Evaluation of acute fixation strength of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant
Abstract
Introduction: The purpose of this study was to evaluate the acute fixation strength of fibrin sealant as an alternative fixation method for laparoscopic ventral hernia repair (LVHR) when utilized with absorbable and nonabsorbable barrier meshes.
Methods: Representative mesh types for LVHR included one nonabsorbable barrier mesh (Bard Composix) and three absorbable barrier meshes (Sepramesh, Proceed, and Parietex Composite Mesh). Macroporous polypropylene mesh (Prolite Ultra) served as a control. An associated pilot study demonstrated that acute fixation was not achieved with C-Qur or Dualmesh, thus these meshes were not subjected to further analysis. Two fibrin sealants varying in thrombin concentration (TISSEEL 500 IU/ml or ARTISS 4 IU/ml) were each utilized to secure a 3 x 4 cm piece of mesh to the peritoneal surface of harvested abdominal wall from New Zealand white rabbits. Ten samples were prepared for each mesh and fibrin sealant combination. The resulting mesh-tissue interface (3 x 3 cm) underwent lap shear testing at a rate of 0.42 mm/s using a tensiometer (Instron 5542, Norwood, MA). The maximum load sustained by the construct was recorded as the acute fixation strength in units of Newtons (N). Data are given as means +/- SEM. Statistical significance (P < 0.05) was determined using a two-way ANOVA and Bonferroni post-test.
Results: There was no significant difference (P > 0.05) in the acute fixation strength achieved by ARTISS versus TISSEEL with Composix, Sepramesh, Proceed, and Parietex Composite Mesh. However, Prolite Ultra fixed with ARTISS (7.099 +/- 1.01 N) had a significantly higher tensile strength (P < 0.001) than Prolite Ultra fixed with TISSEEL (3.405 +/- 0.72 N). Among meshes fixed with TISSEEL, Parietex Composite (3.936 +/- 0.73 N) was significantly stronger than Sepramesh (1.760 +/- 0.33 N) (P < 0.05).
Conclusions: Acute fixation strength is equivalent for TISSEEL and ARTISS fibrin sealants for selected nonabsorbable (Composix) and absorbable barrier-coated (Sepramesh, Proceed, and Parietex Composite Mesh) meshes. Previous evaluations in this model indicated fibrin sealant alone did not appear to be a suitable method of fixation with C-Qur or DualMesh against the peritoneum. Long-term evaluation is needed to evaluate the efficacy of this method of fixation using absorbable and nonabsorbable barrier-coated meshes in a model of LVHR.
Similar articles
-
Evaluation of intraperitoneal placement of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant in a New Zealand white rabbit model.Surg Endosc. 2011 Feb;25(2):604-12. doi: 10.1007/s00464-010-1230-8. Epub 2010 Jul 22. Surg Endosc. 2011. PMID: 20652323
-
Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair.Surg Endosc. 2011 May;25(5):1541-52. doi: 10.1007/s00464-010-1432-0. Epub 2010 Oct 26. Surg Endosc. 2011. PMID: 20976479
-
Histologic evaluation of absorbable and non-absorbable barrier coated mesh secured to the peritoneum with fibrin sealant in a New Zealand white rabbit model.Hernia. 2011 Dec;15(6):677-84. doi: 10.1007/s10029-011-0834-9. Epub 2011 May 24. Hernia. 2011. PMID: 21607571 Free PMC article.
-
A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair.Surg Endosc. 2012 Feb;26(2):566-75. doi: 10.1007/s00464-011-1899-3. Epub 2011 Sep 5. Surg Endosc. 2012. PMID: 21898010 Review.
-
[The problem of mesh shrinkage in laparoscopic incisional hernia repair].Zentralbl Chir. 2009 Jun;134(3):209-13. doi: 10.1055/s-0028-1098779. Epub 2009 Jun 17. Zentralbl Chir. 2009. PMID: 19536713 Review. German.
Cited by
-
Effectiveness and safety of n-butyl-2-cyanoacrylate medical adhesive for noninvasive patch fixation in laparoscopic inguinal hernia repair.Surg Endosc. 2013 Oct;27(10):3792-8. doi: 10.1007/s00464-013-2970-z. Epub 2013 May 10. Surg Endosc. 2013. PMID: 23660719
-
Mesh fixation using novel bio-adhesive coating compared to tack fixation for IPOM hernia repair: in vivo evaluation in a porcine model.Surg Endosc. 2019 Jul;33(7):2364-2375. doi: 10.1007/s00464-019-06806-x. Epub 2019 May 8. Surg Endosc. 2019. PMID: 31069501
-
The impact of hydrophobic hernia mesh coating by omega fatty acid on atraumatic fibrin sealant fixation.Hernia. 2015 Aug;19(4):651-7. doi: 10.1007/s10029-014-1304-y. Epub 2014 Sep 7. Hernia. 2015. PMID: 25194586
-
Fibrin sealant for mesh fixation in laparoscopic umbilical hernia repair: 1-year results of a randomized controlled double-blinded study.Hernia. 2013 Aug;17(4):511-4. doi: 10.1007/s10029-013-1101-z. Epub 2013 May 9. Hernia. 2013. PMID: 23657861 Clinical Trial.
-
Use of fibrin sealant (Tisseel/Tissucol) in hernia repair: a systematic review.Surg Endosc. 2012 Jul;26(7):1803-12. doi: 10.1007/s00464-012-2156-0. Epub 2012 Jan 26. Surg Endosc. 2012. PMID: 22278103
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials