Intraexaminer and interexaminer reliabilities of landmark identification on digitized lateral cephalograms and formatted 3-dimensional cone-beam computerized tomography images
- PMID: 20451778
- DOI: 10.1016/j.ajodo.2008.07.018
Intraexaminer and interexaminer reliabilities of landmark identification on digitized lateral cephalograms and formatted 3-dimensional cone-beam computerized tomography images
Abstract
Introduction: The purposes of this study were to determine and compare the intraexaminer and interexaminer reliabilities of commonly used cephalometric landmarks identified on digitized lateral cephalograms and formatted cone-beam computerized tomography (CBCT) images.
Methods: CBCT images from 10 randomly selected adolescent patients were obtained from the orthodontic records of a private practice. Measurement errors, and intraexaminer, and interexaminer reliability correlation coefficients (ICC) were obtained for all landmark coordinates.
Results: Intraexaminer and interexaminer reliabilities for all coordinates for most landmarks on the digital lateral cephalograms and CBCT images were greater than 0.9 (ICC value). The means of landmark locations differed by approximately 1 mm in most coordinates from the lateral cephalograms and were predominantly higher than 1 mm for all coordinates from the CBCT images.
Conclusions: Intraexaminer and interexaminer reliabilities were high for most landmarks. Coordinates with greater measurement errors in the lateral cephalograms (condylion, gonion, porion, mandibular incisor apex, and posterior nasal spine) were in structures without clearly defined borders. In the CBCT images, gonion, condylion, and porion were located on surfaces that were flat or curved, making it difficult to recognize a specific reference point. Other less reliable landmarks (anterior nasal spine, posterior nasal spine, mandibular incisor apex) were located in structures with lower densities and could not be visualized with 3-dimensional reconstruction; thus, they had high measurement errors.
Copyright (c) 2010 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.
Similar articles
-
Comparison between conventional and cone-beam computed tomography-generated cephalograms.Am J Orthod Dentofacial Orthop. 2008 Dec;134(6):798-802. doi: 10.1016/j.ajodo.2008.07.008. Am J Orthod Dentofacial Orthop. 2008. PMID: 19061807
-
The reliability of cone-beam computed tomography (CBCT) - generated frontal cephalograms.J Craniomaxillofac Surg. 2012 Dec;40(8):e331-6. doi: 10.1016/j.jcms.2012.01.016. Epub 2012 Mar 22. J Craniomaxillofac Surg. 2012. PMID: 22444351
-
Reliability and accuracy of cone-beam computed tomography dental measurements.Am J Orthod Dentofacial Orthop. 2009 Jul;136(1):19-25; discussion 25-8. doi: 10.1016/j.ajodo.2007.09.016. Am J Orthod Dentofacial Orthop. 2009. PMID: 19577143
-
Accuracy of conventional versus cone-beam CT-synthesised lateral cephalograms for cephalometric analysis: A systematic review.J Orthod. 2024 Jun;51(2):160-176. doi: 10.1177/14653125231178038. Epub 2023 Jun 21. J Orthod. 2024. PMID: 37340975
-
Facial asymmetry quantitative evaluation in oculoauriculovertebral spectrum.Clin Oral Investig. 2016 Mar;20(2):219-25. doi: 10.1007/s00784-015-1660-8. Epub 2015 Nov 18. Clin Oral Investig. 2016. PMID: 26578120 Review.
Cited by
-
The Reproducibility of Reference Landmarks in the External Acoustic Meatus (EAM) on Cone Beam Computed Tomography (CBCT) Images.J Clin Med. 2024 Jul 19;13(14):4226. doi: 10.3390/jcm13144226. J Clin Med. 2024. PMID: 39064266 Free PMC article.
-
Intra- and Interexaminer Measurement Variability Analysis of an Orthodontic Gauge Device to Determine Incisor Occlusal Surface Angles in the Horse.Vet Sci. 2022 Sep 7;9(9):481. doi: 10.3390/vetsci9090481. Vet Sci. 2022. PMID: 36136698 Free PMC article.
-
Reliability of different three-dimensional cephalometric landmarks in cone-beam computed tomography : A systematic review.Angle Orthod. 2019 Mar;89(2):317-332. doi: 10.2319/042018-302.1. Epub 2018 Nov 13. Angle Orthod. 2019. PMID: 30423256 Free PMC article.
-
[Use of skeletal-related diagnostic methods in articles published from 2018 to 2020 in four major orthodontics journals].Rev Cient Odontol (Lima). 2021 Mar 11;9(1):e041. doi: 10.21142/2523-2754-0901-2021-041. eCollection 2021 Jan-Mar. Rev Cient Odontol (Lima). 2021. PMID: 38464403 Free PMC article. Spanish.
-
Influence of Head Motion on the Accuracy of 3D Reconstruction with Cone-Beam CT: Landmark Identification Errors in Maxillofacial Surface Model.PLoS One. 2016 Apr 11;11(4):e0153210. doi: 10.1371/journal.pone.0153210. eCollection 2016. PLoS One. 2016. PMID: 27065238 Free PMC article.