The impact of overscan on patient dose with first generation multislice CT scanners
- PMID: 20456992
- DOI: 10.1016/j.ejmp.2010.03.001
The impact of overscan on patient dose with first generation multislice CT scanners
Abstract
Helical scanning requires the irradiation of larger lengths than those planned. This is referred to as overscan and results to an increase of patient dose. Its impact on patient dose was investigated for three first generation multislice CT scanners; a six-, a quad- and a dual-slice. The amount of overscan was determined using the scanners' dose-length product (DLP) indications and films positioned on the CT table. With the preset protocol for the chest examination selected in all CT scanners, the overscan length calculated from the DLP indications was 6.3, 3.5 and 2cm respectively, whereas the corresponding figures derived from the films were 6.6, 4.8 and 2.5 to 3.2cm. For a 30cm scan length, the respective contributions of overscan to the DLP values were 17, 10 and 6%, whereas for a scan length of 20cm the respective values increased to 24, 15 and 9%. For the smallest scan lengths allowed in helical mode, the respective contributions reached 53, 88 and 67% because for the six-slice scanner the smallest scan length was limited to twice the collimation, whereas in the quad and dual scanners no limitation existed. For small scan lengths the presence of overscan cancels out any dose reduction offered by helical scanning with pitch factor values larger than one and therefore the axial mode should be preferred, when this is not prohibited by the diagnostic task in question.
Copyright © 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Primary radiation outside the imaged volume of a multislice helical CT scan.Br J Radiol. 2002 Jun;75(894):518-22. doi: 10.1259/bjr.75.894.750518. Br J Radiol. 2002. PMID: 12124239
-
Dual energy CT of the chest: how about the dose?Invest Radiol. 2010 Jun;45(6):347-53. doi: 10.1097/RLI.0b013e3181df901d. Invest Radiol. 2010. PMID: 20404737
-
Saving dose in triple-rule-out computed tomography examination using a high-pitch dual spiral technique.Invest Radiol. 2010 Feb;45(2):64-71. doi: 10.1097/RLI.0b013e3181c15842. Invest Radiol. 2010. PMID: 20027121
-
Scanning protocols for multislice CT scanners.Br J Radiol. 2004;77 Spec No 1:S2-9. doi: 10.1259/bjr/28755689. Br J Radiol. 2004. PMID: 15546838 Review.
-
Multislice CT: technical principles and future trends.Eur Radiol. 2003 Dec;13 Suppl 5:M3-13. doi: 10.1007/s00330-003-2178-z. Eur Radiol. 2003. PMID: 14989605 Review.
Cited by
-
Patient centring and scan length: how inaccurate practice impacts on radiation dose in CT colonography (CTC).Radiol Med. 2019 Aug;124(8):762-767. doi: 10.1007/s11547-019-01021-z. Epub 2019 Mar 8. Radiol Med. 2019. PMID: 30848421
-
Open-Source Radiation Exposure Extraction Engine (RE3) with Patient-Specific Outlier Detection.J Digit Imaging. 2016 Aug;29(4):406-19. doi: 10.1007/s10278-015-9852-y. J Digit Imaging. 2016. PMID: 26644157 Free PMC article.
-
Overranging and overbeaming measurement in area detector computed tomography: A method for simultaneous measurement in volume helical acquisition.J Appl Clin Med Phys. 2019 Jul;20(7):160-165. doi: 10.1002/acm2.12650. Epub 2019 Jun 4. J Appl Clin Med Phys. 2019. PMID: 31165567 Free PMC article.
-
Efficacy of a dynamic collimator for overranging dose reduction in a second- and third-generation dual source CT scanner.Eur Radiol. 2017 Sep;27(9):3618-3624. doi: 10.1007/s00330-017-4745-8. Epub 2017 Jan 26. Eur Radiol. 2017. PMID: 28127643 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical