The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study
- PMID: 20444248
- PMCID: PMC2876142
- DOI: 10.1186/1471-227X-10-9
The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study
Abstract
Background: Pulse oximetry is routinely used to continuously and noninvasively monitor arterial oxygen saturation (SaO2) in critically ill patients. Although pulse oximeter oxygen saturation (SpO2) has been studied in several patient populations, including the critically ill, its accuracy has never been studied in emergency department (ED) patients with severe sepsis and septic shock. Sepsis results in characteristic microcirculatory derangements that could theoretically affect pulse oximeter accuracy. The purposes of the present study were twofold: 1) to determine the accuracy of pulse oximetry relative to SaO2 obtained from ABG in ED patients with severe sepsis and septic shock, and 2) to assess the impact of specific physiologic factors on this accuracy.
Methods: This analysis consisted of a retrospective cohort of 88 consecutive ED patients with severe sepsis who had a simultaneous arterial blood gas and an SpO2 value recorded. Adult ICU patients that were admitted from any Calgary Health Region adult ED with a pre-specified, sepsis-related admission diagnosis between October 1, 2005 and September 30, 2006, were identified. Accuracy (SpO2 - SaO2) was analyzed by the method of Bland and Altman. The effects of hypoxemia, acidosis, hyperlactatemia, anemia, and the use of vasoactive drugs on bias were determined.
Results: The cohort consisted of 88 subjects, with a mean age of 57 years (19 - 89). The mean difference (SpO2 - SaO2) was 2.75% and the standard deviation of the differences was 3.1%. Subgroup analysis demonstrated that hypoxemia (SaO2 < 90) significantly affected pulse oximeter accuracy. The mean difference was 4.9% in hypoxemic patients and 1.89% in non-hypoxemic patients (p < 0.004). In 50% (11/22) of cases in which SpO2 was in the 90-93% range the SaO2 was <90%. Though pulse oximeter accuracy was not affected by acidoisis, hyperlactatementa, anemia or vasoactive drugs, these factors worsened precision.
Conclusions: Pulse oximetry overestimates ABG-determined SaO2 by a mean of 2.75% in emergency department patients with severe sepsis and septic shock. This overestimation is exacerbated by the presence of hypoxemia. When SaO2 needs to be determined with a high degree of accuracy arterial blood gases are recommended.
Figures
Similar articles
-
Comparative Analysis of Oxygen Saturation by Pulse Oximetry and Arterial Blood Gas in Hypoxemic Patients in a Tertiary Care Hospital.Cureus. 2023 Jul 25;15(7):e42447. doi: 10.7759/cureus.42447. eCollection 2023 Jul. Cureus. 2023. PMID: 37637606 Free PMC article.
-
Pulse oximetry: fundamentals and technology update.Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014. Med Devices (Auckl). 2014. PMID: 25031547 Free PMC article. Review.
-
Accuracy of pulse oximetry in the intensive care unit.Intensive Care Med. 2001 Oct;27(10):1606-13. doi: 10.1007/s001340101064. Intensive Care Med. 2001. PMID: 11685301
-
Accuracy of pulse oximeters in detecting hypoxemia in patients with chronic thromboembolic pulmonary hypertension.PLoS One. 2015 May 15;10(5):e0126979. doi: 10.1371/journal.pone.0126979. eCollection 2015. PLoS One. 2015. PMID: 25978517 Free PMC article.
-
Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review.Cureus. 2023 Sep 27;15(9):e46078. doi: 10.7759/cureus.46078. eCollection 2023 Sep. Cureus. 2023. PMID: 37900526 Free PMC article.
Cited by
-
The role of pulse oximetry in chiropractic practice: a rationale for its use.J Chiropr Med. 2012 Jun;11(2):127-33. doi: 10.1016/j.jcm.2011.10.004. J Chiropr Med. 2012. PMID: 23204957 Free PMC article.
-
The Various Oximetric Techniques Used for the Evaluation of Blood Oxygenation.Sensors (Basel). 2020 Aug 27;20(17):4844. doi: 10.3390/s20174844. Sensors (Basel). 2020. PMID: 32867184 Free PMC article. Review.
-
Accuracy of Multiple Pulse Oximeters in Stable Critically Ill Patients.Respir Care. 2023 May;68(5):565-574. doi: 10.4187/respcare.10582. Epub 2023 Jan 3. Respir Care. 2023. PMID: 36596654 Free PMC article.
-
Comparative Analysis of Oxygen Saturation by Pulse Oximetry and Arterial Blood Gas in Hypoxemic Patients in a Tertiary Care Hospital.Cureus. 2023 Jul 25;15(7):e42447. doi: 10.7759/cureus.42447. eCollection 2023 Jul. Cureus. 2023. PMID: 37637606 Free PMC article.
-
Pulse oximetry: fundamentals and technology update.Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014. Med Devices (Auckl). 2014. PMID: 25031547 Free PMC article. Review.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources