PtM. Thereafter, SVD starts to occur and is mainly of the incompetence-type SVD (79% of cases). In patients with P-PtM, SVD starts to occur after 2 to 3 years after implantation and is mainly of the stenosis-type SVD (81% of cases).
Conclusions: These data suggest that stenosis-type SVD is an early, P-PtM-related, and thus preventable phenomenon. Incompetence-type SVD is a time-dependent, nonspecific wear damage to bioprosthetic valves, which is not related to P-PtM.
Comment in
-
Hemodynamic factors affecting the fate of valvular bioprosthesis.Circulation. 2010 May 18;121(19):2083-4. doi: 10.1161/CIRCULATIONAHA.110.954123. Epub 2010 May 3. Circulation. 2010. PMID: 20439782 No abstract available.
Similar articles
-
Lack of durability of the Mitroflow valve does not affect survival.J Heart Valve Dis. 1999 Jul;8(4):368-74; discussion 374-5. J Heart Valve Dis. 1999. PMID: 10461235
-
Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves.J Thorac Cardiovasc Surg. 2004 Jan;127(1):149-59. doi: 10.1016/j.jtcvs.2003.07.043. J Thorac Cardiovasc Surg. 2004. PMID: 14752425
-
Impact of valve prosthesis-patient mismatch on long-term survival and left ventricular mass regression after aortic valve replacement for aortic stenosis.J Card Surg. 2007 Jul-Aug;22(4):314-9. doi: 10.1111/j.1540-8191.2007.00414.x. J Card Surg. 2007. PMID: 17661773
-
The effect of prosthesis-patient mismatch on aortic bioprosthetic valve hemodynamic performance and patient clinical status.Can J Cardiol. 1996 Apr;12(4):379-87. Can J Cardiol. 1996. PMID: 8608457 Review.
-
[The best of valvular heart disease in 2006].Arch Mal Coeur Vaiss. 2007 Jan;100 Spec No 1:19-28. Arch Mal Coeur Vaiss. 2007. PMID: 17405561 Review. French.
Cited by
-
In Vitro Proof of Concept of a First-Generation Growth-Accommodating Heart Valved Conduit for Pediatric Use.Macromol Biosci. 2023 Jul;23(7):e2300011. doi: 10.1002/mabi.202300011. Epub 2023 Mar 22. Macromol Biosci. 2023. PMID: 36905285 Free PMC article.
-
An ex vivo evaluation of two different suture techniques for the Ozaki aortic neocuspidization procedure.Interact Cardiovasc Thorac Surg. 2021 Oct 4;33(4):518-524. doi: 10.1093/icvts/ivab138. Interact Cardiovasc Thorac Surg. 2021. PMID: 34089604 Free PMC article.
-
Durability of prostheses for transcatheter aortic valve implantation.Nat Rev Cardiol. 2016 Jun;13(6):360-7. doi: 10.1038/nrcardio.2016.43. Epub 2016 Apr 7. Nat Rev Cardiol. 2016. PMID: 27053461 Review.
-
Mechanism of valve failure and efficacy of reintervention through catheterization in patients with bioprosthetic valves in the pulmonary position.Ann Pediatr Cardiol. 2017 Jan-Apr;10(1):11-17. doi: 10.4103/0974-2069.197049. Ann Pediatr Cardiol. 2017. PMID: 28163423 Free PMC article.
-
Aortic valve function post-replacement of severe aortic stenosis by transcatheter procedure versus surgery: a systematic review and metanalysis.Sci Rep. 2021 Jun 7;11(1):11975. doi: 10.1038/s41598-021-91548-x. Sci Rep. 2021. PMID: 34099815 Free PMC article.