Accuracy of amplitude integrated EEG in a neonatal cohort
- PMID: 20444809
- DOI: 10.1136/adc.2009.165969
Accuracy of amplitude integrated EEG in a neonatal cohort
Abstract
Objective: To determine the sensitivity and specificity of amplitude integrated electroencephalogram (aEEG) compared to simultaneous standard electroencephalogram (sEEG) for seizure detection and background discontinuity.
Design: Prospective paired cohort.
Setting: Tertiary academic neonatal intensive care unit.
Patients: Infants were recruited from 2005 to 2008. Neonates requiring sEEG were recruited for simultaneous aEEG.
Interventions: Following sEEG and aEEG, seizures were recorded as present or absent, and background was recorded as normal or discontinuous in each format.
Main outcome measures: Presence of at least one seizure during recording. The background activity was reported as normal or discontinuous. Discontinuity of brain activity was further ranked as mild, moderate or severe.
Results: 51 sEEG and aEEG studies were completed. 44 studies were analysed for presence of seizures and 46 were analysed for background discontinuity. Sensitivity for presence of seizures by aEEG was 80% and specificity was 50%. The proportion of infants with seizures were overdiagnosed by aEEG (63.6% vs 45.5% for sEEG p=0.045). Discontinuity of background activity had higher sensitivity (88.6%) and specificity (54.5%) when compared with seizure detection. When stratified by indication for EEG, hypoxic episode (n=14) or suspected seizures (n=33), similar sensitivity for presence of seizure (80%) was noted by aEEG and sEEG. However the specificity of aEEG for seizure detection was higher in neonates undergoing EEG for suspected seizures (66.7% vs 22.2%).
Conclusions: Background abnormalities were detected with fair accuracy by aEEG but aEEG criteria alone would result in the overdiagnosis of neonatal seizures. Therefore seizures noted on aEEG require sEEG confirmation prior to implementing anticonvulsant therapy for neonatal seizures.
Similar articles
-
Utility of prolonged bedside amplitude-integrated encephalogram in encephalopathic infants.Am J Perinatol. 2008 Nov;25(10):611-5. doi: 10.1055/s-0028-1090598. Epub 2008 Oct 7. Am J Perinatol. 2008. PMID: 18841534
-
Additional value of two-channel amplitude integrated EEG recording in full-term infants with unilateral brain injury.Arch Dis Child Fetal Neonatal Ed. 2010 May;95(3):F160-8. doi: 10.1136/adc.2008.156711. Epub 2009 Oct 8. Arch Dis Child Fetal Neonatal Ed. 2010. PMID: 19815938
-
Accuracy of bedside electroencephalographic monitoring in comparison with simultaneous continuous conventional electroencephalography for seizure detection in term infants.Pediatrics. 2008 Jun;121(6):1146-54. doi: 10.1542/peds.2007-1839. Pediatrics. 2008. PMID: 18519484
-
Using amplitude-integrated EEG in neonatal intensive care.J Perinatol. 2010 Oct;30 Suppl:S73-81. doi: 10.1038/jp.2010.93. J Perinatol. 2010. PMID: 20877412 Review.
-
Brain monitoring in neonates.Early Hum Dev. 2009 Feb;85(2):77-84. doi: 10.1016/j.earlhumdev.2008.11.007. Epub 2009 Jan 17. Early Hum Dev. 2009. PMID: 19150756 Review.
Cited by
-
Lost in Transition: A Systematic Review of Neonatal Electroencephalography in the Delivery Room-Are We Forgetting an Important Biomarker for Newborn Brain Health?Front Pediatr. 2017 Aug 10;5:173. doi: 10.3389/fped.2017.00173. eCollection 2017. Front Pediatr. 2017. PMID: 28848727 Free PMC article. Review.
-
Electroencephalographic monitoring in the pediatric intensive care unit.Curr Neurol Neurosci Rep. 2013 Mar;13(3):330. doi: 10.1007/s11910-012-0330-3. Curr Neurol Neurosci Rep. 2013. PMID: 23335026 Free PMC article. Review.
-
Effect of Treatment of Clinical Seizures vs Electrographic Seizures in Full-Term and Near-Term Neonates: A Randomized Clinical Trial.JAMA Netw Open. 2021 Dec 1;4(12):e2139604. doi: 10.1001/jamanetworkopen.2021.39604. JAMA Netw Open. 2021. PMID: 34919132 Free PMC article. Clinical Trial.
-
Seizure Detection by Critical Care Providers Using Amplitude-Integrated Electroencephalography and Color Density Spectral Array in Pediatric Cardiac Arrest Patients.Pediatr Crit Care Med. 2017 Apr;18(4):363-369. doi: 10.1097/PCC.0000000000001099. Pediatr Crit Care Med. 2017. PMID: 28234810 Free PMC article. Clinical Trial.
-
Neonatal seizures and status epilepticus.J Clin Neurophysiol. 2012 Oct;29(5):441-8. doi: 10.1097/WNP.0b013e31826bd90d. J Clin Neurophysiol. 2012. PMID: 23027101 Free PMC article. Review.