Evaluation of a new chemiluminescence immunoassay for diagnosis of syphilis
- PMID: 20452886
- PMCID: PMC3352047
- DOI: 10.1186/2047-783x-15-2-66
Evaluation of a new chemiluminescence immunoassay for diagnosis of syphilis
Abstract
Objective: To assess the sensitivity, specificity, and feasibility of a new chemiluminescence immunoassay (CLIA) in the diagnosis of syphilis.
Methods: At first, a retrospective study was conducted, using 135 documented cases of syphilis and 30 potentially interfering samples and 80 normal sera. A prospective study was also performed by testing 2, 071 unselected samples for routine screening for syphilis. CLIA was compared with a nontreponemal test (TRUST) and a treponemal test (TPPA).
Results: There was an agreement of 100% between CLIA and TPPA in the respective study. The percentage of agreement among the 245 sera tested was 100.0%. Compared with TPPA, the specificity of CLIA was 99.9% (1817/1819), the sensitivity of CLIA was 100.0% (244/244) in the prospective study. CLIA showed 99.5% agreement with TPPA by testing 2, 071 unselected samples. And CLIA seemed to be more sensitive than TPPA in detecting the samples of primary syphilis.
Conclusions: CLIA is easy to perform and the indicator results are objective and unequivocal. It may be suitable for large-scale screening as a treponemal test substituted for TPPA.
References
-
- de Larrañaga G, Trombetta L, Wingeyer SP, Remondino G. False positive reactions in confirmatory tests for syphilis in presence of antiphospholipid antibodies: misdiagnosis with prognostic and social consequences. Dermatol Online J. 2006;12(4):22. - PubMed
-
- Marks R, Thomas-Kaskel AK, Schmidt D. Steroid refractory episcleritis as early manifestation of neurosyphilis. Eur J Med Res. 2006;11(7):309–312. - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical